Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction.

IF 1.7 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Pace-Pacing and Clinical Electrophysiology Pub Date : 2025-04-01 Epub Date: 2025-02-06 DOI:10.1111/pace.15146
Alberto Preda, Francesco Melillo, Matteo Baroni, Alessandra Marzi, Vincenzo Schillaci, Sara Vargiu, Andrea Caccia, Fabrizio Guarracini, Lorenzo Gigli, Gabriele Paglino, Giulia Massaro, Igor Diemberger, Giosuè Mascioli, Francesco Solimene, Patrizio Mazzone
{"title":"Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction.","authors":"Alberto Preda, Francesco Melillo, Matteo Baroni, Alessandra Marzi, Vincenzo Schillaci, Sara Vargiu, Andrea Caccia, Fabrizio Guarracini, Lorenzo Gigli, Gabriele Paglino, Giulia Massaro, Igor Diemberger, Giosuè Mascioli, Francesco Solimene, Patrizio Mazzone","doi":"10.1111/pace.15146","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice.</p><p><strong>Methods and results: </strong>In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1-year follow-up.</p><p><strong>Conclusion: </strong>This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"436-442"},"PeriodicalIF":1.7000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15146","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/6 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice.

Methods and results: In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1-year follow-up.

Conclusion: This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.

短13Fr机械旋转扩张套用于经静脉铅提取的多中心经验。
背景:在世界范围内,经静脉铅提取(TLE)的需求正在增加。由于锁骨下静脉直到与上腔静脉交界处的过程是铅粘附的常见地方,我们的目的是评估短13Fr双向旋转机械鞘作为首选的安全性和有效性。方法和结果:在这项多中心研究中,202名心脏植入式电子装置(CIED)携带者采用短13 Fr双向旋转机械护套进行TLE,前瞻性入组。所有手术均采用逐步方法进行。TLE的适应症为感染(62%)、功能障碍(32%)和升级(6%)。总共取出了471根导联:65%起搏导联,20%除颤器导联,9%冠状动脉窦导联,6%废弃导联。临床成功率和完全拔铅率分别为97%和95%。在手术开始时,短鞘总是有效地获得静脉通路,并且在67%的病例中足以完成TLE。导联停留时间、除颤器导联、每位患者导联数和导联故障是长时间双向旋转机械护套使用的预测因素。无术中死亡病例,2%的患者出现严重并发症。1年随访时总生存率为97%。结论:使用短的13Fr双向旋转机械护套的多中心试验报告了高安全性和有效性,证明了常规使用较大卡尺的短拔牙护套作为首选的实用性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Pace-Pacing and Clinical Electrophysiology
Pace-Pacing and Clinical Electrophysiology 医学-工程:生物医学
CiteScore
2.70
自引率
5.60%
发文量
209
审稿时长
2-4 weeks
期刊介绍: Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信