{"title":"Computed Tomography Doses Calculation: Do We Really Need a New Dose Assessment Tool?","authors":"Arkadiusz Szarmach, Dominika Sabiniewicz-Ziajka, Małgorzata Grzywińska, Paweł Gać, Maciej Piskunowicz, Magdalena Wszędybył-Winklewska","doi":"10.3390/jcm14041348","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background/Objectives:</b> The increasing use of computed tomography (CT) scans significantly contributes to population exposure to ionizing radiation. Traditional dose metrics, such as dose-length product (DLP) and effective dose (ED), lack precision in reflecting individual radiation exposure. This study introduces a novel parameters such as size-specific effective dose (EDss) and the size-specific dose-length product (DLPss), to improve patient-specific dose estimation. The aim of this study is to enhance dose calculation accuracy, optimize CT protocols, and guide the development of next-generation CT technologies. <b>Methods:</b> A retrospective analysis of 247 abdominal and pelvic CT scans (113 women, 134 men) was conducted. Anthropometric parameters, including body mass index (BMI), cross-sectional dimensions, and dose indices, were measured. EDss and DLPss were calculated using size-specific correction factors, and statistical correlations between these parameters were assessed. <b>Results:</b> The mean BMI was 25.92 ± 5.34. DLPss values ranged from 261.63 to 1217.70 mGy·cm (mean: 627.83 ± 145.32) and were roughly 21% higher than traditional DLP values, with men showing slightly higher mean values than women. EDss values ranged from 6.65 to 15.45 mSv (mean: 9.42 ± 2.18 mSv), approximately 22% higher than traditional ED values, demonstrating improved individualization. Significant correlations were observed between BMI and effective diameter (r = 0.78), with stronger correlations in men (r = 0.85). The mean CTDIvol was 11.37 ± 3.50 mGy, and SSDE averaged 13.91 ± 2.39 mGy. Scan length reductions were observed in 53.8% of cases, with statistically significant differences by gender. <b>Conclusions:</b> EDss and DLPss offer improved accuracy in radiation dose estimation, addressing the limitations of traditional methods. Their adoption into clinical protocols, supported by AI-driven automation, could optimize diagnostic safety and significantly reduce radiation risk for patients. Further multicenter studies and technological advancements are recommended to validate these metrics and facilitate their integration into daily practice.</p>","PeriodicalId":15533,"journal":{"name":"Journal of Clinical Medicine","volume":"14 4","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-02-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/jcm14041348","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background/Objectives: The increasing use of computed tomography (CT) scans significantly contributes to population exposure to ionizing radiation. Traditional dose metrics, such as dose-length product (DLP) and effective dose (ED), lack precision in reflecting individual radiation exposure. This study introduces a novel parameters such as size-specific effective dose (EDss) and the size-specific dose-length product (DLPss), to improve patient-specific dose estimation. The aim of this study is to enhance dose calculation accuracy, optimize CT protocols, and guide the development of next-generation CT technologies. Methods: A retrospective analysis of 247 abdominal and pelvic CT scans (113 women, 134 men) was conducted. Anthropometric parameters, including body mass index (BMI), cross-sectional dimensions, and dose indices, were measured. EDss and DLPss were calculated using size-specific correction factors, and statistical correlations between these parameters were assessed. Results: The mean BMI was 25.92 ± 5.34. DLPss values ranged from 261.63 to 1217.70 mGy·cm (mean: 627.83 ± 145.32) and were roughly 21% higher than traditional DLP values, with men showing slightly higher mean values than women. EDss values ranged from 6.65 to 15.45 mSv (mean: 9.42 ± 2.18 mSv), approximately 22% higher than traditional ED values, demonstrating improved individualization. Significant correlations were observed between BMI and effective diameter (r = 0.78), with stronger correlations in men (r = 0.85). The mean CTDIvol was 11.37 ± 3.50 mGy, and SSDE averaged 13.91 ± 2.39 mGy. Scan length reductions were observed in 53.8% of cases, with statistically significant differences by gender. Conclusions: EDss and DLPss offer improved accuracy in radiation dose estimation, addressing the limitations of traditional methods. Their adoption into clinical protocols, supported by AI-driven automation, could optimize diagnostic safety and significantly reduce radiation risk for patients. Further multicenter studies and technological advancements are recommended to validate these metrics and facilitate their integration into daily practice.
期刊介绍:
Journal of Clinical Medicine (ISSN 2077-0383), is an international scientific open access journal, providing a platform for advances in health care/clinical practices, the study of direct observation of patients and general medical research. This multi-disciplinary journal is aimed at a wide audience of medical researchers and healthcare professionals.
Unique features of this journal:
manuscripts regarding original research and ideas will be particularly welcomed.JCM also accepts reviews, communications, and short notes.
There is no limit to publication length: our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible.