[Design and validation of an automated testing system for essential performance parameters of ventilators].

Q4 Medicine
Yongzhen Li, Wei Wang, Chunyuan Zhang, Xia Zhang, Zhenglong Chen, Zhaoyan Hu
{"title":"[Design and validation of an automated testing system for essential performance parameters of ventilators].","authors":"Yongzhen Li, Wei Wang, Chunyuan Zhang, Xia Zhang, Zhenglong Chen, Zhaoyan Hu","doi":"10.7507/1001-5515.202408009","DOIUrl":null,"url":null,"abstract":"<p><p>Traditional manual testing of ventilator performance is labor-intensive, time-consuming, and prone to errors in data recording, making it difficult to meet the current demands for testing efficiency in the development and manufacturing of ventilators. Therefore, in this study we designed an automated testing system for essential performance parameters of ventilators. The system mainly comprises a ventilator airflow analyzer, an automated switch module for simulated lungs, and a test control platform. Under the control of testing software, this system can perform automated tests of critical performance parameters of ventilators and generate a final test report. To validate the effectiveness of the designed system, tests were conducted on two different brands of ventilators under four different operating conditions, comparing tidal volume, oxygen concentration, and positive end expiratory pressure accuracy using both the automated testing system and traditional manual methods. Bland-Altman statistical analysis indicated good consistency between the accuracy of automated tests and manual tests for all respiratory parameters. In terms of testing efficiency, the automated testing system required approximately one-third of the time needed for manual testing. These results demonstrate that the designed automated testing system provides a novel approach and means for quality inspection and measurement calibration of ventilators, showing broad application prospects.</p>","PeriodicalId":39324,"journal":{"name":"生物医学工程学杂志","volume":"42 1","pages":"164-173"},"PeriodicalIF":0.0000,"publicationDate":"2025-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11955326/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"生物医学工程学杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.7507/1001-5515.202408009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Traditional manual testing of ventilator performance is labor-intensive, time-consuming, and prone to errors in data recording, making it difficult to meet the current demands for testing efficiency in the development and manufacturing of ventilators. Therefore, in this study we designed an automated testing system for essential performance parameters of ventilators. The system mainly comprises a ventilator airflow analyzer, an automated switch module for simulated lungs, and a test control platform. Under the control of testing software, this system can perform automated tests of critical performance parameters of ventilators and generate a final test report. To validate the effectiveness of the designed system, tests were conducted on two different brands of ventilators under four different operating conditions, comparing tidal volume, oxygen concentration, and positive end expiratory pressure accuracy using both the automated testing system and traditional manual methods. Bland-Altman statistical analysis indicated good consistency between the accuracy of automated tests and manual tests for all respiratory parameters. In terms of testing efficiency, the automated testing system required approximately one-third of the time needed for manual testing. These results demonstrate that the designed automated testing system provides a novel approach and means for quality inspection and measurement calibration of ventilators, showing broad application prospects.

求助全文
约1分钟内获得全文 求助全文
来源期刊
生物医学工程学杂志
生物医学工程学杂志 Medicine-Medicine (all)
CiteScore
0.80
自引率
0.00%
发文量
4868
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信