Impact of acceleration treatment on treatment plan and delivery qualities in tomotherapy for lung cancer.

IF 2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Ryosuke Shirata, Tatsuya Inoue, Yugo Ebinuma, Akihiro Yamano, Takayuki Yagihashi, Hironori Nagata, Yumiko Minagawa, Yuki Mukai, Akiko Sato, Motoko Omura
{"title":"Impact of acceleration treatment on treatment plan and delivery qualities in tomotherapy for lung cancer.","authors":"Ryosuke Shirata, Tatsuya Inoue, Yugo Ebinuma, Akihiro Yamano, Takayuki Yagihashi, Hironori Nagata, Yumiko Minagawa, Yuki Mukai, Akiko Sato, Motoko Omura","doi":"10.1002/acm2.70049","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Acceleration treatment (AT) is a novel treatment planning parameter introduced in the tomotherapy-dedicated treatment planning system, Precision. This study explores the effects of AT on tomotherapy plans using helical (TomoHelical) and direct (TomoDirect) irradiation techniques.</p><p><strong>Methods: </strong>This study enrolled 20 patients with lung cancer. Initially, 10 TomoHelical and 10 TomoDirect treatment plans were created for each patient, utilizing patient-specific field width and pitch with an AT setting of 0. These original plans were subsequently reoptimized by changing only the AT values to 1, 4, 7, and 10 without changing other calculation parameters to assess the impact of AT on dosimetric and delivery parameters. Additionally, the deliverability of all plans was evaluated through patient-specific quality assurance using gamma analysis.</p><p><strong>Results: </strong>Increasing the AT from 0 to 10 led to a slight increase in maximum doses and a decrease in minimum doses within the target volume, thereby impairing dose homogeneity. Dose conformity to the target also deteriorated. Conversely, target coverage and delivery time improved considerably with higher AT values. Moreover, doses to organs at risk, including the lung, spinal cord, heart, and esophagus, remained clinically acceptable across all plans. Changes in these doses and the gamma pass rate in patient-specific quality assurance were negligible with variations in AT. This trend was consistent across both delivery techniques.</p><p><strong>Conclusion: </strong>AT is a crucial parameter in tomotherapy planning for modulating plan and delivery qualities. Higher AT values can enhance target coverage and delivery time efficiency.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e70049"},"PeriodicalIF":2.0000,"publicationDate":"2025-02-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.70049","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Acceleration treatment (AT) is a novel treatment planning parameter introduced in the tomotherapy-dedicated treatment planning system, Precision. This study explores the effects of AT on tomotherapy plans using helical (TomoHelical) and direct (TomoDirect) irradiation techniques.

Methods: This study enrolled 20 patients with lung cancer. Initially, 10 TomoHelical and 10 TomoDirect treatment plans were created for each patient, utilizing patient-specific field width and pitch with an AT setting of 0. These original plans were subsequently reoptimized by changing only the AT values to 1, 4, 7, and 10 without changing other calculation parameters to assess the impact of AT on dosimetric and delivery parameters. Additionally, the deliverability of all plans was evaluated through patient-specific quality assurance using gamma analysis.

Results: Increasing the AT from 0 to 10 led to a slight increase in maximum doses and a decrease in minimum doses within the target volume, thereby impairing dose homogeneity. Dose conformity to the target also deteriorated. Conversely, target coverage and delivery time improved considerably with higher AT values. Moreover, doses to organs at risk, including the lung, spinal cord, heart, and esophagus, remained clinically acceptable across all plans. Changes in these doses and the gamma pass rate in patient-specific quality assurance were negligible with variations in AT. This trend was consistent across both delivery techniques.

Conclusion: AT is a crucial parameter in tomotherapy planning for modulating plan and delivery qualities. Higher AT values can enhance target coverage and delivery time efficiency.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信