Glenoid component cyclical failure decreases with increasing baseplate contact: a biomechanical study

Q4 Medicine
Hafiz F. Kassam MD , Josiah Valk DO , Stephen Wiseman DO , Lucein N. Blaine BS , Ana Mata-Fink MD , Steven Tommasini PhD , Theodore A. Blaine MD
{"title":"Glenoid component cyclical failure decreases with increasing baseplate contact: a biomechanical study","authors":"Hafiz F. Kassam MD ,&nbsp;Josiah Valk DO ,&nbsp;Stephen Wiseman DO ,&nbsp;Lucein N. Blaine BS ,&nbsp;Ana Mata-Fink MD ,&nbsp;Steven Tommasini PhD ,&nbsp;Theodore A. Blaine MD","doi":"10.1053/j.sart.2024.08.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Glenoid baseplate loosening remains a common mode of failure in reverse shoulder arthroplasty. One of the key factors to baseplate stability is theorized to be maximization of baseplate backside contact. The purpose of this biomechanical study is to investigate the role of varying degrees of backside bony glenoid support in component stability for reverse total shoulder arthroplasty.</div></div><div><h3>Methods</h3><div>Twenty synthetic scapular models were divided into 3 test groups of 5 scapulae with glenoid baseplate contacts of 40%, 60%, and 75%, and one control group with glenoid baseplate contact of 100%. Standardized application of a commercially available glenoid baseplate and glenosphere was performed. The scapulae were mounted on a linear bearing with a humeral component and polyethylene liner which were affixed to a biaxial servohydraulic fatigue testing system. Each specimen was loaded for 10,000 cycles or to failure, about a 55° arc along the glenosphere at a rate of 1 Hz as a 750 N compression load was applied. Failure was defined as fracture of the scapula with implant fixation compromise. Before and after loading, stability of the baseplate was assessed by quantifying the total motion between the model and the baseplate with digital calipers as a ramp load between 0 and 150 N was applied. Two-sample unpaired <em>t</em>-tests were performed with significance set at <em>P</em> &lt; .05.</div></div><div><h3>Results</h3><div>Baseplate contacts of 40% (1623 ± 227, <em>P</em> = .0001), 60% (3299 ± 1170, <em>P</em> = .0001), and 75% (5615 ± 1587, <em>P</em> = .0077) demonstrated statistically significant decrease in the average number of cycles to failure in all cohorts compared to our control (8641 ± 1070). Cycles taken for initial cracks to progress to failure showed no significant differences; 40% contact (862 ± 452, <em>P</em> = .4751), 60% contact (1651 ± 996, <em>P</em> = .4318), 75% contact (2882 ± 1347, <em>P</em> = .0620), and 100% control (1166 ± 657). Baseplate contacts of 40% (6150.4 ± 444.0, <em>P</em> = .0006), 60% (4647.1 ± 552.3, <em>P</em> = .0072), and 75% (2927.8 ± 918.5, <em>P</em> = .2573) demonstrated increasing micromotion (pre-post cyclical loading) in all cohorts compared to our control (2074.7 ± 1164.6) with statistical significance at 40% and 60%.</div></div><div><h3>Conclusion</h3><div>These biomechanical tests demonstrate that decreasing glenoid baseplate backside contact leads to increased micromotion and fewer cycles to failure. This supports the surgical goal of achieving maximal glenoid baseplate backside contact, suggesting that decreased glenoid baseplate support could contribute to significant loosening.</div></div>","PeriodicalId":39885,"journal":{"name":"Seminars in Arthroplasty","volume":"35 1","pages":"Pages 22-30"},"PeriodicalIF":0.0000,"publicationDate":"2024-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Seminars in Arthroplasty","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1045452724000968","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Glenoid baseplate loosening remains a common mode of failure in reverse shoulder arthroplasty. One of the key factors to baseplate stability is theorized to be maximization of baseplate backside contact. The purpose of this biomechanical study is to investigate the role of varying degrees of backside bony glenoid support in component stability for reverse total shoulder arthroplasty.

Methods

Twenty synthetic scapular models were divided into 3 test groups of 5 scapulae with glenoid baseplate contacts of 40%, 60%, and 75%, and one control group with glenoid baseplate contact of 100%. Standardized application of a commercially available glenoid baseplate and glenosphere was performed. The scapulae were mounted on a linear bearing with a humeral component and polyethylene liner which were affixed to a biaxial servohydraulic fatigue testing system. Each specimen was loaded for 10,000 cycles or to failure, about a 55° arc along the glenosphere at a rate of 1 Hz as a 750 N compression load was applied. Failure was defined as fracture of the scapula with implant fixation compromise. Before and after loading, stability of the baseplate was assessed by quantifying the total motion between the model and the baseplate with digital calipers as a ramp load between 0 and 150 N was applied. Two-sample unpaired t-tests were performed with significance set at P < .05.

Results

Baseplate contacts of 40% (1623 ± 227, P = .0001), 60% (3299 ± 1170, P = .0001), and 75% (5615 ± 1587, P = .0077) demonstrated statistically significant decrease in the average number of cycles to failure in all cohorts compared to our control (8641 ± 1070). Cycles taken for initial cracks to progress to failure showed no significant differences; 40% contact (862 ± 452, P = .4751), 60% contact (1651 ± 996, P = .4318), 75% contact (2882 ± 1347, P = .0620), and 100% control (1166 ± 657). Baseplate contacts of 40% (6150.4 ± 444.0, P = .0006), 60% (4647.1 ± 552.3, P = .0072), and 75% (2927.8 ± 918.5, P = .2573) demonstrated increasing micromotion (pre-post cyclical loading) in all cohorts compared to our control (2074.7 ± 1164.6) with statistical significance at 40% and 60%.

Conclusion

These biomechanical tests demonstrate that decreasing glenoid baseplate backside contact leads to increased micromotion and fewer cycles to failure. This supports the surgical goal of achieving maximal glenoid baseplate backside contact, suggesting that decreased glenoid baseplate support could contribute to significant loosening.
求助全文
约1分钟内获得全文 求助全文
来源期刊
Seminars in Arthroplasty
Seminars in Arthroplasty Medicine-Surgery
CiteScore
1.00
自引率
0.00%
发文量
104
期刊介绍: Each issue of Seminars in Arthroplasty provides a comprehensive, current overview of a single topic in arthroplasty. The journal addresses orthopedic surgeons, providing authoritative reviews with emphasis on new developments relevant to their practice.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信