Fully Threaded Versus Partially Threaded Intramedullary Headless Compression Screw Fixation for Proximal Third Fractures of the Proximal Phalanx: A Biomechanical Study.
Bert Vanmierlo, Bruno Vandekerckhove, Leen Popleu, Bert O Eijnde, Joris Duerinckx, Jean F Goubau
{"title":"Fully Threaded Versus Partially Threaded Intramedullary Headless Compression Screw Fixation for Proximal Third Fractures of the Proximal Phalanx: A Biomechanical Study.","authors":"Bert Vanmierlo, Bruno Vandekerckhove, Leen Popleu, Bert O Eijnde, Joris Duerinckx, Jean F Goubau","doi":"10.1016/j.jhsa.2025.01.006","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The purpose of this study was to compare the maximal elastic force and deformation resistance of two commonly used headless compression screw systems, a fully threaded 2.5 mm screw and a partially threaded screw of the same dimensions, in a cadaveric model of a human transverse proximal third fracture of the proximal phalanx.</p><p><strong>Methods: </strong>Six fresh frozen cadavers were used, and the proximal phalanges of the index, long, and ring fingers of both hands were dissected. A proximal third transverse osteotomy was created, and screws were inserted in an antegrade direction. The three-point bending load-to-failure test measured the maximal elastic force and deformation resistance of each bone-screw construct.</p><p><strong>Results: </strong>The fully threaded screw demonstrated a maximal elastic force of 121 N and a deformation resistance of 134 N/mm. The partially threaded screw showed superior performance with a maximal elastic force of 165 N and deformation resistance of 166 N/mm. The bone-screw construct with the fully threaded screw had a mean maximal elastic force of 155.2 N compared to 101.8 N in the construct with the partially threaded screw. Deformation resistance was 145.2 N/mm for fully threaded screws and 98.6 N/mm for partially threaded screws.</p><p><strong>Conclusions: </strong>Fully threaded intramedullary headless compression screw fixation provides superior maximal elastic force and deformation resistance compared to partially threaded intramedullary headless compression screw fixation in transverse proximal third fractures of the proximal phalanx.</p><p><strong>Clinical relevance: </strong>The findings indicate that fully threaded intramedullary headless compression screw fixation may offer enhanced stability for treating proximal third fractures of the proximal phalanx, potentially leading to better patient outcomes through earlier mobilization and reduced complications compared to partially threaded screws.</p>","PeriodicalId":54815,"journal":{"name":"Journal of Hand Surgery-American Volume","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2025-03-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hand Surgery-American Volume","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jhsa.2025.01.006","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The purpose of this study was to compare the maximal elastic force and deformation resistance of two commonly used headless compression screw systems, a fully threaded 2.5 mm screw and a partially threaded screw of the same dimensions, in a cadaveric model of a human transverse proximal third fracture of the proximal phalanx.
Methods: Six fresh frozen cadavers were used, and the proximal phalanges of the index, long, and ring fingers of both hands were dissected. A proximal third transverse osteotomy was created, and screws were inserted in an antegrade direction. The three-point bending load-to-failure test measured the maximal elastic force and deformation resistance of each bone-screw construct.
Results: The fully threaded screw demonstrated a maximal elastic force of 121 N and a deformation resistance of 134 N/mm. The partially threaded screw showed superior performance with a maximal elastic force of 165 N and deformation resistance of 166 N/mm. The bone-screw construct with the fully threaded screw had a mean maximal elastic force of 155.2 N compared to 101.8 N in the construct with the partially threaded screw. Deformation resistance was 145.2 N/mm for fully threaded screws and 98.6 N/mm for partially threaded screws.
Conclusions: Fully threaded intramedullary headless compression screw fixation provides superior maximal elastic force and deformation resistance compared to partially threaded intramedullary headless compression screw fixation in transverse proximal third fractures of the proximal phalanx.
Clinical relevance: The findings indicate that fully threaded intramedullary headless compression screw fixation may offer enhanced stability for treating proximal third fractures of the proximal phalanx, potentially leading to better patient outcomes through earlier mobilization and reduced complications compared to partially threaded screws.
期刊介绍:
The Journal of Hand Surgery publishes original, peer-reviewed articles related to the pathophysiology, diagnosis, and treatment of diseases and conditions of the upper extremity; these include both clinical and basic science studies, along with case reports. Special features include Review Articles (including Current Concepts and The Hand Surgery Landscape), Reviews of Books and Media, and Letters to the Editor.