Multiple b value diffusion-weighted MRI of liver: A novel respiratory frequency-modulated continuous-wave radar-trigger technique and comparison with free-breathing technique

IF 2.1 4区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Kai Liu , Caizhong Chen , Tingting Shen , Xixi Wen , Mengsu Zeng , Pengju Xu
{"title":"Multiple b value diffusion-weighted MRI of liver: A novel respiratory frequency-modulated continuous-wave radar-trigger technique and comparison with free-breathing technique","authors":"Kai Liu ,&nbsp;Caizhong Chen ,&nbsp;Tingting Shen ,&nbsp;Xixi Wen ,&nbsp;Mengsu Zeng ,&nbsp;Pengju Xu","doi":"10.1016/j.mri.2024.110312","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>The aim of this study was to evaluate a novel respiratory frequency-modulated continuous-wave radar-trigger (FT) technique for multiple -b-value diffusion-weighted imaging (DWI) of liver and compare it with conventional free breathing (FB) DWI technique.</div></div><div><h3>Material and methods</h3><div>39 patients with focal liver lesions underwent both frequency-modulated continuous-wave radar-trigger (FT) and conventional free-breathing (FB) multi-b-value diffusion-weighted imaging (DWI,b = 0,50,400,800 s/mm<sup>2)</sup>. Two abdominal radiologists independently assessed the quality of liver DWI images obtained using both techniques, measured and compared liver signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) at different b-values, as well as apparent diffusion coefficient (ADC) values calculated from all b-values.</div></div><div><h3>Results</h3><div>In terms of image quality, the FT technique is superior to the conventional FB technique, with overall image quality scores (Reader 1, 3.56 ± 0.50 and Reader 2, 3.59 ± 0.55)vs (Reader 1, 2.90 ± 0.75 and Reader 2, 2.97 ± 0.71), respectively. The liver SNR (at b-values of 50,400,and 800 s/mm<sup>2</sup>) obtained by FT was (138.5 ± 43.48, 96.67 ± 31.95, 71.54 ± 22.03), respectively, which was significantly higher than that obtained by conventional FB (110.90 ± 39.28, 80.86 ± 29.13, 60.43 ± 18.61, <em>P</em> &lt; 0.05). The lesion CNR with FT was significantly higher than that with conventional FB (258.99 ± 151.38 vs 174.60 ± 99.90; 164.56 ± 87.25 vs 111.12 ± 42.43; 118.83 ± 68.76 vs 76.01 ± 35.48, <em>P</em> &lt; 0.001). There was no significant difference in ADC values of liver and lesions between the two techniques: ADCliver-L and ADCliver-R: (FT 1479.3 ± 270.0 vs FB 1529.3 ± 275.5 and FT 1219.6 ± 127.4 vs FB 1248.7 ± 168.2, <em>P</em> &gt; 0.05); ADC lesion:FT(969.0 ± 261.3) vs FB (1017.5 ± 240.4, <em>P</em> &gt; 0.05).</div></div><div><h3>Conclusion</h3><div>For multi-b-value liver diffusion-weighted imaging, FT technique has higher image quality and better lesion visibility than conventional FB technique and there is no significant difference in ADC values of liver and lesions between the two techniques.</div></div>","PeriodicalId":18165,"journal":{"name":"Magnetic resonance imaging","volume":"117 ","pages":"Article 110312"},"PeriodicalIF":2.1000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic resonance imaging","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0730725X24002935","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

The aim of this study was to evaluate a novel respiratory frequency-modulated continuous-wave radar-trigger (FT) technique for multiple -b-value diffusion-weighted imaging (DWI) of liver and compare it with conventional free breathing (FB) DWI technique.

Material and methods

39 patients with focal liver lesions underwent both frequency-modulated continuous-wave radar-trigger (FT) and conventional free-breathing (FB) multi-b-value diffusion-weighted imaging (DWI,b = 0,50,400,800 s/mm2). Two abdominal radiologists independently assessed the quality of liver DWI images obtained using both techniques, measured and compared liver signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) at different b-values, as well as apparent diffusion coefficient (ADC) values calculated from all b-values.

Results

In terms of image quality, the FT technique is superior to the conventional FB technique, with overall image quality scores (Reader 1, 3.56 ± 0.50 and Reader 2, 3.59 ± 0.55)vs (Reader 1, 2.90 ± 0.75 and Reader 2, 2.97 ± 0.71), respectively. The liver SNR (at b-values of 50,400,and 800 s/mm2) obtained by FT was (138.5 ± 43.48, 96.67 ± 31.95, 71.54 ± 22.03), respectively, which was significantly higher than that obtained by conventional FB (110.90 ± 39.28, 80.86 ± 29.13, 60.43 ± 18.61, P < 0.05). The lesion CNR with FT was significantly higher than that with conventional FB (258.99 ± 151.38 vs 174.60 ± 99.90; 164.56 ± 87.25 vs 111.12 ± 42.43; 118.83 ± 68.76 vs 76.01 ± 35.48, P < 0.001). There was no significant difference in ADC values of liver and lesions between the two techniques: ADCliver-L and ADCliver-R: (FT 1479.3 ± 270.0 vs FB 1529.3 ± 275.5 and FT 1219.6 ± 127.4 vs FB 1248.7 ± 168.2, P > 0.05); ADC lesion:FT(969.0 ± 261.3) vs FB (1017.5 ± 240.4, P > 0.05).

Conclusion

For multi-b-value liver diffusion-weighted imaging, FT technique has higher image quality and better lesion visibility than conventional FB technique and there is no significant difference in ADC values of liver and lesions between the two techniques.
肝脏多重b值弥散加权MRI:一种新的呼吸调频连续波雷达触发技术及与自由呼吸技术的比较。
目的:研究一种新的呼吸调频连续波雷达触发(FT)技术用于肝脏多b值弥散加权成像(DWI),并与传统自由呼吸(FB) DWI技术进行比较。材料与方法:39例局灶性肝脏病变患者行调频连续波雷达触发(FT)和常规自由呼吸(FB)多b值弥散加权成像(DWI,b = 0,50,400,800 s/mm2)。两名腹部放射科医生独立评估了使用这两种技术获得的肝脏DWI图像的质量,测量并比较了不同b值下肝脏的信噪比(SNR)和噪声对比比(CNR),以及由所有b值计算的表观扩散系数(ADC)值。结果:在图像质量方面,FT技术优于传统FB技术,其总体图像质量得分分别为(Reader 1, 3.56 ± 0.50和Reader 2, 3.59 ± 0.55)vs (Reader 1, 2.90 ± 0.75和Reader 2, 2.97 ± 0.71)。肝脏信噪比(b值的50400年和800年)英国《金融时报》获得的是(138.5 ± 43.48,96.67 ± 31.95,71.54 ± 22.03),分别获得的显著高于传统的FB(110.90 ± 39.28,80.86 ± 29.13,60.43 ± 18.61,P  0.05);ADC病变:英尺(969.0 ±261.3 )和facebook(1017.5 ± 240.4,P > 0.05)。结论:对于多b值肝脏弥散加权成像,FT技术比常规FB技术具有更高的图像质量和更好的病变可见性,两种技术在肝脏和病变的ADC值上无显著差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Magnetic resonance imaging
Magnetic resonance imaging 医学-核医学
CiteScore
4.70
自引率
4.00%
发文量
194
审稿时长
83 days
期刊介绍: Magnetic Resonance Imaging (MRI) is the first international multidisciplinary journal encompassing physical, life, and clinical science investigations as they relate to the development and use of magnetic resonance imaging. MRI is dedicated to both basic research, technological innovation and applications, providing a single forum for communication among radiologists, physicists, chemists, biochemists, biologists, engineers, internists, pathologists, physiologists, computer scientists, and mathematicians.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信