Spatial resolution improvements for photon counting detectors using coincidence counting and frequency weighted reconstruction.

Medical physics Pub Date : 2025-02-06 DOI:10.1002/mp.17664
Scott S Hsieh, Katsuyuki Taguchi, Shuai Leng, Cynthia H McCollough
{"title":"Spatial resolution improvements for photon counting detectors using coincidence counting and frequency weighted reconstruction.","authors":"Scott S Hsieh, Katsuyuki Taguchi, Shuai Leng, Cynthia H McCollough","doi":"10.1002/mp.17664","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Photon counting X-ray detectors (PCDs) provide better spatial resolution than energy integrating X-ray detectors, but even higher resolution is desired in some applications. Certain charge sharing compensation techniques such as coincidence counting preferentially detect photons that arrive at the boundary between pixels, and this could be used for subpixel localization of incident photons and enhanced spatial resolution.</p><p><strong>Purpose: </strong>To estimate improvements to spatial resolution and detective quantum efficiency that are possible when using coincidence counting for high-resolution, non-spectral imaging.</p><p><strong>Methods: </strong>The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were estimated using numerical simulations of a two-dimensional parallel-beam CT system. Coincidence counters were modeled either geometrically or with Monte Carlo simulations. The geometric model consisted of narrow coincidence counters that interlaced with wider ordinary pixels, and showed that using standard filtered backprojection decreases low-frequency DQE, but that a frequency weighting technique could be used to restore DQE. The Monte Carlo simulations were used to estimate the possible improvements that could be expected from real systems. The pixel pitch was 0.25 mm and the source apertures considered were 0, 0.125, and 0.5 mm. A numerical stent phantom was also used to illustrate possible improvements.</p><p><strong>Results: </strong>Assuming a 0.125 mm source aperture and the Monte Carlo model, the limiting MTF (at 10%) increased from 20 to 40 lp/cm using coincidence counters. This can be explained by the increased sampling (and higher Nyquist limit) possible from coincidence counters. In the geometric model, coincidence counters are compared to conventional double sampling techniques such as in-plane flying focal spot, and the MTF at 38 lp/cm increased from 5% to 53%. Without the frequency weighting technique, low-frequency DQE was reduced by about 20%, but these losses are recovered with frequency weighting. Improvements are much more modest with the 0.25 mm source aperture because the system becomes source-limited.</p><p><strong>Conclusions: </strong>Coincidence counting could be used to increase spatial resolution in PCDs. The increases in system resolution could be large if a high-resolution X-ray source were available.</p>","PeriodicalId":94136,"journal":{"name":"Medical physics","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/mp.17664","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Photon counting X-ray detectors (PCDs) provide better spatial resolution than energy integrating X-ray detectors, but even higher resolution is desired in some applications. Certain charge sharing compensation techniques such as coincidence counting preferentially detect photons that arrive at the boundary between pixels, and this could be used for subpixel localization of incident photons and enhanced spatial resolution.

Purpose: To estimate improvements to spatial resolution and detective quantum efficiency that are possible when using coincidence counting for high-resolution, non-spectral imaging.

Methods: The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were estimated using numerical simulations of a two-dimensional parallel-beam CT system. Coincidence counters were modeled either geometrically or with Monte Carlo simulations. The geometric model consisted of narrow coincidence counters that interlaced with wider ordinary pixels, and showed that using standard filtered backprojection decreases low-frequency DQE, but that a frequency weighting technique could be used to restore DQE. The Monte Carlo simulations were used to estimate the possible improvements that could be expected from real systems. The pixel pitch was 0.25 mm and the source apertures considered were 0, 0.125, and 0.5 mm. A numerical stent phantom was also used to illustrate possible improvements.

Results: Assuming a 0.125 mm source aperture and the Monte Carlo model, the limiting MTF (at 10%) increased from 20 to 40 lp/cm using coincidence counters. This can be explained by the increased sampling (and higher Nyquist limit) possible from coincidence counters. In the geometric model, coincidence counters are compared to conventional double sampling techniques such as in-plane flying focal spot, and the MTF at 38 lp/cm increased from 5% to 53%. Without the frequency weighting technique, low-frequency DQE was reduced by about 20%, but these losses are recovered with frequency weighting. Improvements are much more modest with the 0.25 mm source aperture because the system becomes source-limited.

Conclusions: Coincidence counting could be used to increase spatial resolution in PCDs. The increases in system resolution could be large if a high-resolution X-ray source were available.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信