{"title":"A multicenter study comparing a point-of-care blood glucose metre with a blood gas analyser in infants by the Shikoku Neonatal Medical Research Group.","authors":"Toru Kuboi, Masashiro Sugino, Takaaki Sadamura, Nana Kawaguchi, Yoko Tadatomo, Kosei Takada, Natsumi Okamoto, Kiwako Miyamoto, Akiko Nakano, Noriko Miura, Yusei Nakata, Kenichi Suga, Masaaki Ota, Shinji Nakamura, Kosuke Koyano, Takashi Kusaka","doi":"10.1177/00045632251326460","DOIUrl":null,"url":null,"abstract":"<p><p>BackgroundThere have been no reports on blood glucose metres for measurements in a wide variety of infant patients, from very preterm infants to mature infants and from the early neonatal period onwards. In this study, we evaluated the accuracy of the Glutest Mint II, a point-of-care blood glucose metre, by comparing blood glucose levels measured by this device with those measured by a blood gas analyser in infants of all gestational ages and birth weights at a number of time points after birth.MethodsInfants aged 22 weeks and 0 days of gestation or older who were born at any of six tertiary neonatal intensive care units between March 2022 and January 2023 and needed blood glucose monitoring were enrolled. Samples were collected into capillary tubes when the physician determined that a blood glucose test was necessary and could be taken at any time after birth and any number of times from the same individual. Blood glucose was measured using a Glutest Mint II and then using a blood gas analyser.ResultsIn total, 2943 blood glucose points were measured in 285 infants. Blood glucose levels measured using the Glutest Mint II were significantly correlated with those measured using a blood gas analyser. Neonatal-specific parameters such as hematocrit and total serum bilirubin levels may not have an effect.ConclusionsThe Glutest Mint II device can measure blood glucose levels with very high accuracy in the range used in the neonatal setting, comparable to the blood gas analyser.</p>","PeriodicalId":8005,"journal":{"name":"Annals of Clinical Biochemistry","volume":" ","pages":"45632251326460"},"PeriodicalIF":2.1000,"publicationDate":"2025-02-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Clinical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00045632251326460","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
BackgroundThere have been no reports on blood glucose metres for measurements in a wide variety of infant patients, from very preterm infants to mature infants and from the early neonatal period onwards. In this study, we evaluated the accuracy of the Glutest Mint II, a point-of-care blood glucose metre, by comparing blood glucose levels measured by this device with those measured by a blood gas analyser in infants of all gestational ages and birth weights at a number of time points after birth.MethodsInfants aged 22 weeks and 0 days of gestation or older who were born at any of six tertiary neonatal intensive care units between March 2022 and January 2023 and needed blood glucose monitoring were enrolled. Samples were collected into capillary tubes when the physician determined that a blood glucose test was necessary and could be taken at any time after birth and any number of times from the same individual. Blood glucose was measured using a Glutest Mint II and then using a blood gas analyser.ResultsIn total, 2943 blood glucose points were measured in 285 infants. Blood glucose levels measured using the Glutest Mint II were significantly correlated with those measured using a blood gas analyser. Neonatal-specific parameters such as hematocrit and total serum bilirubin levels may not have an effect.ConclusionsThe Glutest Mint II device can measure blood glucose levels with very high accuracy in the range used in the neonatal setting, comparable to the blood gas analyser.
背景:目前还没有关于血糖仪用于各种婴儿患者测量的报道,从非常早产的婴儿到成熟的婴儿以及从新生儿早期开始。在这项研究中,我们评估了Glutest Mint II的准确性,这是一种即时护理血糖仪,通过比较该设备测量的血糖水平与血气分析仪在出生后多个时间点测量的所有胎龄和出生体重的婴儿的血糖水平。方法:纳入2022年3月至2023年1月期间在6个三级新生儿重症监护病房中出生并需要血糖监测的22周零0天或以上的婴儿。当医生确定有必要进行血糖测试时,将样本收集到毛细血管中,并且可以在出生后的任何时间和同一个体的任何次数进行血糖测试。血糖测量使用Glutest Mint II,然后使用血气分析仪。结果:285例婴儿共测得2943个血糖点。使用Glutest Mint II测量的血糖水平与使用血气分析仪测量的血糖水平显著相关。新生儿特异性参数,如红细胞压积和血清总胆红素水平可能没有影响。结论:Glutest Mint II设备可以在新生儿环境中以非常高的准确度测量血糖水平,与血气分析仪相当。
期刊介绍:
Annals of Clinical Biochemistry is the fully peer reviewed international journal of the Association for Clinical Biochemistry and Laboratory Medicine.
Annals of Clinical Biochemistry accepts papers that contribute to knowledge in all fields of laboratory medicine, especially those pertaining to the understanding, diagnosis and treatment of human disease. It publishes papers on clinical biochemistry, clinical audit, metabolic medicine, immunology, genetics, biotechnology, haematology, microbiology, computing and management where they have both biochemical and clinical relevance. Papers describing evaluation or implementation of commercial reagent kits or the performance of new analysers require substantial original information. Unless of exceptional interest and novelty, studies dealing with the redox status in various diseases are not generally considered within the journal''s scope. Studies documenting the association of single nucleotide polymorphisms (SNPs) with particular phenotypes will not normally be considered, given the greater strength of genome wide association studies (GWAS). Research undertaken in non-human animals will not be considered for publication in the Annals.
Annals of Clinical Biochemistry is also the official journal of NVKC (de Nederlandse Vereniging voor Klinische Chemie) and JSCC (Japan Society of Clinical Chemistry).