{"title":"Effectiveness of Radiation Shields to Minimize Operator Dose in the Bronchoscopy Suite: A Phantom Study and Clinical Application.","authors":"Hosang Jeon, Dong Woon Kim, Ji Hyeon Joo, Yongkan Ki, Suk-Woong Kang, Won Chul Shin, Seong Hoon Yoon, Yun Seong Kim, Seung Hyun Yong, Hyun Sung Chung, Taehoon Lee, Hee Yun Seol","doi":"10.3390/jcm14062114","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background/Objectives</b>: Fluoroscopy has been widely adopted in interventional pulmonology, as it facilitates real-time visualization of the bronchoscope, endobronchial ultrasound, and biopsy tools during procedures. The purpose of this study was to evaluate the effectiveness of radiation shields in minimizing scattered X-ray dose to the bronchoscopist in a phantom study and to determine the dose of scattered X-ray dose to medical staff with radiation shields in clinical application. <b>Methods</b>: An anthropomorphic torso phantom was positioned on the fluoroscopic table between the C-arm X-ray tube and the image detector to mimic bronchoscopic operations. Upper and lower body lead shields were used to examine the effectiveness of radiation shielding. Scatter radiation rates were assessed at a first operator location using real-time dosimeters with and without protective devices. In clinical application, the scattered X-ray dose of the first operator and main assistant was measured using wearable radiation dosimeters during 20 procedures. <b>Results</b>: In the phantom study, scattered radiation without shielding was 266.34 ± 8.86 μSv/h (glabella), 483.90 ± 8.01 μSv/h (upper thorax), 143.97 ± 8.20 μSv/h (hypogastrium), and 7.22 ± 0.28 μSv/h (ankle). The combination of upper and lower body lead shields reduced the scattered X-ray dose by 98.7%, 98.3%, 66.2%, and 79.9% at these levels, respectively. In clinical application, mean scattered X-ray dose rates were 0.14 ± 0.05 μSv/procedure (eye), 0.46 ± 0.51 μSv/procedure (chest), 0.67 ± 0.50 μSv/procedure (hypogastrium), and 1.57 ± 2.84 μSv/procedure (assistant's wrist). <b>Conclusions</b>: The combination of radiation shields significantly reduced the scattered X-ray dose at the operator site in the phantom study. The scattered X-ray dose to medical staff during bronchoscopy can be kept at a low level with the aid of a shielding system.</p>","PeriodicalId":15533,"journal":{"name":"Journal of Clinical Medicine","volume":"14 6","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-03-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11943078/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/jcm14062114","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background/Objectives: Fluoroscopy has been widely adopted in interventional pulmonology, as it facilitates real-time visualization of the bronchoscope, endobronchial ultrasound, and biopsy tools during procedures. The purpose of this study was to evaluate the effectiveness of radiation shields in minimizing scattered X-ray dose to the bronchoscopist in a phantom study and to determine the dose of scattered X-ray dose to medical staff with radiation shields in clinical application. Methods: An anthropomorphic torso phantom was positioned on the fluoroscopic table between the C-arm X-ray tube and the image detector to mimic bronchoscopic operations. Upper and lower body lead shields were used to examine the effectiveness of radiation shielding. Scatter radiation rates were assessed at a first operator location using real-time dosimeters with and without protective devices. In clinical application, the scattered X-ray dose of the first operator and main assistant was measured using wearable radiation dosimeters during 20 procedures. Results: In the phantom study, scattered radiation without shielding was 266.34 ± 8.86 μSv/h (glabella), 483.90 ± 8.01 μSv/h (upper thorax), 143.97 ± 8.20 μSv/h (hypogastrium), and 7.22 ± 0.28 μSv/h (ankle). The combination of upper and lower body lead shields reduced the scattered X-ray dose by 98.7%, 98.3%, 66.2%, and 79.9% at these levels, respectively. In clinical application, mean scattered X-ray dose rates were 0.14 ± 0.05 μSv/procedure (eye), 0.46 ± 0.51 μSv/procedure (chest), 0.67 ± 0.50 μSv/procedure (hypogastrium), and 1.57 ± 2.84 μSv/procedure (assistant's wrist). Conclusions: The combination of radiation shields significantly reduced the scattered X-ray dose at the operator site in the phantom study. The scattered X-ray dose to medical staff during bronchoscopy can be kept at a low level with the aid of a shielding system.
期刊介绍:
Journal of Clinical Medicine (ISSN 2077-0383), is an international scientific open access journal, providing a platform for advances in health care/clinical practices, the study of direct observation of patients and general medical research. This multi-disciplinary journal is aimed at a wide audience of medical researchers and healthcare professionals.
Unique features of this journal:
manuscripts regarding original research and ideas will be particularly welcomed.JCM also accepts reviews, communications, and short notes.
There is no limit to publication length: our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible.