Risk of Fracture at External Fixator Pin Hole After Lateral Tibial Plateau Fracture Plating: A Biomechanical Comparison of Different Screw Configurations.

IF 2.3 Q2 ORTHOPEDICS
JBJS Open Access Pub Date : 2025-02-03 eCollection Date: 2025-01-01 DOI:10.2106/JBJS.OA.24.00094
Patrick A Massey, Wayne Scalisi, Chloe Duval, Michael Lowery, Brad Chauvin, Giovanni F Solitro
{"title":"Risk of Fracture at External Fixator Pin Hole After Lateral Tibial Plateau Fracture Plating: A Biomechanical Comparison of Different Screw Configurations.","authors":"Patrick A Massey, Wayne Scalisi, Chloe Duval, Michael Lowery, Brad Chauvin, Giovanni F Solitro","doi":"10.2106/JBJS.OA.24.00094","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>External fixation of tibial plateau fractures commonly provides temporary stabilization before definitive fixation with plate and screws. The purpose of this study was to determine if an external fixator pin hole distal to a tibial plate in a synthetic fracture model would increase the risk of fracture after fixation. Another objective was to determine the ideal configuration when placing tibial plate screws near an external fixator pin hole.</p><p><strong>Methods: </strong>Thirty synthetic tibiae were tested and evenly divided into 5 groups. Tibial plateau plates were placed with 4 different screw configurations for the distal-most screw near the external fixator pin hole. The 5 groups tested were control (fixation with no external fixator hole), unicortical (distal fixation with a unicortical locking screw), bicortical (distal fixation with a bicortical locking screw), oblique (distal fixation with an oblique cortical screw angled 30° proximally from the external fixator hole), and hole-bridging (hole-bridging fixation in which the plate was placed bridging the external fixator hole). The bone surrogates were potted and tested using an Instron 8874 Testing System.</p><p><strong>Results: </strong>There was a significant difference in failure load among the 5 groups (p = 0.005). The mean peak loads were 1,259 N (control), 835 N (unicortical), 831 N (bicortical), 943 N (oblique), and 993 N (hole-bridging). There was a higher failure load in the control group compared with the bicortical group (p = 0.007) and the unicortical group (p = 0.007). There was no difference in failure load between the control group and the hole-bridging group (p = 0.16) and the oblique group (p = 0.067).</p><p><strong>Conclusions: </strong>External fixator pin holes distal to a tibial plateau plate may increase the risk of tibial fracture through the pin hole. This risk may be mitigated by placing the distal screw oblique and angled proximally away from the external fixator pin hole or by placing the external fixator pin proximally with subsequent bridging of the external fixator pin hole with the plate.</p>","PeriodicalId":36492,"journal":{"name":"JBJS Open Access","volume":"10 1","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-02-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11778079/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JBJS Open Access","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2106/JBJS.OA.24.00094","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: External fixation of tibial plateau fractures commonly provides temporary stabilization before definitive fixation with plate and screws. The purpose of this study was to determine if an external fixator pin hole distal to a tibial plate in a synthetic fracture model would increase the risk of fracture after fixation. Another objective was to determine the ideal configuration when placing tibial plate screws near an external fixator pin hole.

Methods: Thirty synthetic tibiae were tested and evenly divided into 5 groups. Tibial plateau plates were placed with 4 different screw configurations for the distal-most screw near the external fixator pin hole. The 5 groups tested were control (fixation with no external fixator hole), unicortical (distal fixation with a unicortical locking screw), bicortical (distal fixation with a bicortical locking screw), oblique (distal fixation with an oblique cortical screw angled 30° proximally from the external fixator hole), and hole-bridging (hole-bridging fixation in which the plate was placed bridging the external fixator hole). The bone surrogates were potted and tested using an Instron 8874 Testing System.

Results: There was a significant difference in failure load among the 5 groups (p = 0.005). The mean peak loads were 1,259 N (control), 835 N (unicortical), 831 N (bicortical), 943 N (oblique), and 993 N (hole-bridging). There was a higher failure load in the control group compared with the bicortical group (p = 0.007) and the unicortical group (p = 0.007). There was no difference in failure load between the control group and the hole-bridging group (p = 0.16) and the oblique group (p = 0.067).

Conclusions: External fixator pin holes distal to a tibial plateau plate may increase the risk of tibial fracture through the pin hole. This risk may be mitigated by placing the distal screw oblique and angled proximally away from the external fixator pin hole or by placing the external fixator pin proximally with subsequent bridging of the external fixator pin hole with the plate.

求助全文
约1分钟内获得全文 求助全文
来源期刊
JBJS Open Access
JBJS Open Access Medicine-Surgery
CiteScore
5.00
自引率
0.00%
发文量
77
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信