{"title":"New gantry angle-dependent beam control optimization with Elekta linear accelerator for VMAT delivery.","authors":"Adriaan Abraham van Appeldoorn, Johannes Gerardus Maria Kok, Jochem Willem Heiko Wolthaus","doi":"10.1002/acm2.14598","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>This paper describes a method to improve gantry-dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions. These interruptions are known to contribute to unwanted intrafraction motion and longer treatment times.</p><p><strong>Methods: </strong>Compared to the default method of the manufacturer, this new method takes both clockwise and counterclockwise rotation to compensate for magnetic hysteresis as well as previous configuration and noise filtering into account. The improved method to determine the lookup table uses service graphing information from the linac without the need for additional symmetry information. The clinical configuration of the flattened beam energies remains untouched during the data record.</p><p><strong>Results: </strong>This method results in a configuration where the gantry-dependent steering is optimized over the full arc with optimal balance in the hysteresis and minimizing the effect of errors in the steering values. This method is a less error-prone process compared to the methodology described in previous research, still achieving a reduction of interruption of about 60 percent compared to the Elekta method.</p><p><strong>Conclusion: </strong>This study shows a simplified way to optimize linac stability with improved LUT. The optimized LUT results in a lower number of interruptions, preventing downtime, and a lower risk of intrafraction motion due to longer treatment time.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14598"},"PeriodicalIF":2.0000,"publicationDate":"2025-01-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14598","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: This paper describes a method to improve gantry-dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions. These interruptions are known to contribute to unwanted intrafraction motion and longer treatment times.
Methods: Compared to the default method of the manufacturer, this new method takes both clockwise and counterclockwise rotation to compensate for magnetic hysteresis as well as previous configuration and noise filtering into account. The improved method to determine the lookup table uses service graphing information from the linac without the need for additional symmetry information. The clinical configuration of the flattened beam energies remains untouched during the data record.
Results: This method results in a configuration where the gantry-dependent steering is optimized over the full arc with optimal balance in the hysteresis and minimizing the effect of errors in the steering values. This method is a less error-prone process compared to the methodology described in previous research, still achieving a reduction of interruption of about 60 percent compared to the Elekta method.
Conclusion: This study shows a simplified way to optimize linac stability with improved LUT. The optimized LUT results in a lower number of interruptions, preventing downtime, and a lower risk of intrafraction motion due to longer treatment time.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic