Differences in Pharmacokinetic and Histopathological Effects of Five Drug-Coated Balloons: An Experimental Study in Rabbit.

IF 1.7 2区 医学 Q3 PERIPHERAL VASCULAR DISEASE
Chenlin Zeng, Zhaoxi Peng, Xiangxiang Li, Qun Huang, Zhijue Xu, Junchao Liu, Zhaoyu Wu, Jiahao Lei, Hongji Pu, Weiqing Wei, Weimin Li, Jinbao Qin, Xinwu Lu
{"title":"Differences in Pharmacokinetic and Histopathological Effects of Five Drug-Coated Balloons: An Experimental Study in Rabbit.","authors":"Chenlin Zeng, Zhaoxi Peng, Xiangxiang Li, Qun Huang, Zhijue Xu, Junchao Liu, Zhaoyu Wu, Jiahao Lei, Hongji Pu, Weiqing Wei, Weimin Li, Jinbao Qin, Xinwu Lu","doi":"10.1177/15266028251326848","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The optimal design concept and the safety of drug-coated balloons (DCBs) have not been completely determined yet. Also, the optimal approach for DCB-based revascularization of peripheral artery diseases still remains undefined. This study was designed to explore the in vivo pharmacokinetic and histopathological effects of DCBs using Ranger and 4 Chinese DCBs after implantation and administration in New Zealand rabbits.</p><p><strong>Materials and methods: </strong>Fifty New Zealand rabbits were divided into 5 groups with 10 rabbits in each group according to the DCB used: Ranger (Boston Scientific), Orchid (Acotec), Reewarm (Endovastec), Ultrafree (Zylox), and Yaohang (Polyrey). After being guided to the lower segment of the abdominal aorta, the DCB was inflated for 3 minutes. Plasma, inflated infrarenal aorta, vastus lateralis muscle, anterior tibial muscle, and right toes were harvested for histological and paclitaxel concentration analyses 4 hours or 28 days after the angioplasty.</p><p><strong>Results: </strong>At 4 hours after aortic angioplasty, the overall paclitaxel concentrations in aortic wall were not statistically different (p = 0.050), but the paclitaxel concentrations in vastus lateralis muscle (p = 0.002), anterior tibial muscle (p = 0.006) and toe (p < 0.001) were not totally same according to the results of Kruskal-Wallis test. In toe, concentrations of paclitaxel were significantly lower for the Ranger (120.8 ng/g) DCB than for the Orchid (1880 ng/g; p = 0.008), Reewarm (347 ng/g; p = 0.016), and Ultrafree (261 ng/g; p = 0.016) DCBs. Concentrations of paclitaxel in the toe were not statistically different between Ranger and Yaohang DCBs (p = 0.421). Neointimal area (p < 0.001), neointimal thickness (p < 0.001), and percentage of luminal stenosis (p < 0.001) were less for Ranger DCB than for other DCBs 28 days after aortic angioplasty. The differences in paclitaxel concentrations in tissues 28 days after aortic angioplasty were not statistically significant.</p><p><strong>Conclusion: </strong>Different design concepts will make a difference in the in vivo pharmacokinetic and histopathological effects of DCBs. The Ranger DCB can achieve similar drug delivery efficiency as other higher-dose DCBs and fewer neointimal hyperplasia. Although clinical implications remain to be further investigated, the present results may provide implications for the design and use of DCBs.Clinical ImpactThe current preclinical study suggested that different design concepts would make a difference in the pharmacokinetic and histopathological effects of drug-coated balloons (DCBs). Though Ranger DCB had the lowest paclitaxel loading in this study, it still achieved similar drug delivery efficiency with other higher-dose DCBs. Also, neointimal hyperplasia was less for the Ranger DCB than for the other DCBs 28 days after aortic angioplasty. Although clinical implications remain to be further investigated, the present results may provide implications for the design and use of DCBs.</p>","PeriodicalId":50210,"journal":{"name":"Journal of Endovascular Therapy","volume":" ","pages":"15266028251326848"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Endovascular Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15266028251326848","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The optimal design concept and the safety of drug-coated balloons (DCBs) have not been completely determined yet. Also, the optimal approach for DCB-based revascularization of peripheral artery diseases still remains undefined. This study was designed to explore the in vivo pharmacokinetic and histopathological effects of DCBs using Ranger and 4 Chinese DCBs after implantation and administration in New Zealand rabbits.

Materials and methods: Fifty New Zealand rabbits were divided into 5 groups with 10 rabbits in each group according to the DCB used: Ranger (Boston Scientific), Orchid (Acotec), Reewarm (Endovastec), Ultrafree (Zylox), and Yaohang (Polyrey). After being guided to the lower segment of the abdominal aorta, the DCB was inflated for 3 minutes. Plasma, inflated infrarenal aorta, vastus lateralis muscle, anterior tibial muscle, and right toes were harvested for histological and paclitaxel concentration analyses 4 hours or 28 days after the angioplasty.

Results: At 4 hours after aortic angioplasty, the overall paclitaxel concentrations in aortic wall were not statistically different (p = 0.050), but the paclitaxel concentrations in vastus lateralis muscle (p = 0.002), anterior tibial muscle (p = 0.006) and toe (p < 0.001) were not totally same according to the results of Kruskal-Wallis test. In toe, concentrations of paclitaxel were significantly lower for the Ranger (120.8 ng/g) DCB than for the Orchid (1880 ng/g; p = 0.008), Reewarm (347 ng/g; p = 0.016), and Ultrafree (261 ng/g; p = 0.016) DCBs. Concentrations of paclitaxel in the toe were not statistically different between Ranger and Yaohang DCBs (p = 0.421). Neointimal area (p < 0.001), neointimal thickness (p < 0.001), and percentage of luminal stenosis (p < 0.001) were less for Ranger DCB than for other DCBs 28 days after aortic angioplasty. The differences in paclitaxel concentrations in tissues 28 days after aortic angioplasty were not statistically significant.

Conclusion: Different design concepts will make a difference in the in vivo pharmacokinetic and histopathological effects of DCBs. The Ranger DCB can achieve similar drug delivery efficiency as other higher-dose DCBs and fewer neointimal hyperplasia. Although clinical implications remain to be further investigated, the present results may provide implications for the design and use of DCBs.Clinical ImpactThe current preclinical study suggested that different design concepts would make a difference in the pharmacokinetic and histopathological effects of drug-coated balloons (DCBs). Though Ranger DCB had the lowest paclitaxel loading in this study, it still achieved similar drug delivery efficiency with other higher-dose DCBs. Also, neointimal hyperplasia was less for the Ranger DCB than for the other DCBs 28 days after aortic angioplasty. Although clinical implications remain to be further investigated, the present results may provide implications for the design and use of DCBs.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.30
自引率
15.40%
发文量
203
审稿时长
6-12 weeks
期刊介绍: The Journal of Endovascular Therapy (formerly the Journal of Endovascular Surgery) was established in 1994 as a forum for all physicians, scientists, and allied healthcare professionals who are engaged or interested in peripheral endovascular techniques and technology. An official publication of the International Society of Endovascular Specialists (ISEVS), the Journal of Endovascular Therapy publishes peer-reviewed articles of interest to clinicians and researchers in the field of peripheral endovascular interventions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信