Presence of Ineffective Cardiac Resynchronization Therapy Pacing Provides Insights Into Hidden Causes and Therapeutic Targets of Nonresponder.

IF 2.3 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Satoshi Oka, Nobuhiko Ueda, Mitsuru Wada, Kohei Ishibashi, Toshihiro Nakamura, Yuichiro Miyazaki, Akinori Wakamiya, Hideaki Kanzaki, Chisato Izumi, Kengo Kusano
{"title":"Presence of Ineffective Cardiac Resynchronization Therapy Pacing Provides Insights Into Hidden Causes and Therapeutic Targets of Nonresponder.","authors":"Satoshi Oka, Nobuhiko Ueda, Mitsuru Wada, Kohei Ishibashi, Toshihiro Nakamura, Yuichiro Miyazaki, Akinori Wakamiya, Hideaki Kanzaki, Chisato Izumi, Kengo Kusano","doi":"10.1111/jce.16605","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>An effective cardiac resynchronization therapy (CRT) algorithm classifies each left ventricular (LV) pace as effective or ineffective based on electrogram morphology. Loss of capture, latency, pseudo-fusion during sinus rhythm, and rapid ventricular responses due to atrial fibrillation (AF) and premature ventricular conductions (PVCs) are potential causes of ineffective CRT pacing. However, little is known about the relationship between percentage of ineffective CRT pacing (%i-CRT) and CRT response.</p><p><strong>Methods and results: </strong>Among 136 consecutive CRT recipients, we evaluated 44 patients with an effective CRT algorithm. The primary outcome was the prevalence of CRT responders, defined as a reduction in the LV end-systolic volume by ≥ 10% 6 months after implantation. The median values of percent ventricular pacing and %i-CRT were 97.5% (interquartile range [IQR]: 96.4%-98.3%) and 0.3% (IQR: 0.1%-1.0%), respectively. Patients in the highest quartile of %i-CRT (%i-CRT ≥ 1%) showed a significantly lower prevalence of CRT responders (4 [36.4%] vs. 28 [84.8%], p = 0.004) than the remaining quartiles. The univariate analysis showed that %i-CRT ≥ 1% was negatively associated with the CRT response (odds ratio: 0.10, 95% confidence interval: 0.02-0.48). The most frequent cause of ineffective pacing was rapid ventricular responses due to AF and/or PVCs, whereas the cause of the highest %i-CRT in this cohort was latency. A total of three nonresponders with a high %i-CRT showed favorable responses after therapeutic interventions for their hidden causes.</p><p><strong>Conclusion: </strong>High %i-CRT could be associated with an unfavorable CRT response and provide insights into hidden causes and therapeutic targets.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-02-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16605","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: An effective cardiac resynchronization therapy (CRT) algorithm classifies each left ventricular (LV) pace as effective or ineffective based on electrogram morphology. Loss of capture, latency, pseudo-fusion during sinus rhythm, and rapid ventricular responses due to atrial fibrillation (AF) and premature ventricular conductions (PVCs) are potential causes of ineffective CRT pacing. However, little is known about the relationship between percentage of ineffective CRT pacing (%i-CRT) and CRT response.

Methods and results: Among 136 consecutive CRT recipients, we evaluated 44 patients with an effective CRT algorithm. The primary outcome was the prevalence of CRT responders, defined as a reduction in the LV end-systolic volume by ≥ 10% 6 months after implantation. The median values of percent ventricular pacing and %i-CRT were 97.5% (interquartile range [IQR]: 96.4%-98.3%) and 0.3% (IQR: 0.1%-1.0%), respectively. Patients in the highest quartile of %i-CRT (%i-CRT ≥ 1%) showed a significantly lower prevalence of CRT responders (4 [36.4%] vs. 28 [84.8%], p = 0.004) than the remaining quartiles. The univariate analysis showed that %i-CRT ≥ 1% was negatively associated with the CRT response (odds ratio: 0.10, 95% confidence interval: 0.02-0.48). The most frequent cause of ineffective pacing was rapid ventricular responses due to AF and/or PVCs, whereas the cause of the highest %i-CRT in this cohort was latency. A total of three nonresponders with a high %i-CRT showed favorable responses after therapeutic interventions for their hidden causes.

Conclusion: High %i-CRT could be associated with an unfavorable CRT response and provide insights into hidden causes and therapeutic targets.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.20
自引率
14.80%
发文量
433
审稿时长
3-6 weeks
期刊介绍: Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信