Clinical Application of Volumetric Absorptive Microsampling for Therapeutic Drug Monitoring of Oral Targeted Anticancer Drugs.

IF 2.8 4区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY
Marinda Meertens, Nikki Kerssemakers, Niels de Vries, Hilde Rosing, Neeltje Steeghs, Jos H Beijnen, Alwin D R Huitema
{"title":"Clinical Application of Volumetric Absorptive Microsampling for Therapeutic Drug Monitoring of Oral Targeted Anticancer Drugs.","authors":"Marinda Meertens, Nikki Kerssemakers, Niels de Vries, Hilde Rosing, Neeltje Steeghs, Jos H Beijnen, Alwin D R Huitema","doi":"10.1097/FTD.0000000000001315","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Therapeutic Drug Monitoring optimizes oral anticancer drug treatment by measuring plasma levels. Volumetric absorptive microsampling (VAMS) allows home sampling with a minimal blood sample. However, methods for converting whole blood into plasma are required to interpret these results. This study aimed to establish conversion methods for abiraterone, alectinib, cabozantinib, imatinib, olaparib, sunitinib, and their metabolites, while assessing the differences between venous and capillary blood. The feasibility of home sampling was also evaluated.</p><p><strong>Methods: </strong>Plasma and VAMS samples, both from venipuncture-collected whole blood tubes and from a finger prick, were collected from each patient. The VAMS samples were deemed comparable if their concentrations were within ±20% of each other for ≥2/3rd of the patients. The Passing-Bablok regression and conversion factor methods were tested for the plasma and VAMS finger prick samples. The estimated plasma concentrations using both methods were required to be within ±20% of the measured plasma concentrations for ≥2/3rd of the pairs.</p><p><strong>Results: </strong>Overall, 153 patients were enrolled in this study. Conversion methods were applied to the VAMS samples, and the acceptance criteria were met for alectinib-M4, cabozantinib, imatinib, N-desmethyl imatinib, olaparib, sunitinib, and N-desethyl sunitinib but not for abiraterone, D4A, or alectinib. The capillary and venous VAMS concentrations were similar, except for that of D4A. Patients were positive toward home sampling.</p><p><strong>Conclusions: </strong>The established VAMS conversion methods for 7 out of 10 oral targeted anticancer drugs or metabolites met the acceptance criteria. Future studies need to validate the conversion methods with an independent cohort and integrate home sampling via VAMS to provide patients with an alternative to venipuncture at the outpatient clinic.</p>","PeriodicalId":23052,"journal":{"name":"Therapeutic Drug Monitoring","volume":" ","pages":""},"PeriodicalIF":2.8000,"publicationDate":"2025-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Therapeutic Drug Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/FTD.0000000000001315","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Therapeutic Drug Monitoring optimizes oral anticancer drug treatment by measuring plasma levels. Volumetric absorptive microsampling (VAMS) allows home sampling with a minimal blood sample. However, methods for converting whole blood into plasma are required to interpret these results. This study aimed to establish conversion methods for abiraterone, alectinib, cabozantinib, imatinib, olaparib, sunitinib, and their metabolites, while assessing the differences between venous and capillary blood. The feasibility of home sampling was also evaluated.

Methods: Plasma and VAMS samples, both from venipuncture-collected whole blood tubes and from a finger prick, were collected from each patient. The VAMS samples were deemed comparable if their concentrations were within ±20% of each other for ≥2/3rd of the patients. The Passing-Bablok regression and conversion factor methods were tested for the plasma and VAMS finger prick samples. The estimated plasma concentrations using both methods were required to be within ±20% of the measured plasma concentrations for ≥2/3rd of the pairs.

Results: Overall, 153 patients were enrolled in this study. Conversion methods were applied to the VAMS samples, and the acceptance criteria were met for alectinib-M4, cabozantinib, imatinib, N-desmethyl imatinib, olaparib, sunitinib, and N-desethyl sunitinib but not for abiraterone, D4A, or alectinib. The capillary and venous VAMS concentrations were similar, except for that of D4A. Patients were positive toward home sampling.

Conclusions: The established VAMS conversion methods for 7 out of 10 oral targeted anticancer drugs or metabolites met the acceptance criteria. Future studies need to validate the conversion methods with an independent cohort and integrate home sampling via VAMS to provide patients with an alternative to venipuncture at the outpatient clinic.

用于口服靶向抗癌药物治疗药物监测的体积吸收微采样的临床应用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Therapeutic Drug Monitoring
Therapeutic Drug Monitoring 医学-毒理学
CiteScore
5.00
自引率
8.00%
发文量
213
审稿时长
4-8 weeks
期刊介绍: Therapeutic Drug Monitoring is a peer-reviewed, multidisciplinary journal directed to an audience of pharmacologists, clinical chemists, laboratorians, pharmacists, drug researchers and toxicologists. It fosters the exchange of knowledge among the various disciplines–clinical pharmacology, pathology, toxicology, analytical chemistry–that share a common interest in Therapeutic Drug Monitoring. The journal presents studies detailing the various factors that affect the rate and extent drugs are absorbed, metabolized, and excreted. Regular features include review articles on specific classes of drugs, original articles, case reports, technical notes, and continuing education articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信