{"title":"Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance-Guided Catheter.","authors":"Hitoshi Mori, Hidehira Fukaya, Kazuhisa Matsumoto, Masataka Narita, Tsukasa Naganuma, Wataru Sasaki, Naomichi Tanaka, Daisuke Kawano, Yoshifumi Ikeda, Kazuo Matsumoto, Ritsushi Kato","doi":"10.1111/pace.15144","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The IntellaNav MiFi OI catheter (MiFi) is equipped with a sensor for local impedance (LI) monitoring and three mini-electrodes. In this study, we investigated the target LI values for a successful pulmonary vein isolation (PVI) under the pacing and ablation technique using the MiFi catheter.</p><p><strong>Methods: </strong>Twenty-seven patients underwent PVI using the MiFi catheter under mini electrode pacing from the MiFi catheter. The local impedance (LI) changes, generator impedance (GI) changes, and the time to capture loss were evaluated.</p><p><strong>Results: </strong>First-pass isolations were obtained in 15 patients (57.7 %) for right PVs and in 22 patients (84.6 %) for left PVs. At gap sites, the impedance decrease was smaller than at non-gap sites (non-gap sites vs. gap sites; LI drop, 23.2 [±10.3] vs. 15.6 [±7.7] Ω, p < 0.0001; GI drop, 4.8 [±4.1] vs. 2.7 [3.9] Ω, p = 0.0026; %LI drop, -19.3 [±7.4] vs. -13.1 [±6.1] %, p < 0.0001; % GI drop, -5.1 [±4.2] vs. -2.9 [±4.2] %, p = 0.0020), suggesting that changes in impedance could be useful for predicting gaps. The cutoff values for predicting no gaps were identified as 15.0 Ω for the LI drop and -13.74% for the %LI drop.</p><p><strong>Conclusion: </strong>The LI showed greater changes than the GI and was also useful for predicting gaps. The cutoff values of 15.0 Ω for the LI drop and -13.74% for the %LI drop could predict conduction gaps. Under the monitoring of the LI, the pacing and ablation technique proved useful for PVI, even though the MiFi catheter does not have a CF sensor or ablation indices.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"216-223"},"PeriodicalIF":1.7000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11822088/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15144","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/27 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The IntellaNav MiFi OI catheter (MiFi) is equipped with a sensor for local impedance (LI) monitoring and three mini-electrodes. In this study, we investigated the target LI values for a successful pulmonary vein isolation (PVI) under the pacing and ablation technique using the MiFi catheter.
Methods: Twenty-seven patients underwent PVI using the MiFi catheter under mini electrode pacing from the MiFi catheter. The local impedance (LI) changes, generator impedance (GI) changes, and the time to capture loss were evaluated.
Results: First-pass isolations were obtained in 15 patients (57.7 %) for right PVs and in 22 patients (84.6 %) for left PVs. At gap sites, the impedance decrease was smaller than at non-gap sites (non-gap sites vs. gap sites; LI drop, 23.2 [±10.3] vs. 15.6 [±7.7] Ω, p < 0.0001; GI drop, 4.8 [±4.1] vs. 2.7 [3.9] Ω, p = 0.0026; %LI drop, -19.3 [±7.4] vs. -13.1 [±6.1] %, p < 0.0001; % GI drop, -5.1 [±4.2] vs. -2.9 [±4.2] %, p = 0.0020), suggesting that changes in impedance could be useful for predicting gaps. The cutoff values for predicting no gaps were identified as 15.0 Ω for the LI drop and -13.74% for the %LI drop.
Conclusion: The LI showed greater changes than the GI and was also useful for predicting gaps. The cutoff values of 15.0 Ω for the LI drop and -13.74% for the %LI drop could predict conduction gaps. Under the monitoring of the LI, the pacing and ablation technique proved useful for PVI, even though the MiFi catheter does not have a CF sensor or ablation indices.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.