Alberto Preda, Francesco Melillo, Matteo Baroni, Alessandra Marzi, Vincenzo Schillaci, Sara Vargiu, Andrea Caccia, Fabrizio Guarracini, Lorenzo Gigli, Gabriele Paglino, Giulia Massaro, Igor Diemberger, Giosuè Mascioli, Francesco Solimene, Patrizio Mazzone
{"title":"Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction.","authors":"Alberto Preda, Francesco Melillo, Matteo Baroni, Alessandra Marzi, Vincenzo Schillaci, Sara Vargiu, Andrea Caccia, Fabrizio Guarracini, Lorenzo Gigli, Gabriele Paglino, Giulia Massaro, Igor Diemberger, Giosuè Mascioli, Francesco Solimene, Patrizio Mazzone","doi":"10.1111/pace.15146","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice.</p><p><strong>Methods and results: </strong>In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1-year follow-up.</p><p><strong>Conclusion: </strong>This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15146","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice.
Methods and results: In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1-year follow-up.
Conclusion: This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.