Activated Clotting Time Requires Adaptation Across Altered Measurement Devices: Determination of Appropriate Range During Atrial Fibrillation Ablation.

IF 2.3 4区 医学 Q2 HEMATOLOGY
Haruna Sakanoue, Hirosuke Yamaji, Sayaka Okamoto, Kumi Okano, Yuka Fujita, Shunichi Higashiya, Takashi Murakami, Satoshi Hirohata, Shozo Kusachi
{"title":"Activated Clotting Time Requires Adaptation Across Altered Measurement Devices: Determination of Appropriate Range During Atrial Fibrillation Ablation.","authors":"Haruna Sakanoue, Hirosuke Yamaji, Sayaka Okamoto, Kumi Okano, Yuka Fujita, Shunichi Higashiya, Takashi Murakami, Satoshi Hirohata, Shozo Kusachi","doi":"10.1177/10760296251332938","DOIUrl":null,"url":null,"abstract":"<p><p>BackgroundMethods for measuring activated clotting time (ACT) are not yet standardized.ObjectivesTo adjust and compare values between two measurement systems and to optimize ACT during atrial fibrillation (AF) ablation.MethodsTwo systems were compared: electromagnetic detection using a rotating tube (EM system; Hemochron Response) and photo-optical detection using a cartridge immersed in blood (PO system; ACT CA-300TM).ResultsACT was measured simultaneously in 124 instances in 53 patients before and during AF ablations using both methods. A linear regression analysis showed ACT (EM system) = 1.19 × ACT (PO system) + 9.03 (p < .001, r = 0.90). Bland-Altman plots indicated an average difference of 50 s between the two systems. In 3364 ACT measurements from 1161 ablations, the EM system recorded a mean ACT of 320 ± 44 s (range 156-487 s). Estimating the target range as mean ± 1 SD range, the EM system's range was 275-365 s, in 5-s increments. The pre-ablation ACT measured on the EM system was 143 ± 28 s (115-170 s). Cardiac tamponade occurred in 4 out of 2085 ablations (0.19%) over 5 years, with ACT values ranging from 330 to 391 s on the EM system. Based on these findings, the estimated optimal ACT range for the PO system was adjusted to 225-300 s to align with the EM system's range of 275-365 s.ConclusionsACT target ranges should be system-specific, and direct extrapolation between devices is not recommended. Adjustment is clinically necessary when switching systems.</p>","PeriodicalId":10335,"journal":{"name":"Clinical and Applied Thrombosis/Hemostasis","volume":"31 ","pages":"10760296251332938"},"PeriodicalIF":2.3000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11954560/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical and Applied Thrombosis/Hemostasis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/10760296251332938","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/3/29 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

BackgroundMethods for measuring activated clotting time (ACT) are not yet standardized.ObjectivesTo adjust and compare values between two measurement systems and to optimize ACT during atrial fibrillation (AF) ablation.MethodsTwo systems were compared: electromagnetic detection using a rotating tube (EM system; Hemochron Response) and photo-optical detection using a cartridge immersed in blood (PO system; ACT CA-300TM).ResultsACT was measured simultaneously in 124 instances in 53 patients before and during AF ablations using both methods. A linear regression analysis showed ACT (EM system) = 1.19 × ACT (PO system) + 9.03 (p < .001, r = 0.90). Bland-Altman plots indicated an average difference of 50 s between the two systems. In 3364 ACT measurements from 1161 ablations, the EM system recorded a mean ACT of 320 ± 44 s (range 156-487 s). Estimating the target range as mean ± 1 SD range, the EM system's range was 275-365 s, in 5-s increments. The pre-ablation ACT measured on the EM system was 143 ± 28 s (115-170 s). Cardiac tamponade occurred in 4 out of 2085 ablations (0.19%) over 5 years, with ACT values ranging from 330 to 391 s on the EM system. Based on these findings, the estimated optimal ACT range for the PO system was adjusted to 225-300 s to align with the EM system's range of 275-365 s.ConclusionsACT target ranges should be system-specific, and direct extrapolation between devices is not recommended. Adjustment is clinically necessary when switching systems.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
4.40
自引率
3.40%
发文量
150
审稿时长
2 months
期刊介绍: CATH is a peer-reviewed bi-monthly journal that addresses the practical clinical and laboratory issues involved in managing bleeding and clotting disorders, especially those related to thrombosis, hemostasis, and vascular disorders. CATH covers clinical trials, studies on etiology, pathophysiology, diagnosis and treatment of thrombohemorrhagic disorders.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信