Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device.

IF 0.9 Q4 REHABILITATION
Elise Yee Yan Li, Luke Steven Robinson, Claire Stockdale
{"title":"Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device.","authors":"Elise Yee Yan Li, Luke Steven Robinson, Claire Stockdale","doi":"10.1177/17589983251322537","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Evaluation of maximum grip strength (MGS) is commonly conducted to guide intervention planning and evaluate treatment outcomes for individuals with hand or upper limb impairments. While the hydraulic Jamar dynamometer is recognised as the gold standard measurement instrument, it has several limitations and there has not been a consensus on whether alternative instruments can replace or complement it. This study aimed to determine whether a new electronic rehabilitation device, the GripAble, is a reliable and valid instrument for measuring the MGS of healthy adults when compared to the gold standard hydraulic Jamar dynamometer. It also sought to compare the user's experience when using the two instruments.</p><p><strong>Methods: </strong>Fifty-five healthy adults completed two MGS assessments ten minutes apart using a standardised protocol. Test-retest and inter-instrument reliability were evaluated using intraclass correlation coefficient. Concurrent validity was established using a Bland-Altman plot. Relationship and level of agreement between measurements on the two devices were evaluated. Participants' user experience was explored through a quantitative survey.</p><p><strong>Results: </strong>The GripAble was found to have excellent test-retest reliability (ICC = 0.971-0.975) and good-to-excellent inter-instrument reliability (ICC = 0.898-0.922) with the Jamar dynamometer. Its measurement output is equivalent to approximately 63.3% of that of the Jamar dynamometer. The mean difference value between the measurements on the two devices was 9.44 kg (SD = 4.47). Users preferred the GripAble.</p><p><strong>Discussion: </strong>While the GripAble is a reliable MGS assessment instrument, it should not be used interchangeably with the Jamar dynamometer for repeated measurements of the same individual owing to the differences in their readings.</p>","PeriodicalId":43971,"journal":{"name":"Hand Therapy","volume":" ","pages":"17589983251322537"},"PeriodicalIF":0.9000,"publicationDate":"2025-03-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11909645/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hand Therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/17589983251322537","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"REHABILITATION","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Evaluation of maximum grip strength (MGS) is commonly conducted to guide intervention planning and evaluate treatment outcomes for individuals with hand or upper limb impairments. While the hydraulic Jamar dynamometer is recognised as the gold standard measurement instrument, it has several limitations and there has not been a consensus on whether alternative instruments can replace or complement it. This study aimed to determine whether a new electronic rehabilitation device, the GripAble, is a reliable and valid instrument for measuring the MGS of healthy adults when compared to the gold standard hydraulic Jamar dynamometer. It also sought to compare the user's experience when using the two instruments.

Methods: Fifty-five healthy adults completed two MGS assessments ten minutes apart using a standardised protocol. Test-retest and inter-instrument reliability were evaluated using intraclass correlation coefficient. Concurrent validity was established using a Bland-Altman plot. Relationship and level of agreement between measurements on the two devices were evaluated. Participants' user experience was explored through a quantitative survey.

Results: The GripAble was found to have excellent test-retest reliability (ICC = 0.971-0.975) and good-to-excellent inter-instrument reliability (ICC = 0.898-0.922) with the Jamar dynamometer. Its measurement output is equivalent to approximately 63.3% of that of the Jamar dynamometer. The mean difference value between the measurements on the two devices was 9.44 kg (SD = 4.47). Users preferred the GripAble.

Discussion: While the GripAble is a reliable MGS assessment instrument, it should not be used interchangeably with the Jamar dynamometer for repeated measurements of the same individual owing to the differences in their readings.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Hand Therapy
Hand Therapy REHABILITATION-
CiteScore
1.60
自引率
10.00%
发文量
13
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信