Comparative evaluation of peptide vs. protein-based calibration for quantification of cardiac troponin I using ID-LC-MS/MS.

IF 3.8 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Meltem Asicioglu, Claudia Swart, Evren Saban, Emrah Yurek, Nevin Gul Karaguler, Merve Oztug
{"title":"Comparative evaluation of peptide vs. protein-based calibration for quantification of cardiac troponin I using ID-LC-MS/MS.","authors":"Meltem Asicioglu, Claudia Swart, Evren Saban, Emrah Yurek, Nevin Gul Karaguler, Merve Oztug","doi":"10.1515/cclm-2024-0999","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.</p><p><strong>Methods: </strong>To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.</p><p><strong>Results: </strong>The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).</p><p><strong>Conclusions: </strong>This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2025-01-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry and laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1515/cclm-2024-0999","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.

Methods: To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.

Results: The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).

Conclusions: This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical chemistry and laboratory medicine
Clinical chemistry and laboratory medicine 医学-医学实验技术
CiteScore
11.30
自引率
16.20%
发文量
306
审稿时长
3 months
期刊介绍: Clinical Chemistry and Laboratory Medicine (CCLM) publishes articles on novel teaching and training methods applicable to laboratory medicine. CCLM welcomes contributions on the progress in fundamental and applied research and cutting-edge clinical laboratory medicine. It is one of the leading journals in the field, with an impact factor over 3. CCLM is issued monthly, and it is published in print and electronically. CCLM is the official journal of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) and publishes regularly EFLM recommendations and news. CCLM is the official journal of the National Societies from Austria (ÖGLMKC); Belgium (RBSLM); Germany (DGKL); Hungary (MLDT); Ireland (ACBI); Italy (SIBioC); Portugal (SPML); and Slovenia (SZKK); and it is affiliated to AACB (Australia) and SFBC (France). Topics: - clinical biochemistry - clinical genomics and molecular biology - clinical haematology and coagulation - clinical immunology and autoimmunity - clinical microbiology - drug monitoring and analysis - evaluation of diagnostic biomarkers - disease-oriented topics (cardiovascular disease, cancer diagnostics, diabetes) - new reagents, instrumentation and technologies - new methodologies - reference materials and methods - reference values and decision limits - quality and safety in laboratory medicine - translational laboratory medicine - clinical metrology Follow @cclm_degruyter on Twitter!
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信