Marcela Mercado-Montoya, Tatiana Gomez-Bustamante, Steven R Mickelsen, Erik Kulstad, Ana González-Suárez, Lawrence J Overzet
{"title":"Thermal side effects during pulsed field ablation: an analysis using computer modelling.","authors":"Marcela Mercado-Montoya, Tatiana Gomez-Bustamante, Steven R Mickelsen, Erik Kulstad, Ana González-Suárez, Lawrence J Overzet","doi":"10.1093/europace/euaf035","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>Pulsed field ablation (PFA) is described as non-thermal, but data from oncology and cardiology show thermal effects occur. The specific waveform parameters influencing thermal energy development during PFA are unclear. The aim of this study is to numerically evaluate the thermal effects of PFA on myocardial and oesophageal tissue at various peak voltage conditions.</p><p><strong>Methods and results: </strong>A three-dimensional computer model of the left atrium quantified thermal effects from PFA at peak voltages of 1, 1.5, and 2 kV. Energy was applied using a bipolar configuration with far-field and symmetry boundaries set as electrically insulating. A monophasic waveform with a 100 μs pulse width and a 1 s gap between pulses was applied for a total of 50 pulses, mimicking clinical conditions. Minimal temperature rise in the oesophagus was observed with 1 kV pulses (214.5 J). At 1.5 and 2 kV (570.3 and 1.23 kJ), temperatures reached 46.3°C and >62°C, respectively, after a single pulse train. These findings suggest that repeated applications could lead to even higher temperatures, especially if good tissue contact is obtained. These results align with data from other medical fields using pulsed field treatments.</p><p><strong>Conclusion: </strong>Thermal effects from PFA depend on the total energy deposited, with peak voltage being a significant factor. Current commercially available PFA systems have the potential to induce collateral thermal injury with repeated applications of pulsed field energy. This highlights the need for careful monitoring and adjustment of PFA parameters in clinical settings.</p>","PeriodicalId":11981,"journal":{"name":"Europace","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2025-03-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11879925/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Europace","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/europace/euaf035","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Aims: Pulsed field ablation (PFA) is described as non-thermal, but data from oncology and cardiology show thermal effects occur. The specific waveform parameters influencing thermal energy development during PFA are unclear. The aim of this study is to numerically evaluate the thermal effects of PFA on myocardial and oesophageal tissue at various peak voltage conditions.
Methods and results: A three-dimensional computer model of the left atrium quantified thermal effects from PFA at peak voltages of 1, 1.5, and 2 kV. Energy was applied using a bipolar configuration with far-field and symmetry boundaries set as electrically insulating. A monophasic waveform with a 100 μs pulse width and a 1 s gap between pulses was applied for a total of 50 pulses, mimicking clinical conditions. Minimal temperature rise in the oesophagus was observed with 1 kV pulses (214.5 J). At 1.5 and 2 kV (570.3 and 1.23 kJ), temperatures reached 46.3°C and >62°C, respectively, after a single pulse train. These findings suggest that repeated applications could lead to even higher temperatures, especially if good tissue contact is obtained. These results align with data from other medical fields using pulsed field treatments.
Conclusion: Thermal effects from PFA depend on the total energy deposited, with peak voltage being a significant factor. Current commercially available PFA systems have the potential to induce collateral thermal injury with repeated applications of pulsed field energy. This highlights the need for careful monitoring and adjustment of PFA parameters in clinical settings.
期刊介绍:
EP - Europace - European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology. The journal aims to provide an avenue of communication of top quality European and international original scientific work and reviews in the fields of Arrhythmias, Pacing and Cellular Electrophysiology. The Journal offers the reader a collection of contemporary original peer-reviewed papers, invited papers and editorial comments together with book reviews and correspondence.