Francesco Bosco , Giuseppe Rovere , Carmelo Burgio , Giorgia Lo Bue , Claudio Domenico Cobisi , Riccardo Giai Via , Ludovico Lucenti , Lawrence Camarda
{"title":"Accuracy and learning curve of imageless robotic-assisted total knee arthroplasty","authors":"Francesco Bosco , Giuseppe Rovere , Carmelo Burgio , Giorgia Lo Bue , Claudio Domenico Cobisi , Riccardo Giai Via , Ludovico Lucenti , Lawrence Camarda","doi":"10.1016/j.jor.2024.12.029","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Total knee arthroplasty (TKA) is widely used to manage severe knee osteoarthritis. However, conventional TKA (C-TKA) often leaves patients dissatisfied due to suboptimal alignment and soft-tissue balance. Robotic-assisted TKA (RA-TKA), particularly with imageless systems like the NAVIO Surgical System, promises enhanced accuracy and improved outcomes. This study aims to validate the accuracy of RA-TKA in achieving functional alignment (FA) and to explore the learning curve associated with this technique.</div></div><div><h3>Materials and methods</h3><div>A retrospective analysis included 101 patients undergoing RA-TKA with the NAVIO system from July 2021 to April 2024. Data on alignment angles, gap balance, and surgical times were analyzed. Patients were categorized by preoperative coronal alignment (valgus, neutral, and varus), with subgroups within the varus category. Accuracy was defined as deviations ≤3° for alignment and ≤1 mm for gap balance. Learning curve trends were analyzed using segmented regression.</div></div><div><h3>Results</h3><div>The study demonstrated a mean alignment error of 1.18° (±1.21) and a gap balance accuracy of 84 %, with no significant differences across knee morphologies. The RA-TKA system achieved an overall implant alignment accuracy rate of 95 %. Varus knees with greater deformities (>6°) showed comparable or superior accuracy to less severe cases. Surgical time averaged 72.3 min (±5.6), with significant reductions observed after the first 11 cases, reflecting procedural efficiency improvements without compromising accuracy.</div></div><div><h3>Conclusion</h3><div>The RA-TKA reliably achieves precise FA across diverse knee morphologies with a rapid learning curve. Future studies should evaluate long-term outcomes and implant survivorship to confirm these promising findings.</div></div><div><h3>Level of evidence</h3><div>IV.</div></div>","PeriodicalId":16633,"journal":{"name":"Journal of orthopaedics","volume":"66 ","pages":"Pages 77-83"},"PeriodicalIF":1.5000,"publicationDate":"2025-01-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11779652/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of orthopaedics","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0972978X24004574","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Total knee arthroplasty (TKA) is widely used to manage severe knee osteoarthritis. However, conventional TKA (C-TKA) often leaves patients dissatisfied due to suboptimal alignment and soft-tissue balance. Robotic-assisted TKA (RA-TKA), particularly with imageless systems like the NAVIO Surgical System, promises enhanced accuracy and improved outcomes. This study aims to validate the accuracy of RA-TKA in achieving functional alignment (FA) and to explore the learning curve associated with this technique.
Materials and methods
A retrospective analysis included 101 patients undergoing RA-TKA with the NAVIO system from July 2021 to April 2024. Data on alignment angles, gap balance, and surgical times were analyzed. Patients were categorized by preoperative coronal alignment (valgus, neutral, and varus), with subgroups within the varus category. Accuracy was defined as deviations ≤3° for alignment and ≤1 mm for gap balance. Learning curve trends were analyzed using segmented regression.
Results
The study demonstrated a mean alignment error of 1.18° (±1.21) and a gap balance accuracy of 84 %, with no significant differences across knee morphologies. The RA-TKA system achieved an overall implant alignment accuracy rate of 95 %. Varus knees with greater deformities (>6°) showed comparable or superior accuracy to less severe cases. Surgical time averaged 72.3 min (±5.6), with significant reductions observed after the first 11 cases, reflecting procedural efficiency improvements without compromising accuracy.
Conclusion
The RA-TKA reliably achieves precise FA across diverse knee morphologies with a rapid learning curve. Future studies should evaluate long-term outcomes and implant survivorship to confirm these promising findings.
期刊介绍:
Journal of Orthopaedics aims to be a leading journal in orthopaedics and contribute towards the improvement of quality of orthopedic health care. The journal publishes original research work and review articles related to different aspects of orthopaedics including Arthroplasty, Arthroscopy, Sports Medicine, Trauma, Spine and Spinal deformities, Pediatric orthopaedics, limb reconstruction procedures, hand surgery, and orthopaedic oncology. It also publishes articles on continuing education, health-related information, case reports and letters to the editor. It is requested to note that the journal has an international readership and all submissions should be aimed at specifying something about the setting in which the work was conducted. Authors must also provide any specific reasons for the research and also provide an elaborate description of the results.