Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?

IF 1.6 3区 医学 Q3 PERIPHERAL VASCULAR DISEASE
Arthur Muller, Yann Rétory, Laura Négrier, Benoit Lemarié, Anne Pfeifer, Alexandra Schmidt, Christian Dupont, Clara Virbel-Fleischman
{"title":"Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?","authors":"Arthur Muller, Yann Rétory, Laura Négrier, Benoit Lemarié, Anne Pfeifer, Alexandra Schmidt, Christian Dupont, Clara Virbel-Fleischman","doi":"10.1177/11297298251314615","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ). An in vitro study was conducted, following the guidelines of EP and drug manufacturers, to assess the impact of three parameters on the accuracy of the infusion flowrate: composition, volume, and temperature of the infused solution.</p><p><strong>Methods: </strong>A test bench was created to evaluate the accuracy of flowrate infusion when changing one parameter at a time (addition of TAZ in the infused solution, minimal filling volume and flow regulator at ambient temperature) compared to the standard conditions provided by the manufacturer. Two types of EP were used and connected to a weighing scale. Instantaneous and average mass flowrates over the total infusion time were determined from the reservoir mass loss. These values  were compared with the nominal flowrates announced by the manufacturers to evaluate the performance of EP under the different conditions. The time the flowrate was within the expected ±15% around the nominal flowrate was calculated as the time in range (TIR).</p><p><strong>Results: </strong>Deviations of the measured mean flowrate from the nominal flowrate range from -37% to 18% depending on the condition. TIR went from 0% to 80% of the total infusion duration. The test bench showed a reliability of more than 96%.</p><p><strong>Conclusions: </strong>The test bench highlights differences in performance. These results raise several questions about the sensitivity of EPs for TAZ infusion in OPAT, the norm's requirements and the need to prescribe a medical device whose limits are well known, as well as therapeutic education or the course of care for treatment administration.</p>","PeriodicalId":56113,"journal":{"name":"Journal of Vascular Access","volume":" ","pages":"11297298251314615"},"PeriodicalIF":1.6000,"publicationDate":"2025-01-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Vascular Access","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11297298251314615","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ). An in vitro study was conducted, following the guidelines of EP and drug manufacturers, to assess the impact of three parameters on the accuracy of the infusion flowrate: composition, volume, and temperature of the infused solution.

Methods: A test bench was created to evaluate the accuracy of flowrate infusion when changing one parameter at a time (addition of TAZ in the infused solution, minimal filling volume and flow regulator at ambient temperature) compared to the standard conditions provided by the manufacturer. Two types of EP were used and connected to a weighing scale. Instantaneous and average mass flowrates over the total infusion time were determined from the reservoir mass loss. These values  were compared with the nominal flowrates announced by the manufacturers to evaluate the performance of EP under the different conditions. The time the flowrate was within the expected ±15% around the nominal flowrate was calculated as the time in range (TIR).

Results: Deviations of the measured mean flowrate from the nominal flowrate range from -37% to 18% depending on the condition. TIR went from 0% to 80% of the total infusion duration. The test bench showed a reliability of more than 96%.

Conclusions: The test bench highlights differences in performance. These results raise several questions about the sensitivity of EPs for TAZ infusion in OPAT, the norm's requirements and the need to prescribe a medical device whose limits are well known, as well as therapeutic education or the course of care for treatment administration.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Vascular Access
Journal of Vascular Access 医学-外周血管病
CiteScore
3.40
自引率
31.60%
发文量
181
审稿时长
6-12 weeks
期刊介绍: The Journal of Vascular Access (JVA) is issued six times per year; it considers the publication of original manuscripts dealing with clinical and laboratory investigations in the fast growing field of vascular access. In addition reviews, case reports and clinical trials are welcome, as well as papers dedicated to more practical aspects covering new devices and techniques. All contributions, coming from all over the world, undergo the peer-review process. The Journal of Vascular Access is divided into independent sections, each led by Editors of the highest scientific level: • Dialysis • Oncology • Interventional radiology • Nutrition • Nursing • Intensive care Correspondence related to published papers is also welcome.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信