Second-Generation Wide-Field Visualization Devices for 5-ALA-Induced Fluorescence and Concepts for Validation in Neurosurgery-A Systematic Review.

Neurosurgery practice Pub Date : 2023-10-06 eCollection Date: 2023-12-01 DOI:10.1227/neuprac.0000000000000059
Zeynep Özdemir, Eric Suero Molina, Sönke J Hellwig, Herbert Stepp, Walter Stummer
{"title":"Second-Generation Wide-Field Visualization Devices for 5-ALA-Induced Fluorescence and Concepts for Validation in Neurosurgery-A Systematic Review.","authors":"Zeynep Özdemir, Eric Suero Molina, Sönke J Hellwig, Herbert Stepp, Walter Stummer","doi":"10.1227/neuprac.0000000000000059","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and objectives: </strong>Fluorescence-guided resection (FGR) of malignant gliomas with five-aminolevulinic acid (5-ALA) is an established method using surgical microscopes equipped with filter systems for observing fluorescence. Over the past decade, new technologies have been introduced for the same purpose, with available publications evaluating their clinical efficacy based on varying criteria. This study aims to review technologies and concepts of validation in the context of 5-ALA-mediated FGR.</p><p><strong>Methods: </strong>A systematic review following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses statement was performed to identify devices capable of detecting 5-ALA-induced fluorescence. Articles found eligible for this review were analyzed, focusing on the methods of validation used for novel devices. A qualitative analysis is presented.</p><p><strong>Results: </strong>Using predefined eligibility criteria, 22 studies were analyzed. Publications on the following visualization devices were reviewed: FL400 (Leica Microsystems), Aeos (Aesculap), BLUE400 and BLUE400 AR Filter System (Carl Zeiss Meditec AG), Endoscope with D-Light C (Karl Storz), Fiberscope N-4L (Machida), ORBEYE 4K 3D Digital Video Microscope (Olympus), and several customized surgical loupe systems. In many cases, validation seemed unstandardized, with inherent biases and limited reproducibility.</p><p><strong>Conclusion: </strong>This review illustrates the significance of device validation within the framework of FGR. It emphasizes the criticality of validating devices in accordance with established standard, i.e. the BLUE400 filter system, which was employed in the approval studies of 5-ALA. Furthermore, standardized concepts of validation are required to assess whether new devices are, in fact, a reliable or superior alternative in the field of FGR. Published guidelines should be considered when performing future studies.</p>","PeriodicalId":74298,"journal":{"name":"Neurosurgery practice","volume":"4 4","pages":"e00059"},"PeriodicalIF":0.0000,"publicationDate":"2023-10-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11809958/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neurosurgery practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1227/neuprac.0000000000000059","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background and objectives: Fluorescence-guided resection (FGR) of malignant gliomas with five-aminolevulinic acid (5-ALA) is an established method using surgical microscopes equipped with filter systems for observing fluorescence. Over the past decade, new technologies have been introduced for the same purpose, with available publications evaluating their clinical efficacy based on varying criteria. This study aims to review technologies and concepts of validation in the context of 5-ALA-mediated FGR.

Methods: A systematic review following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses statement was performed to identify devices capable of detecting 5-ALA-induced fluorescence. Articles found eligible for this review were analyzed, focusing on the methods of validation used for novel devices. A qualitative analysis is presented.

Results: Using predefined eligibility criteria, 22 studies were analyzed. Publications on the following visualization devices were reviewed: FL400 (Leica Microsystems), Aeos (Aesculap), BLUE400 and BLUE400 AR Filter System (Carl Zeiss Meditec AG), Endoscope with D-Light C (Karl Storz), Fiberscope N-4L (Machida), ORBEYE 4K 3D Digital Video Microscope (Olympus), and several customized surgical loupe systems. In many cases, validation seemed unstandardized, with inherent biases and limited reproducibility.

Conclusion: This review illustrates the significance of device validation within the framework of FGR. It emphasizes the criticality of validating devices in accordance with established standard, i.e. the BLUE400 filter system, which was employed in the approval studies of 5-ALA. Furthermore, standardized concepts of validation are required to assess whether new devices are, in fact, a reliable or superior alternative in the field of FGR. Published guidelines should be considered when performing future studies.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信