Dose-volume parameter evaluation of a sub-fractionation workflow for adaptive radiotherapy of prostate cancer patients on a 1.5 T magnetic resonance imaging radiotherapy system

IF 3.4 Q2 ONCOLOGY
Georgios Tsekas, Cornel Zachiu, Gijsbert H. Bol, Jochem R.N. van der Voort van Zyp, Sandrine M.G. van de Pol, Johannes C.J. de Boer, Bas W. Raaymakers
{"title":"Dose-volume parameter evaluation of a sub-fractionation workflow for adaptive radiotherapy of prostate cancer patients on a 1.5 T magnetic resonance imaging radiotherapy system","authors":"Georgios Tsekas,&nbsp;Cornel Zachiu,&nbsp;Gijsbert H. Bol,&nbsp;Jochem R.N. van der Voort van Zyp,&nbsp;Sandrine M.G. van de Pol,&nbsp;Johannes C.J. de Boer,&nbsp;Bas W. Raaymakers","doi":"10.1016/j.phro.2025.100706","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and purpose:</h3><div>This study focuses on evaluating a sub-fractionation workflow for intrafraction motion mitigation of prostate cancer patients on a 1.5 T magnetic resonance imaging radiotherapy system.</div></div><div><h3>Materials and methods:</h3><div>The investigated workflow consisted of two sub-fractions where intrafraction drift correction steps were applied based on a daily reference plan. However, the daily contours were only rigidly shifted to match the intrafraction anatomies and therefore the clinical dosimetric constraints might be violated. In this work, daily contours were deformed to match the intrafraction anatomies and the online plans were re-calculated for a total of 15 patients. The deformed prostate contours were inspected by radiation oncologists and corrections were performed when necessary. Finally, a dose-volume parameter evaluation was performed on a sub-fraction level using the clinical plan parameters.</div></div><div><h3>Results:</h3><div>Clinically acceptable coverage was reported for the target structures resulting in mean V<sub>95%</sub> of 99.7 % and 97.8 % for the clinical target volume (CTV) and planning target volume (PTV) respectively. Sub-fractions with insufficient CTV dose can be explained by the presence of intrafraction rotations and deformations that were not taken into account during intrafraction corrections. Additionally, for no sub-fraction the dose to the organs-at-risk exceeded the clinical constraints.</div></div><div><h3>Conclusion:</h3><div>Given our results on the CTV coverage we can conclude that the sub-fractionation workflow met the dosimetric constraints for the hypofractionated treatment of the analyzed group of prostate cancer patients. A future dose accumulation study can provide further insights into the suitability of the clinical margins.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"33 ","pages":"Article 100706"},"PeriodicalIF":3.4000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631625000119","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose:

This study focuses on evaluating a sub-fractionation workflow for intrafraction motion mitigation of prostate cancer patients on a 1.5 T magnetic resonance imaging radiotherapy system.

Materials and methods:

The investigated workflow consisted of two sub-fractions where intrafraction drift correction steps were applied based on a daily reference plan. However, the daily contours were only rigidly shifted to match the intrafraction anatomies and therefore the clinical dosimetric constraints might be violated. In this work, daily contours were deformed to match the intrafraction anatomies and the online plans were re-calculated for a total of 15 patients. The deformed prostate contours were inspected by radiation oncologists and corrections were performed when necessary. Finally, a dose-volume parameter evaluation was performed on a sub-fraction level using the clinical plan parameters.

Results:

Clinically acceptable coverage was reported for the target structures resulting in mean V95% of 99.7 % and 97.8 % for the clinical target volume (CTV) and planning target volume (PTV) respectively. Sub-fractions with insufficient CTV dose can be explained by the presence of intrafraction rotations and deformations that were not taken into account during intrafraction corrections. Additionally, for no sub-fraction the dose to the organs-at-risk exceeded the clinical constraints.

Conclusion:

Given our results on the CTV coverage we can conclude that the sub-fractionation workflow met the dosimetric constraints for the hypofractionated treatment of the analyzed group of prostate cancer patients. A future dose accumulation study can provide further insights into the suitability of the clinical margins.
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信