Long-term use of Minimed™ 780G in children and adolescents with type 1 diabetes under real-world conditions: The benefits of optimal settings

IF 5.4 2区 医学 Q1 ENDOCRINOLOGY & METABOLISM
Bruno Bombaci MD, Stefano Passanisi PhD, Marco Calderone MD, Fabio Macrì MD, Fortunato Lombardo MD, Giuseppina Salzano PhD
{"title":"Long-term use of Minimed™ 780G in children and adolescents with type 1 diabetes under real-world conditions: The benefits of optimal settings","authors":"Bruno Bombaci MD,&nbsp;Stefano Passanisi PhD,&nbsp;Marco Calderone MD,&nbsp;Fabio Macrì MD,&nbsp;Fortunato Lombardo MD,&nbsp;Giuseppina Salzano PhD","doi":"10.1111/dom.16226","DOIUrl":null,"url":null,"abstract":"<p>The MiniMed 780G™ (Medtronic, Northridge, CA), launched in 2020, is currently available in more than 40 countries worldwide. This device uses a proportional–integral–derivative algorithm combined with fuzzy logic features, adjusting basal insulin delivery rate based on customizable glycaemic targets (100, 110 or 120 mg/dL) and delivering automated correction boluses. Furthermore, the algorithm features meal detection and safe bolus technologies to minimize post-meal glycaemic excursions.<span><sup>1</sup></span></p><p>Although numerous experimental and real-world studies have demonstrated the benefits related to the MiniMed 780G™ use in youth with type 1 diabetes (T1D), there is still no strong evidence of its sustained effectiveness beyond 1 year of use.<span><sup>2-5</sup></span></p><p>The aim of this study is to evaluate glycaemic outcomes in a population of children and adolescents with T1D during their first 24 months of Minimed 780G™ system use in real-world conditions.</p><p>A single-centre, longitudinal, real-world design was adopted. Children and adolescents with T1D who began using the Minimed™ 780G system between November 2020 and May 2024 were enrolled. Local ethical committee approval was obtained (n. 39-23). Participants were included if they met the following criteria: a diagnosis of T1D, age &lt;18 years at the time of starting automated insulin delivery (AID) therapy, and provided informed consent for remote access to clinical data.</p><p>The following time points were considered across the study period: 2 weeks preceding automatic mode activation (T0), 2 weeks after activation of the automatic mode (T1), 12 months after automatic mode activation (T2), 24 months post-activation (T3). Prior to May 2022, all participants used a 3-day infusion set, whereas afterward, all users were gradually switched to a 7-day infusion set.</p><p>At each time point, we collected CGM metrics including mean sensor glucose and its standard deviation (SD), percentage of time between 70 and 180 mg/dL (TIR), percentage of time between 70 and 140 mg/dL (TITR), percentage of time above 180 mg/dL (TAR), percentage of time between 180 and 250 mg/dL (TAR<sub>Level1</sub>), percentage of time above 250 mg/dL (TAR<sub>Level2</sub>), percentage of time below 70 mg/dL (TBR), percentage of time between 54 and 70 mg/dL (TBR<sub>Level1</sub>), percentage of time below 54 mg/dL (TBR<sub>Level2</sub>), glucose management indicator (GMI), coefficient of variation (CV) and glycaemia risk index (GRI). All metrics were acquired from CareLink System® platform, considering time intervals of 15 days.</p><p>Numerical data were expressed as mean and standard deviation, while categorical variables were reported as absolute frequencies and percentages. A parametric approach was adopted since the numerical variables followed a normal distribution, as confirmed by the Kolmogorov–Smirnov test. To compare clinical variables, including glycaemic control indicators, across the different time points, paired-samples Student <i>t</i> test was used to compare baseline and the first 2 weeks of automatic mode use, and repeated-measures analysis of variance (ANOVA) was applied to compare T1, T2 and T3. Post hoc pairwise comparisons were conducted using Tukey test, with Bonferroni correction applied for multiple comparisons, yielding a corrected significance level of 0.017.</p><p>At time points T1, T2 and T3, clinical variables and system settings were compared between achievers and non-achievers of CGM targets of TITR ≥50% and TBR ≤4% using Student <i>t</i> test for numerical variables and the chi-square test for categorical variables. Additionally, a logistic regression model was applied to identify baseline predictors for the simultaneous achievement of TITR ≥50% and TBR ≤4%. Statistical analyses were performed using IBM SPSS software for Windows, Version 22 (IBM Corp., Armonk, NY, United States), with a <i>p</i> value of less than 0.05 considered statistically significant.</p><p>We included 111 children and adolescents starting therapy with the Minimed™ 780G, with a slight prevalence of males (55.0%). The mean age of the study participants was 12.7 ± 3.3 years, with an average duration of diabetes of 4.7 ± 3.4 years. Prior to using the Minimed™ 780G system, 46 individuals (41.4%) were on multiple daily insulin injections, while 65 (58.6%) were already using an insulin pump. The average HbA1c in the year preceding enrollment was 7.2% ± 0.9%. At T0, the mean BMI <i>Z</i>-score was 0.8 ± 1.2.</p><p>Based on the time of initiation of device use, all participants completed a 6-month observational period, 97 (87.4%) achieved 12-month use of the Minimed 780G, and 46 (41.4%) completed the 24-month follow-up period.</p><p>Immediately after the switch from manual to automatic mode, we observed significant improvement of TBR<sub>Level1</sub> (2.2 ± 1.7 vs. 1.9% ± 1.7%; <i>p</i> = 0.015), TBR (2.6 ± 2.2 vs. 2.4% ± 2.4%; <i>p</i> = 0.045), TITR (43.1% ± 10.1% vs. 51.5% ± 8.6%; <i>p</i> &lt;0.001), TIR (66.9 ± 10.8 vs. 76.0% ± 7.6%; <i>p</i> &lt;0.001), TAR<sub>Level1</sub> (24.0 ± 7.2 vs. 18.3 ± 5.8; <i>p</i> &lt;0.001), TAR<sub>Level2</sub> (6.5 ± 4.8 vs. 3.6% ± 3.0%; <i>p</i> &lt;0.001), TAR (30.6 ± 10.8 vs. 21.9% ± 7.8%; <i>p</i> &lt;0.001), CV (34.6 ± 4.0 vs. 33.0% ± 4.0%; <i>p</i> &lt;0.001), mean sensor glucose (156.5 ± 15.6 vs. 144.7 ± 11.3 mg/dL; <i>p</i> &lt;0.001) and GRI (36.3 ± 12.5 vs. 23.4 ± 8.0; <i>p</i> &lt;0.001).</p><p>Glucose metrics remained statistically unchanged over the subsequent 24 months of follow-up, with no significant differences between T1, T2 and T3 (Figure 1).</p><p>When comparing achievers and non-achievers of TITR ≥50% and a TBR ≤4% at T1, we found lower percentage of automatic correction boluses (<i>p</i> = 0.009) and more frequent use of active insulin time (AIT) 2 h (<i>p</i> = 0.005) and optimal settings, consisting of both AIT 2 h and SmartGuard target 100 mg/dL (<i>p</i> = 0.003) among achievers. Similar findings were observed at T2 (Table 1).</p><p>The logistic regression analysis did not identify any baseline predictors for the simultaneous achievement of TITR ≥50% and TBR ≤4%, except for an inverse association between the mean HbA1c value in the 12 months prior to enrollment and target achievement at T2 (<i>B</i> = −1.126; 95% CI 0.173–0.607; <i>p</i> &lt;0.001).</p><p>Our data confirm the immediate benefits of the Minimed™ 780G use in automatic mode, resulting in significant improvements of key CGM metrics, including TIR and TITR, and reduction in hypoglycaemic events. These results align with previous studies that reported substantial short-term benefits of AID systems in youths, starting within 2 weeks after the activation of automatic mode.<span><sup>5, 6</sup></span></p><p>Notably, our data revealed that improvements in CGM data were sustained up to 24 months, with no significant differences between 12- and 24-month metrics. Similarly, a prospective study on 50 youths showed sustained performance of the device in terms of TIR across 2-year study period.<span><sup>7</sup></span> Another follow-up analysis on 35 preschoolers participating to a non-randomized, prospective, single-arm clinical trial, revealed persistence of positive effect of the Minimed™ 780G on glycaemic control up to 18 months.<span><sup>8</sup></span> The stability of glycaemic outcomes up to 2 years is particularly noteworthy given the real-world challenges of device adherence and behavioural variability in children and adolescents. Diabetes management in youths is often characterized by numerous challenges, including specific physiological factors that bring significant daily glucose excursions and psychological and behavioural issues typical of this age, with a significant portion of adolescents not achieving recommended glycaemic targets.<span><sup>9, 10</sup></span></p><p>In our analysis, we focused on the characteristics of subjects achieving a TITR of 50% without exceeding the recommended threshold for time spent in hypoglycemia. TITR has been recently proposed as an alternative metric to TIR for clinical practice, as it more accurately reflects time spent in the euglycaemic range. Interestingly, we found that participants who met glycaemic targets of TITR ≥50% and TBR ≤4% were more likely to use more stringent device settings. However, this association was not confirmed after 24 months. Further studies with larger cohorts are needed to validate this trend and to elucidate potential underlying determinants. AIT of 2 h along with SmartGuard target set at 100 mg/dL have already been identified as predictors of improved TIR among 12 870 Minimed™ 780G users.<span><sup>11</sup></span> A subanalysis focusing on individuals aged ≤15 years from the same population further demonstrated that the percentage of individuals simultaneously achieving the recommended targets for TIR, TBR and GMI was significantly higher among those using optimal settings than the general population.<span><sup>12</sup></span> Similarly, consistent use of AIT of 2 h and a SmartGuard target of 100 mg/dL has been strongly associated to higher TITR levels.<span><sup>6</sup></span></p><p>Finally, our analysis revealed a reduced percentage of automatic correction boluses among participants who simultaneously achieved TITR and TBR targets. This finding was consistent across all time points and aligns with previous studies demonstrating an association between a higher proportion of automatic correction boluses and suboptimal glycaemic outcomes among AID users.<span><sup>5, 11</sup></span> A study on youths with T1D using the MiniMed™ 780G reported that an automatic correction bolus percentage below 30% was associated with higher levels of TIR and TITR.<span><sup>13</sup></span> This trend underscores the critical role of user-initiated boluses and accurate carbohydrate counting in achieving optimal glycaemic control.<span><sup>14</sup></span></p><p>Despite its strengths, our study has some limitations, including the decreasing sample size over the 24-month period, which reflects the fact that some participants were enrolled more recently and the single-center design, that may limit the generalizability of the findings.</p><p>In conclusion, our study demonstrated that the Minimed™ 780G provides sustained improvements in glycaemic control for up to 24 months in children and adolescents with T1D. Achieving time in tight range targets without increasing the risk of hypoglycemia is strongly associated with the use of optimal device settings.</p><p>BB and SP conceptualized the study and wrote the first draft of the article. FM and MC researched data. GS reviewed and edited the manuscript. FL contributed to discussion and reviewed and edited the manuscript. All authors approved the final version of the manuscript. SP is the guarantor of this work and, as such, had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.</p><p>No funding was received for this article.</p><p>BB reports travel grants from Movi SpA and Abbott. FL has received speaker and consultant honoraria from Sanofi and speaking honoraria from Movi SpA. SP received speaking honoraria from Movi SpA.</p>","PeriodicalId":158,"journal":{"name":"Diabetes, Obesity & Metabolism","volume":"27 4","pages":"2309-2312"},"PeriodicalIF":5.4000,"publicationDate":"2025-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/dom.16226","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diabetes, Obesity & Metabolism","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/dom.16226","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

Abstract

The MiniMed 780G™ (Medtronic, Northridge, CA), launched in 2020, is currently available in more than 40 countries worldwide. This device uses a proportional–integral–derivative algorithm combined with fuzzy logic features, adjusting basal insulin delivery rate based on customizable glycaemic targets (100, 110 or 120 mg/dL) and delivering automated correction boluses. Furthermore, the algorithm features meal detection and safe bolus technologies to minimize post-meal glycaemic excursions.1

Although numerous experimental and real-world studies have demonstrated the benefits related to the MiniMed 780G™ use in youth with type 1 diabetes (T1D), there is still no strong evidence of its sustained effectiveness beyond 1 year of use.2-5

The aim of this study is to evaluate glycaemic outcomes in a population of children and adolescents with T1D during their first 24 months of Minimed 780G™ system use in real-world conditions.

A single-centre, longitudinal, real-world design was adopted. Children and adolescents with T1D who began using the Minimed™ 780G system between November 2020 and May 2024 were enrolled. Local ethical committee approval was obtained (n. 39-23). Participants were included if they met the following criteria: a diagnosis of T1D, age <18 years at the time of starting automated insulin delivery (AID) therapy, and provided informed consent for remote access to clinical data.

The following time points were considered across the study period: 2 weeks preceding automatic mode activation (T0), 2 weeks after activation of the automatic mode (T1), 12 months after automatic mode activation (T2), 24 months post-activation (T3). Prior to May 2022, all participants used a 3-day infusion set, whereas afterward, all users were gradually switched to a 7-day infusion set.

At each time point, we collected CGM metrics including mean sensor glucose and its standard deviation (SD), percentage of time between 70 and 180 mg/dL (TIR), percentage of time between 70 and 140 mg/dL (TITR), percentage of time above 180 mg/dL (TAR), percentage of time between 180 and 250 mg/dL (TARLevel1), percentage of time above 250 mg/dL (TARLevel2), percentage of time below 70 mg/dL (TBR), percentage of time between 54 and 70 mg/dL (TBRLevel1), percentage of time below 54 mg/dL (TBRLevel2), glucose management indicator (GMI), coefficient of variation (CV) and glycaemia risk index (GRI). All metrics were acquired from CareLink System® platform, considering time intervals of 15 days.

Numerical data were expressed as mean and standard deviation, while categorical variables were reported as absolute frequencies and percentages. A parametric approach was adopted since the numerical variables followed a normal distribution, as confirmed by the Kolmogorov–Smirnov test. To compare clinical variables, including glycaemic control indicators, across the different time points, paired-samples Student t test was used to compare baseline and the first 2 weeks of automatic mode use, and repeated-measures analysis of variance (ANOVA) was applied to compare T1, T2 and T3. Post hoc pairwise comparisons were conducted using Tukey test, with Bonferroni correction applied for multiple comparisons, yielding a corrected significance level of 0.017.

At time points T1, T2 and T3, clinical variables and system settings were compared between achievers and non-achievers of CGM targets of TITR ≥50% and TBR ≤4% using Student t test for numerical variables and the chi-square test for categorical variables. Additionally, a logistic regression model was applied to identify baseline predictors for the simultaneous achievement of TITR ≥50% and TBR ≤4%. Statistical analyses were performed using IBM SPSS software for Windows, Version 22 (IBM Corp., Armonk, NY, United States), with a p value of less than 0.05 considered statistically significant.

We included 111 children and adolescents starting therapy with the Minimed™ 780G, with a slight prevalence of males (55.0%). The mean age of the study participants was 12.7 ± 3.3 years, with an average duration of diabetes of 4.7 ± 3.4 years. Prior to using the Minimed™ 780G system, 46 individuals (41.4%) were on multiple daily insulin injections, while 65 (58.6%) were already using an insulin pump. The average HbA1c in the year preceding enrollment was 7.2% ± 0.9%. At T0, the mean BMI Z-score was 0.8 ± 1.2.

Based on the time of initiation of device use, all participants completed a 6-month observational period, 97 (87.4%) achieved 12-month use of the Minimed 780G, and 46 (41.4%) completed the 24-month follow-up period.

Immediately after the switch from manual to automatic mode, we observed significant improvement of TBRLevel1 (2.2 ± 1.7 vs. 1.9% ± 1.7%; p = 0.015), TBR (2.6 ± 2.2 vs. 2.4% ± 2.4%; p = 0.045), TITR (43.1% ± 10.1% vs. 51.5% ± 8.6%; p <0.001), TIR (66.9 ± 10.8 vs. 76.0% ± 7.6%; p <0.001), TARLevel1 (24.0 ± 7.2 vs. 18.3 ± 5.8; p <0.001), TARLevel2 (6.5 ± 4.8 vs. 3.6% ± 3.0%; p <0.001), TAR (30.6 ± 10.8 vs. 21.9% ± 7.8%; p <0.001), CV (34.6 ± 4.0 vs. 33.0% ± 4.0%; p <0.001), mean sensor glucose (156.5 ± 15.6 vs. 144.7 ± 11.3 mg/dL; p <0.001) and GRI (36.3 ± 12.5 vs. 23.4 ± 8.0; p <0.001).

Glucose metrics remained statistically unchanged over the subsequent 24 months of follow-up, with no significant differences between T1, T2 and T3 (Figure 1).

When comparing achievers and non-achievers of TITR ≥50% and a TBR ≤4% at T1, we found lower percentage of automatic correction boluses (p = 0.009) and more frequent use of active insulin time (AIT) 2 h (p = 0.005) and optimal settings, consisting of both AIT 2 h and SmartGuard target 100 mg/dL (p = 0.003) among achievers. Similar findings were observed at T2 (Table 1).

The logistic regression analysis did not identify any baseline predictors for the simultaneous achievement of TITR ≥50% and TBR ≤4%, except for an inverse association between the mean HbA1c value in the 12 months prior to enrollment and target achievement at T2 (B = −1.126; 95% CI 0.173–0.607; p <0.001).

Our data confirm the immediate benefits of the Minimed™ 780G use in automatic mode, resulting in significant improvements of key CGM metrics, including TIR and TITR, and reduction in hypoglycaemic events. These results align with previous studies that reported substantial short-term benefits of AID systems in youths, starting within 2 weeks after the activation of automatic mode.5, 6

Notably, our data revealed that improvements in CGM data were sustained up to 24 months, with no significant differences between 12- and 24-month metrics. Similarly, a prospective study on 50 youths showed sustained performance of the device in terms of TIR across 2-year study period.7 Another follow-up analysis on 35 preschoolers participating to a non-randomized, prospective, single-arm clinical trial, revealed persistence of positive effect of the Minimed™ 780G on glycaemic control up to 18 months.8 The stability of glycaemic outcomes up to 2 years is particularly noteworthy given the real-world challenges of device adherence and behavioural variability in children and adolescents. Diabetes management in youths is often characterized by numerous challenges, including specific physiological factors that bring significant daily glucose excursions and psychological and behavioural issues typical of this age, with a significant portion of adolescents not achieving recommended glycaemic targets.9, 10

In our analysis, we focused on the characteristics of subjects achieving a TITR of 50% without exceeding the recommended threshold for time spent in hypoglycemia. TITR has been recently proposed as an alternative metric to TIR for clinical practice, as it more accurately reflects time spent in the euglycaemic range. Interestingly, we found that participants who met glycaemic targets of TITR ≥50% and TBR ≤4% were more likely to use more stringent device settings. However, this association was not confirmed after 24 months. Further studies with larger cohorts are needed to validate this trend and to elucidate potential underlying determinants. AIT of 2 h along with SmartGuard target set at 100 mg/dL have already been identified as predictors of improved TIR among 12 870 Minimed™ 780G users.11 A subanalysis focusing on individuals aged ≤15 years from the same population further demonstrated that the percentage of individuals simultaneously achieving the recommended targets for TIR, TBR and GMI was significantly higher among those using optimal settings than the general population.12 Similarly, consistent use of AIT of 2 h and a SmartGuard target of 100 mg/dL has been strongly associated to higher TITR levels.6

Finally, our analysis revealed a reduced percentage of automatic correction boluses among participants who simultaneously achieved TITR and TBR targets. This finding was consistent across all time points and aligns with previous studies demonstrating an association between a higher proportion of automatic correction boluses and suboptimal glycaemic outcomes among AID users.5, 11 A study on youths with T1D using the MiniMed™ 780G reported that an automatic correction bolus percentage below 30% was associated with higher levels of TIR and TITR.13 This trend underscores the critical role of user-initiated boluses and accurate carbohydrate counting in achieving optimal glycaemic control.14

Despite its strengths, our study has some limitations, including the decreasing sample size over the 24-month period, which reflects the fact that some participants were enrolled more recently and the single-center design, that may limit the generalizability of the findings.

In conclusion, our study demonstrated that the Minimed™ 780G provides sustained improvements in glycaemic control for up to 24 months in children and adolescents with T1D. Achieving time in tight range targets without increasing the risk of hypoglycemia is strongly associated with the use of optimal device settings.

BB and SP conceptualized the study and wrote the first draft of the article. FM and MC researched data. GS reviewed and edited the manuscript. FL contributed to discussion and reviewed and edited the manuscript. All authors approved the final version of the manuscript. SP is the guarantor of this work and, as such, had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

No funding was received for this article.

BB reports travel grants from Movi SpA and Abbott. FL has received speaker and consultant honoraria from Sanofi and speaking honoraria from Movi SpA. SP received speaking honoraria from Movi SpA.

Abstract Image

求助全文
约1分钟内获得全文 求助全文
来源期刊
Diabetes, Obesity & Metabolism
Diabetes, Obesity & Metabolism 医学-内分泌学与代谢
CiteScore
10.90
自引率
6.90%
发文量
319
审稿时长
3-8 weeks
期刊介绍: Diabetes, Obesity and Metabolism is primarily a journal of clinical and experimental pharmacology and therapeutics covering the interrelated areas of diabetes, obesity and metabolism. The journal prioritises high-quality original research that reports on the effects of new or existing therapies, including dietary, exercise and lifestyle (non-pharmacological) interventions, in any aspect of metabolic and endocrine disease, either in humans or animal and cellular systems. ‘Metabolism’ may relate to lipids, bone and drug metabolism, or broader aspects of endocrine dysfunction. Preclinical pharmacology, pharmacokinetic studies, meta-analyses and those addressing drug safety and tolerability are also highly suitable for publication in this journal. Original research may be published as a main paper or as a research letter.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信