Kevin G Buda, Yader Sandoval, Stephen W Smith, Barrett Wagner, Karen Schulz, Anne Sexter, Fred S Apple
{"title":"Novel Mindray high sensitivity cardiac troponin I assay for single sample and 0/2-hour rule out of myocardial infarction: MERITnI study.","authors":"Kevin G Buda, Yader Sandoval, Stephen W Smith, Barrett Wagner, Karen Schulz, Anne Sexter, Fred S Apple","doi":"10.1515/cclm-2024-1387","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>We determined the efficacy of a high sensitivity cardiac troponin I (hs-cTnI) assay for newly derived 0 h and 0/2-h rule-out concentrations for myocardial infarction and determined the safety of incremental changes at low concentrations.</p><p><strong>Methods: </strong>Consecutive, emergency department patients undergoing serial hs-cTnI testing on clinical indication were studied in the 'Mindray hs-cTnI Assay Analytical and Clinical Evaluation for the Diagnosis and RIsk Assessment of Myocardial InfarctIon' (MERITnI) trial. Primary safety outcome was the composite of cardiac death and MI at 30 days.</p><p><strong>Results: </strong>In 1,556 patients (60.7 % male, 43.3 % White, 45.8 % Black. 34.8 % chest pain), 26.9 % patients had at least one hs-cTnI >99th percentile. 2.7 % had type 1 MI, 2.7 % type 2 MI, and 21.5 % non-MI myocardial injury. Single-sample MI rule-out using a normal ECG plus hs-cTnI of <2 ng/L (rounded) ruled out 13.8 % of patients, including early (<2 h) presenters. A 2-h delta of <2 ng/L ruled out an additional 39 % of patients. Based on European Society of Cardiology guidance, derived 0 h<15 ng/L and derived delta of <5 ng/L ruled out 42 % of patients. The Mindray assay showed incremental (non-rounded) analysis discriminated rule out from baseline of <0.1 ng/L at 1.2 % to 2.0 ng/L at 17.0 %. There were no missed adverse outcomes at 30-day assessment for composite of MI and cardiac death.</p><p><strong>Conclusions: </strong>The novel Mindray hs-cTnI assay enabled safe and early rule out of MI and cardiac death at very low concentrations in a diverse, cohort utilizing both single sample and 0/2-h rule out protocols, including early presenters.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2025-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry and laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1515/cclm-2024-1387","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: We determined the efficacy of a high sensitivity cardiac troponin I (hs-cTnI) assay for newly derived 0 h and 0/2-h rule-out concentrations for myocardial infarction and determined the safety of incremental changes at low concentrations.
Methods: Consecutive, emergency department patients undergoing serial hs-cTnI testing on clinical indication were studied in the 'Mindray hs-cTnI Assay Analytical and Clinical Evaluation for the Diagnosis and RIsk Assessment of Myocardial InfarctIon' (MERITnI) trial. Primary safety outcome was the composite of cardiac death and MI at 30 days.
Results: In 1,556 patients (60.7 % male, 43.3 % White, 45.8 % Black. 34.8 % chest pain), 26.9 % patients had at least one hs-cTnI >99th percentile. 2.7 % had type 1 MI, 2.7 % type 2 MI, and 21.5 % non-MI myocardial injury. Single-sample MI rule-out using a normal ECG plus hs-cTnI of <2 ng/L (rounded) ruled out 13.8 % of patients, including early (<2 h) presenters. A 2-h delta of <2 ng/L ruled out an additional 39 % of patients. Based on European Society of Cardiology guidance, derived 0 h<15 ng/L and derived delta of <5 ng/L ruled out 42 % of patients. The Mindray assay showed incremental (non-rounded) analysis discriminated rule out from baseline of <0.1 ng/L at 1.2 % to 2.0 ng/L at 17.0 %. There were no missed adverse outcomes at 30-day assessment for composite of MI and cardiac death.
Conclusions: The novel Mindray hs-cTnI assay enabled safe and early rule out of MI and cardiac death at very low concentrations in a diverse, cohort utilizing both single sample and 0/2-h rule out protocols, including early presenters.
期刊介绍:
Clinical Chemistry and Laboratory Medicine (CCLM) publishes articles on novel teaching and training methods applicable to laboratory medicine. CCLM welcomes contributions on the progress in fundamental and applied research and cutting-edge clinical laboratory medicine. It is one of the leading journals in the field, with an impact factor over 3. CCLM is issued monthly, and it is published in print and electronically.
CCLM is the official journal of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) and publishes regularly EFLM recommendations and news. CCLM is the official journal of the National Societies from Austria (ÖGLMKC); Belgium (RBSLM); Germany (DGKL); Hungary (MLDT); Ireland (ACBI); Italy (SIBioC); Portugal (SPML); and Slovenia (SZKK); and it is affiliated to AACB (Australia) and SFBC (France).
Topics:
- clinical biochemistry
- clinical genomics and molecular biology
- clinical haematology and coagulation
- clinical immunology and autoimmunity
- clinical microbiology
- drug monitoring and analysis
- evaluation of diagnostic biomarkers
- disease-oriented topics (cardiovascular disease, cancer diagnostics, diabetes)
- new reagents, instrumentation and technologies
- new methodologies
- reference materials and methods
- reference values and decision limits
- quality and safety in laboratory medicine
- translational laboratory medicine
- clinical metrology
Follow @cclm_degruyter on Twitter!