Effects of In Vivo Contact Force on Pulsed-Field Ablation Efficacy in Porcine Ventricles.

IF 2.3 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Taro Koya, Naoto Otsuka, Jason A Tri, Yang Ying, Elyana Espinoza, Samuel J Asirvatham, Duy T Nguyen
{"title":"Effects of In Vivo Contact Force on Pulsed-Field Ablation Efficacy in Porcine Ventricles.","authors":"Taro Koya, Naoto Otsuka, Jason A Tri, Yang Ying, Elyana Espinoza, Samuel J Asirvatham, Duy T Nguyen","doi":"10.1111/jce.16581","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Pulsed-field ablation (PFA) is an innovative non-thermal method for arrhythmia treatment. The efficacy of various PFA configurations in relation to contact force (CF) has not been well-studied in vivo.</p><p><strong>Objectives: </strong>This study evaluated the effect of CF on acute bipolar PFA lesions in both a vegetal and an in vivo porcine heart model.</p><p><strong>Methods: </strong>Bipolar PFA (500 V, 20 μs for 50 pulses) was applied to eight porcine ventricles, with CF categorized as standard (5-15 g), high (16-30 g), or very high (> 30 g) using the CARTO 3 system. PFA lesions were assessed by gross pathology, histology, and post-ablation voltage amplitude. Additionally, a vegetal model using oscillating russet potatoes replicated the motion of a beating heart to compare lesion parameters under dynamic conditions and various PFA configurations (monopolar vs. bipolar).</p><p><strong>Results: </strong>A total of 61 in vivo bipolar PFA lesions across all groups were analyzed. No significant differences in lesion surface diameter, depth, or volume were found among CF groups. In the vegetal model, monopolar setting had a better correlation between CF and lesion depth compared to bipolar setting. In addition, oscillations and catheter position (perpendicular or parallel) significantly affected lesion dimensions.</p><p><strong>Conclusion: </strong>Increasing CF has a minimal effect on the size of acute bipolar PFA lesions in the porcine ventricle, suggesting that, for bipolar PFA, adequate CF is more critical than its intensity in a beating heart. For monopolar PFA, greater CF can impact lesion depth.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-01-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16581","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Pulsed-field ablation (PFA) is an innovative non-thermal method for arrhythmia treatment. The efficacy of various PFA configurations in relation to contact force (CF) has not been well-studied in vivo.

Objectives: This study evaluated the effect of CF on acute bipolar PFA lesions in both a vegetal and an in vivo porcine heart model.

Methods: Bipolar PFA (500 V, 20 μs for 50 pulses) was applied to eight porcine ventricles, with CF categorized as standard (5-15 g), high (16-30 g), or very high (> 30 g) using the CARTO 3 system. PFA lesions were assessed by gross pathology, histology, and post-ablation voltage amplitude. Additionally, a vegetal model using oscillating russet potatoes replicated the motion of a beating heart to compare lesion parameters under dynamic conditions and various PFA configurations (monopolar vs. bipolar).

Results: A total of 61 in vivo bipolar PFA lesions across all groups were analyzed. No significant differences in lesion surface diameter, depth, or volume were found among CF groups. In the vegetal model, monopolar setting had a better correlation between CF and lesion depth compared to bipolar setting. In addition, oscillations and catheter position (perpendicular or parallel) significantly affected lesion dimensions.

Conclusion: Increasing CF has a minimal effect on the size of acute bipolar PFA lesions in the porcine ventricle, suggesting that, for bipolar PFA, adequate CF is more critical than its intensity in a beating heart. For monopolar PFA, greater CF can impact lesion depth.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.20
自引率
14.80%
发文量
433
审稿时长
3-6 weeks
期刊介绍: Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信