Cephalomedullary Nailing for Subtrochanteric and Reverse-Oblique Femoral Fractures: Comparison of a Single Versus Dual Lag Screw Device.

IF 4.4 1区 医学 Q1 ORTHOPEDICS
Andrew D Ablett, Conor McCann, Tony Feng, Victoria Macaskill, William M Oliver, John F Keating
{"title":"Cephalomedullary Nailing for Subtrochanteric and Reverse-Oblique Femoral Fractures: Comparison of a Single Versus Dual Lag Screw Device.","authors":"Andrew D Ablett, Conor McCann, Tony Feng, Victoria Macaskill, William M Oliver, John F Keating","doi":"10.2106/JBJS.24.00404","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Subtrochanteric proximal femoral fractures are generally treated with cephalomedullary nail fixation. We aimed to compare outcomes of subtrochanteric fracture fixation using a single lag screw (Gamma3 nail, GN) or dual lag screw (INTERTAN nail, IN) device.</p><p><strong>Methods: </strong>The primary outcome measure was mechanical failure, defined as lag screw cut-out or back-out, nail breakage, or peri-implant fracture. Secondary outcomes included reoperation for mechanical failure, deep infection, or nonunion, and technical predictors of mechanical failure. Adult patients (≥18 years of age) with a subtrochanteric proximal femoral fracture treated at a single center were retrospectively identified using electronic records. All patients who underwent fixation using either a long GN (November 2010 to January 2017) or IN (March 2017 to April 2022) were included. Medical records and radiographs were reviewed to identify operative complications.</p><p><strong>Results: </strong>A total of 587 patients were included: 336 in the GN group (median age, 82 years; 73% female) and 251 in the IN group (median age, 82 years; 71% female). The risk of mechanical failure was 3-fold higher in the GN group (adjusted hazard ratio [aHR], 2.87; p = 0.010), with screw cut-out (p = 0.04) and back-out (p = 0.04) only observed in the GN group. We observed a greater risk of reoperation for mechanical failure in the GN group, but this did not achieve significance (aHR, 2.02; p = 0.16). Independent predictors of mechanical failure included varus malalignment of >5° for cut-out (aHR, 17.43; p = 0.012), a tip-to-apex distance of >25 mm for back-out (aHR, 9.47; p = 0.019), and shortening of >1 cm for peri-implant fracture (aHR, 5.44; p = 0.001).</p><p><strong>Conclusions: </strong>For older patients with subtrochanteric and reverse-oblique femoral fractures, the dual lag screw design of the IN nail was associated with a lower risk of mechanical failure compared with the single lag screw design of the GN nail.</p><p><strong>Level of evidence: </strong>Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.</p>","PeriodicalId":15273,"journal":{"name":"Journal of Bone and Joint Surgery, American Volume","volume":" ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2024-12-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Bone and Joint Surgery, American Volume","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2106/JBJS.24.00404","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Subtrochanteric proximal femoral fractures are generally treated with cephalomedullary nail fixation. We aimed to compare outcomes of subtrochanteric fracture fixation using a single lag screw (Gamma3 nail, GN) or dual lag screw (INTERTAN nail, IN) device.

Methods: The primary outcome measure was mechanical failure, defined as lag screw cut-out or back-out, nail breakage, or peri-implant fracture. Secondary outcomes included reoperation for mechanical failure, deep infection, or nonunion, and technical predictors of mechanical failure. Adult patients (≥18 years of age) with a subtrochanteric proximal femoral fracture treated at a single center were retrospectively identified using electronic records. All patients who underwent fixation using either a long GN (November 2010 to January 2017) or IN (March 2017 to April 2022) were included. Medical records and radiographs were reviewed to identify operative complications.

Results: A total of 587 patients were included: 336 in the GN group (median age, 82 years; 73% female) and 251 in the IN group (median age, 82 years; 71% female). The risk of mechanical failure was 3-fold higher in the GN group (adjusted hazard ratio [aHR], 2.87; p = 0.010), with screw cut-out (p = 0.04) and back-out (p = 0.04) only observed in the GN group. We observed a greater risk of reoperation for mechanical failure in the GN group, but this did not achieve significance (aHR, 2.02; p = 0.16). Independent predictors of mechanical failure included varus malalignment of >5° for cut-out (aHR, 17.43; p = 0.012), a tip-to-apex distance of >25 mm for back-out (aHR, 9.47; p = 0.019), and shortening of >1 cm for peri-implant fracture (aHR, 5.44; p = 0.001).

Conclusions: For older patients with subtrochanteric and reverse-oblique femoral fractures, the dual lag screw design of the IN nail was associated with a lower risk of mechanical failure compared with the single lag screw design of the GN nail.

Level of evidence: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
8.90
自引率
7.50%
发文量
660
审稿时长
1 months
期刊介绍: The Journal of Bone & Joint Surgery (JBJS) has been the most valued source of information for orthopaedic surgeons and researchers for over 125 years and is the gold standard in peer-reviewed scientific information in the field. A core journal and essential reading for general as well as specialist orthopaedic surgeons worldwide, The Journal publishes evidence-based research to enhance the quality of care for orthopaedic patients. Standards of excellence and high quality are maintained in everything we do, from the science of the content published to the customer service we provide. JBJS is an independent, non-profit journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信