Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique.

IF 2.1 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Thibault Willaume, Matthieu Ehlinger, Henri Favreau, Noëlle Weingertner, Pierre-Emmanuel Zorn, Jean-Philippe Dillenseger, Guillaume Koch, Michel Velten, Guillaume Bierry
{"title":"Assessment of knee cartilage using accelerated 3 T MRI: Evaluation of an isotropic 3D fast spin-echo sequence (CUBE) with compressed sensing technique.","authors":"Thibault Willaume, Matthieu Ehlinger, Henri Favreau, Noëlle Weingertner, Pierre-Emmanuel Zorn, Jean-Philippe Dillenseger, Guillaume Koch, Michel Velten, Guillaume Bierry","doi":"10.1016/j.mri.2025.110321","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Compressed Sensing (CS) is an emerging technique to accelerate MRI acquisitions. The aim of this study was to assess the reliability and accuracy of cartilage thickness measurements in the knee using a CS-enabled isotropic 3D Fast Spin-Echo (FSE) sequence on a 3-T MRI scanner.</p><p><strong>Methods: </strong>Twenty-eight tibial condyle sections were collected from 14 adult patients who underwent total knee arthroplasty. An isotropic 3D PDw FSE with variable flip-angle (CUBE) sequence with CS was used to acquire MR images of the tibial condyle sections. Minimum cartilage thickness measurements were independently performed by two experienced readers (R1 and R2) on MR images and compared to corresponding anatomical sections measurements. Intraclass correlation coefficients (ICCs) and Bland-Altman analyses, were used to assess agreement between MR and anatomical measurements.</p><p><strong>Results: </strong>A total of 84 paired cartilage areas were analyzed [cartilage thickness measurements ranged from 0 to 3.40 mm at anatomical evaluation (mean, 1.08 mm ± 0.83)]. The agreements between MR and anatomical measurements were excellent (mean differences, 0.06 ± 0.31 mm for R1 and 0.03 ± 0.43 mm for R2) with respective ICC values of 0.93 and 0.88. Bland-Altman analyses revealed small differences between MR and anatomical measurements, with 95 % Limit of Agreements values falling within clinically acceptable ranges (-0.54 to 0.66 mm for R1, -0.87 to 0.80 mm for R2).</p><p><strong>Conclusion: </strong>The 3D PDw FSE sequence with Compressed Sensing acceleration technique demonstrated reliable and accurate assessment of cartilage thickness of the tibial condyles within a timeframe suitable for routine clinical practice.</p>","PeriodicalId":18165,"journal":{"name":"Magnetic resonance imaging","volume":" ","pages":"110321"},"PeriodicalIF":2.1000,"publicationDate":"2025-01-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic resonance imaging","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.mri.2025.110321","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Compressed Sensing (CS) is an emerging technique to accelerate MRI acquisitions. The aim of this study was to assess the reliability and accuracy of cartilage thickness measurements in the knee using a CS-enabled isotropic 3D Fast Spin-Echo (FSE) sequence on a 3-T MRI scanner.

Methods: Twenty-eight tibial condyle sections were collected from 14 adult patients who underwent total knee arthroplasty. An isotropic 3D PDw FSE with variable flip-angle (CUBE) sequence with CS was used to acquire MR images of the tibial condyle sections. Minimum cartilage thickness measurements were independently performed by two experienced readers (R1 and R2) on MR images and compared to corresponding anatomical sections measurements. Intraclass correlation coefficients (ICCs) and Bland-Altman analyses, were used to assess agreement between MR and anatomical measurements.

Results: A total of 84 paired cartilage areas were analyzed [cartilage thickness measurements ranged from 0 to 3.40 mm at anatomical evaluation (mean, 1.08 mm ± 0.83)]. The agreements between MR and anatomical measurements were excellent (mean differences, 0.06 ± 0.31 mm for R1 and 0.03 ± 0.43 mm for R2) with respective ICC values of 0.93 and 0.88. Bland-Altman analyses revealed small differences between MR and anatomical measurements, with 95 % Limit of Agreements values falling within clinically acceptable ranges (-0.54 to 0.66 mm for R1, -0.87 to 0.80 mm for R2).

Conclusion: The 3D PDw FSE sequence with Compressed Sensing acceleration technique demonstrated reliable and accurate assessment of cartilage thickness of the tibial condyles within a timeframe suitable for routine clinical practice.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Magnetic resonance imaging
Magnetic resonance imaging 医学-核医学
CiteScore
4.70
自引率
4.00%
发文量
194
审稿时长
83 days
期刊介绍: Magnetic Resonance Imaging (MRI) is the first international multidisciplinary journal encompassing physical, life, and clinical science investigations as they relate to the development and use of magnetic resonance imaging. MRI is dedicated to both basic research, technological innovation and applications, providing a single forum for communication among radiologists, physicists, chemists, biochemists, biologists, engineers, internists, pathologists, physiologists, computer scientists, and mathematicians.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信