Rescue and Redeploy: Successful Valve-in-Valve Implantation After Transcatheter Aortic Valve Embolization.

IF 1.7 2区 医学 Q3 PERIPHERAL VASCULAR DISEASE
Ahmad Abdelrehim, Ameer Abdelrahman, Ahmad Almagazzachi, Harit Desai, David Drucker
{"title":"Rescue and Redeploy: Successful Valve-in-Valve Implantation After Transcatheter Aortic Valve Embolization.","authors":"Ahmad Abdelrehim, Ameer Abdelrahman, Ahmad Almagazzachi, Harit Desai, David Drucker","doi":"10.1177/15266028251325656","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pacing failure is a common cause of transcatheter aortic valve replacement (TAVR) malposition and embolization. Transcatheter aortic valve embolization that compromises aortic branches usually necessitates open surgery. However, a salvage technique can address this complication percutaneously.</p><p><strong>Case summary: </strong>We report a case involving a 26 mm Edwards transcatheter valve that was cephalically embolized during rapid pacing due to ventricular tachycardia with a single capture beat. The semiinflated embolized valve was successfully pulled back to a position just distal to the left subclavian artery. Subsequently, a second 26 mm valve was implanted through the embolized valve and deployed successfully in the correct position.</p><p><strong>Conclusion: </strong>Carefully pulling back the embolized transcatheter aortic valve (TAV) to the area distal to the left subclavian artery is a viable salvage technique for embolized TAV. Another attempt of TAVR through the embolized valve can be safely performed, with careful monitoring of left ventricular capture during rapid pacing.Clinical Impact<b>Question:</b> Can transcatheter aortic valve (TAV) embolization be solved percutaneously? <b>Findings:</b> The semi-inflated embolized valve was successfully pulled back to a position just distal to the left subclavian artery. Subsequently, a second 26 mm valve was implanted through the embolized valve and deployed successfully in the correct position. <b>Results:</b> Carefully pulling back an embolized TAV to the descending aorta is a viable salvage technique for managing embolized TAV.</p>","PeriodicalId":50210,"journal":{"name":"Journal of Endovascular Therapy","volume":" ","pages":"15266028251325656"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Endovascular Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15266028251325656","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Pacing failure is a common cause of transcatheter aortic valve replacement (TAVR) malposition and embolization. Transcatheter aortic valve embolization that compromises aortic branches usually necessitates open surgery. However, a salvage technique can address this complication percutaneously.

Case summary: We report a case involving a 26 mm Edwards transcatheter valve that was cephalically embolized during rapid pacing due to ventricular tachycardia with a single capture beat. The semiinflated embolized valve was successfully pulled back to a position just distal to the left subclavian artery. Subsequently, a second 26 mm valve was implanted through the embolized valve and deployed successfully in the correct position.

Conclusion: Carefully pulling back the embolized transcatheter aortic valve (TAV) to the area distal to the left subclavian artery is a viable salvage technique for embolized TAV. Another attempt of TAVR through the embolized valve can be safely performed, with careful monitoring of left ventricular capture during rapid pacing.Clinical ImpactQuestion: Can transcatheter aortic valve (TAV) embolization be solved percutaneously? Findings: The semi-inflated embolized valve was successfully pulled back to a position just distal to the left subclavian artery. Subsequently, a second 26 mm valve was implanted through the embolized valve and deployed successfully in the correct position. Results: Carefully pulling back an embolized TAV to the descending aorta is a viable salvage technique for managing embolized TAV.

背景:起搏失败是经导管主动脉瓣置换术(TAVR)定位不良和栓塞的常见原因。经导管主动脉瓣栓塞危及主动脉分支,通常需要进行开放手术。病例摘要:我们报告了一个病例,该病例涉及一个 26 毫米 Edwards 经导管瓣膜,在快速起搏过程中因单次捕获性室性心动过速而发生头端栓塞。半充气的栓塞瓣膜被成功拉回至左锁骨下动脉远端位置。随后,通过栓塞的瓣膜植入了第二个26毫米瓣膜,并在正确位置成功展开:结论:将栓塞的经导管主动脉瓣(TAV)小心地拉回至左锁骨下动脉远端是栓塞TAV的一种可行抢救技术。通过栓塞的瓣膜再次尝试 TAVR 可以安全进行,同时在快速起搏时仔细监测左心室捕获情况:经导管主动脉瓣(TAV)栓塞能否经皮解决?研究结果:半充气的栓塞瓣膜被成功拉回至左锁骨下动脉远端位置。随后,通过栓塞的瓣膜植入第二个 26 毫米瓣膜,并在正确位置成功展开。手术结果将栓塞的 TAV 小心地拉回至降主动脉是处理栓塞 TAV 的一种可行抢救技术。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.30
自引率
15.40%
发文量
203
审稿时长
6-12 weeks
期刊介绍: The Journal of Endovascular Therapy (formerly the Journal of Endovascular Surgery) was established in 1994 as a forum for all physicians, scientists, and allied healthcare professionals who are engaged or interested in peripheral endovascular techniques and technology. An official publication of the International Society of Endovascular Specialists (ISEVS), the Journal of Endovascular Therapy publishes peer-reviewed articles of interest to clinicians and researchers in the field of peripheral endovascular interventions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信