Validation of an In Situ Self-Administered Hearing Test with a Self-Fitting Device for Mild to Moderately Severe Hearing Loss in Various Acoustic Environments.
{"title":"Validation of an In Situ Self-Administered Hearing Test with a Self-Fitting Device for Mild to Moderately Severe Hearing Loss in Various Acoustic Environments.","authors":"Tun-Shin Lo, Hung-Yue Chang","doi":"10.1177/00034894251325617","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>This study compared a simplified in situ self-administered hearing screening test, conducted with a neckband-type self-fitting device, with conventional pure-tone audiometry. It evaluated the maximum speech-shaped noise level for screening (MSNLS), crucial for evaluating the feasibility of this in situ screening test in quiet environments.</p><p><strong>Methods: </strong>This study included 30 adults with normal hearing and 30 adults with mild to moderately severe hearing impairment. A binaural neckband-type self-fitting device was developed. The results of an in situ hearing screening test conducted using the self-fitting device were compared with those obtained using traditional pure-tone audiometry conducted using TDH-50 earphones. Subsequently, MSNLS was determined by assessing noise-masking effects on screening outcomes. All tests were conducted in an audiometric booth, with the hearing screening test conducted in the booth with the door open.</p><p><strong>Results: </strong>Strong positive correlations were observed between the results of pure-tone audiometry and those of hearing screening tests across all test frequencies, with the strongest correlation observed at 2000 Hz (<i>r<sub>s</sub></i> = 0.793, <i>P</i> < .001) and the weakest correlation observed at 500 Hz (<i>r<sub>s</sub></i> = 0.625, <i>P</i> < .001). Comparisons of screening tests results with pure-tone thresholds across all test frequencies revealed differences of approximately 10 dB HL for 80% of all ears. The sensitivity and specificity of the hearing screening test in detecting candidates with hearing loss (>30 dB HL) who are suitable for this device were 93% and 90%, respectively. The hearing-impaired group exhibited MSNLSs, such as 57 dB SPL at 500 Hz, exceeding ambient noise levels in an empty classroom.</p><p><strong>Conclusion: </strong>The in situ hearing screening test, conducted using a self-fitting device, exhibited reasonable accuracy for self-fitting scenarios in general quiet environments. This test can be used for monitoring mild to moderate hearing loss or fluctuating hearing loss, such as that associated with Ménière's disease.</p>","PeriodicalId":50975,"journal":{"name":"Annals of Otology Rhinology and Laryngology","volume":" ","pages":"34894251325617"},"PeriodicalIF":1.3000,"publicationDate":"2025-03-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Otology Rhinology and Laryngology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00034894251325617","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OTORHINOLARYNGOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: This study compared a simplified in situ self-administered hearing screening test, conducted with a neckband-type self-fitting device, with conventional pure-tone audiometry. It evaluated the maximum speech-shaped noise level for screening (MSNLS), crucial for evaluating the feasibility of this in situ screening test in quiet environments.
Methods: This study included 30 adults with normal hearing and 30 adults with mild to moderately severe hearing impairment. A binaural neckband-type self-fitting device was developed. The results of an in situ hearing screening test conducted using the self-fitting device were compared with those obtained using traditional pure-tone audiometry conducted using TDH-50 earphones. Subsequently, MSNLS was determined by assessing noise-masking effects on screening outcomes. All tests were conducted in an audiometric booth, with the hearing screening test conducted in the booth with the door open.
Results: Strong positive correlations were observed between the results of pure-tone audiometry and those of hearing screening tests across all test frequencies, with the strongest correlation observed at 2000 Hz (rs = 0.793, P < .001) and the weakest correlation observed at 500 Hz (rs = 0.625, P < .001). Comparisons of screening tests results with pure-tone thresholds across all test frequencies revealed differences of approximately 10 dB HL for 80% of all ears. The sensitivity and specificity of the hearing screening test in detecting candidates with hearing loss (>30 dB HL) who are suitable for this device were 93% and 90%, respectively. The hearing-impaired group exhibited MSNLSs, such as 57 dB SPL at 500 Hz, exceeding ambient noise levels in an empty classroom.
Conclusion: The in situ hearing screening test, conducted using a self-fitting device, exhibited reasonable accuracy for self-fitting scenarios in general quiet environments. This test can be used for monitoring mild to moderate hearing loss or fluctuating hearing loss, such as that associated with Ménière's disease.
期刊介绍:
The Annals of Otology, Rhinology & Laryngology publishes original manuscripts of clinical and research importance in otolaryngology–head and neck medicine and surgery, otology, neurotology, bronchoesophagology, laryngology, rhinology, head and neck oncology and surgery, plastic and reconstructive surgery, pediatric otolaryngology, audiology, and speech pathology. In-depth studies (supplements), papers of historical interest, and reviews of computer software and applications in otolaryngology are also published, as well as imaging, pathology, and clinicopathology studies, book reviews, and letters to the editor. AOR is the official journal of the American Broncho-Esophagological Association.