{"title":"A Prospective Pilot Study Demonstrating Noninvasive Calibration-Free Glucose Measurement.","authors":"Martina Rothenbühler, Aritz Lizoain, Fabien Rebeaud, Adler Perotte, Marc Stoffel, J Hans DeVries","doi":"10.1177/19322968251313811","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Glucose is an essential molecule in energy metabolism. Dysregulated glucose metabolism, the defining feature of diabetes, requires active monitoring and treatment to prevent significant morbidity and mortality. Current technologies for intermittent and continuous glucose measurement are invasive. Noninvasive glucose measurement would eliminate this barrier toward making glucose monitoring more accessible, extending the benefits from people living with diabetes to prediabetes and the healthy.</p><p><strong>Methods: </strong>A novel spectroscopy-based system for measuring glucose noninvasively was used in an exploratory, prospective, single-center clinical study (NCT06272136) to develop and test a machine learning-based computational model for continuous glucose monitoring without per-subject calibration. The study design blinded the development investigators to the validation analyses.</p><p><strong>Results: </strong>Twenty subjects were enrolled. Fifteen were used for the development set, and five in the validation set. All study participants were adults with insulin-treated diabetes and median glycated hemoglobin (HbA<sub>1c</sub>) of 7.3% (interquartile range [IQR] = 6.7-7.7). The computational model resulted in a mean absolute relative difference (MARD) of 14.5% and 96.5% of the paired glucose data points in the A plus B zones of the Diabetes Technology Society (DTS) error grid. The correlation between the average model sensitivity by wavelength and the spectrum of glucose was 0.45 (<i>P</i> < .001).</p><p><strong>Conclusions: </strong>Our findings suggest that Raman spectroscopy coupled with advanced computational methods can enable continuous, noninvasive glucose measurement without per-subject invasive calibration.</p>","PeriodicalId":15475,"journal":{"name":"Journal of Diabetes Science and Technology","volume":" ","pages":"19322968251313811"},"PeriodicalIF":4.1000,"publicationDate":"2025-01-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11780617/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19322968251313811","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Glucose is an essential molecule in energy metabolism. Dysregulated glucose metabolism, the defining feature of diabetes, requires active monitoring and treatment to prevent significant morbidity and mortality. Current technologies for intermittent and continuous glucose measurement are invasive. Noninvasive glucose measurement would eliminate this barrier toward making glucose monitoring more accessible, extending the benefits from people living with diabetes to prediabetes and the healthy.
Methods: A novel spectroscopy-based system for measuring glucose noninvasively was used in an exploratory, prospective, single-center clinical study (NCT06272136) to develop and test a machine learning-based computational model for continuous glucose monitoring without per-subject calibration. The study design blinded the development investigators to the validation analyses.
Results: Twenty subjects were enrolled. Fifteen were used for the development set, and five in the validation set. All study participants were adults with insulin-treated diabetes and median glycated hemoglobin (HbA1c) of 7.3% (interquartile range [IQR] = 6.7-7.7). The computational model resulted in a mean absolute relative difference (MARD) of 14.5% and 96.5% of the paired glucose data points in the A plus B zones of the Diabetes Technology Society (DTS) error grid. The correlation between the average model sensitivity by wavelength and the spectrum of glucose was 0.45 (P < .001).
Conclusions: Our findings suggest that Raman spectroscopy coupled with advanced computational methods can enable continuous, noninvasive glucose measurement without per-subject invasive calibration.
期刊介绍:
The Journal of Diabetes Science and Technology (JDST) is a bi-monthly, peer-reviewed scientific journal published by the Diabetes Technology Society. JDST covers scientific and clinical aspects of diabetes technology including glucose monitoring, insulin and metabolic peptide delivery, the artificial pancreas, digital health, precision medicine, social media, cybersecurity, software for modeling, physiologic monitoring, technology for managing obesity, and diagnostic tests of glycation. The journal also covers the development and use of mobile applications and wireless communication, as well as bioengineered tools such as MEMS, new biomaterials, and nanotechnology to develop new sensors. Articles in JDST cover both basic research and clinical applications of technologies being developed to help people with diabetes.