The Effect of Repeated Closing Torque on Torque Loss and Angular Deviation: An In Vitro Study.

Ruoyan Guo, Yantao Qian, Guohai Lin, Run Chen
{"title":"The Effect of Repeated Closing Torque on Torque Loss and Angular Deviation: An In Vitro Study.","authors":"Ruoyan Guo, Yantao Qian, Guohai Lin, Run Chen","doi":"10.1563/aaid-joi-D-24-00143","DOIUrl":null,"url":null,"abstract":"<p><p>Loosening and fracturing of the central screw are common mechanical complications after implant restoration. However, the relationship between these complications and the repetitive tightening and loosening of the central screw during the fabrication and maintenance of the implant-supported restorations remains unknown. The purpose of this study was to evaluate the torque loss after repetitive tightening and loosening of the central screws on implants with different diameters as well as the changes in the angle deviation of the central screw relative to the implant. Twenty implants were divided into 2 groups based on diameter: 3.7 mm (group A) and 4.5 mm (group B) with 10 implants in each group. Each group was subdivided into 4 subgroups: A1, A2, B1, and B2 (n = 5). A closing torque of 15 N.cm was applied to groups A1 and B1, whereas a closing torque of 35 N.cm was applied to groups A2 and B2. Reverse torque measurements were taken 10 times for each group. The angular deviation of the central screw relative to the implant was recorded, and the surface wear of the central screw was observed under a scanning electron microscope. The data were analyzed using repeated measures 2-way analysis of variance (α = 0.05). Torque loss showed a significant upward trend across all groups with increased tightening cycle (P < .05). Implant diameter significantly influenced torque loss with smaller diameters exhibiting greater torque loss (P < .05). In addition, the angular deviation of the central screw relative to the implant was not affected by different diameters (P > .05). Still, it was affected by the closing torque and the cycles of multiple tightening and loosening procedures (P < .05). Under a 35 N.cm closing torque, initial torque loss ranged from 9.12 N.cm to 10.98 N.cm. Peak torque loss occurred at the 10th cycle with 16.40 N.cm values for 3.7-mm implants and 12.42 N.cm for 4.5-mm implants. Repeated tightening and loosening procedures increased both torque loss and angular deviation. The diameter of the implant may impact the torque loss with a larger diameter showing less torque loss. To reduce the risk of potential complications, it is suggested that the number of tightening cycles for narrow-diameter implants be limited.</p>","PeriodicalId":519890,"journal":{"name":"The Journal of oral implantology","volume":" ","pages":"202-209"},"PeriodicalIF":0.0000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of oral implantology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1563/aaid-joi-D-24-00143","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Loosening and fracturing of the central screw are common mechanical complications after implant restoration. However, the relationship between these complications and the repetitive tightening and loosening of the central screw during the fabrication and maintenance of the implant-supported restorations remains unknown. The purpose of this study was to evaluate the torque loss after repetitive tightening and loosening of the central screws on implants with different diameters as well as the changes in the angle deviation of the central screw relative to the implant. Twenty implants were divided into 2 groups based on diameter: 3.7 mm (group A) and 4.5 mm (group B) with 10 implants in each group. Each group was subdivided into 4 subgroups: A1, A2, B1, and B2 (n = 5). A closing torque of 15 N.cm was applied to groups A1 and B1, whereas a closing torque of 35 N.cm was applied to groups A2 and B2. Reverse torque measurements were taken 10 times for each group. The angular deviation of the central screw relative to the implant was recorded, and the surface wear of the central screw was observed under a scanning electron microscope. The data were analyzed using repeated measures 2-way analysis of variance (α = 0.05). Torque loss showed a significant upward trend across all groups with increased tightening cycle (P < .05). Implant diameter significantly influenced torque loss with smaller diameters exhibiting greater torque loss (P < .05). In addition, the angular deviation of the central screw relative to the implant was not affected by different diameters (P > .05). Still, it was affected by the closing torque and the cycles of multiple tightening and loosening procedures (P < .05). Under a 35 N.cm closing torque, initial torque loss ranged from 9.12 N.cm to 10.98 N.cm. Peak torque loss occurred at the 10th cycle with 16.40 N.cm values for 3.7-mm implants and 12.42 N.cm for 4.5-mm implants. Repeated tightening and loosening procedures increased both torque loss and angular deviation. The diameter of the implant may impact the torque loss with a larger diameter showing less torque loss. To reduce the risk of potential complications, it is suggested that the number of tightening cycles for narrow-diameter implants be limited.

重复关闭转矩对转矩损失和角度偏差的影响:体外研究。
信息。中心螺钉松动和断裂是种植体修复后常见的机械并发症。然而,在种植体支撑修复体的制作和维护过程中,这些并发症与中心螺钉的反复拧紧和松动之间的关系尚不清楚。目的:评价不同直径种植体中心螺钉反复拧紧和松动后的扭矩损失,以及中心螺钉相对种植体角度偏差的变化。材料与方法:20个种植体按直径分为2组:3.7 mm(A组)和4.5 mm(B组),每组10个种植体。每组再细分为A1、A2、B1、B2 4个亚组(n=5)。A1组和B1组的关闭扭矩为15 N·cm, A2组和B2组的关闭扭矩为35 N·cm。每组进行10次反向扭矩测量。记录中心螺钉相对于种植体的角度偏差,扫描电镜下观察中心螺钉的表面磨损情况。资料分析采用重复测量双方差分析(α=0.05)。结果:扭矩损失随拧紧周期的增加呈显著上升趋势(P0.05),但受闭合扭矩和多次拧紧和松动次数的影响(P0.05)。结论:反复拧紧和松动均增加扭矩损失和角度偏差。植入物的直径可能影响转矩损失,直径越大转矩损失越小。建议限制窄径种植体的拧紧次数,以减少潜在并发症的风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信