Multiple b value diffusion-weighted MRI of liver: A novel respiratory frequency-modulated continuous-wave radar-trigger technique and comparison with free-breathing technique.
IF 2.1 4区 医学Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
{"title":"Multiple b value diffusion-weighted MRI of liver: A novel respiratory frequency-modulated continuous-wave radar-trigger technique and comparison with free-breathing technique.","authors":"Kai Liu, Caizhong Chen, Tingting Shen, Xixi Wen, Mengsu Zeng, Pengju Xu","doi":"10.1016/j.mri.2024.110312","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The aim of this study was to evaluate a novel respiratory frequency-modulated continuous-wave radar-trigger (FT) technique for multiple -b-value diffusion-weighted imaging (DWI) of liver and compare it with conventional free breathing (FB) DWI technique.</p><p><strong>Material and methods: </strong>39 patients with focal liver lesions underwent both frequency-modulated continuous-wave radar-trigger (FT) and conventional free-breathing (FB) multi-b-value diffusion-weighted imaging (DWI,b = 0,50,400,800 s/mm<sup>2)</sup>. Two abdominal radiologists independently assessed the quality of liver DWI images obtained using both techniques, measured and compared liver signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) at different b-values, as well as apparent diffusion coefficient (ADC) values calculated from all b-values.</p><p><strong>Results: </strong>In terms of image quality, the FT technique is superior to the conventional FB technique, with overall image quality scores (Reader 1, 3.56 ± 0.50 and Reader 2, 3.59 ± 0.55)vs (Reader 1, 2.90 ± 0.75 and Reader 2, 2.97 ± 0.71), respectively. The liver SNR (at b-values of 50,400,and 800 s/mm<sup>2</sup>) obtained by FT was (138.5 ± 43.48, 96.67 ± 31.95, 71.54 ± 22.03), respectively, which was significantly higher than that obtained by conventional FB (110.90 ± 39.28, 80.86 ± 29.13, 60.43 ± 18.61, P < 0.05). The lesion CNR with FT was significantly higher than that with conventional FB (258.99 ± 151.38 vs 174.60 ± 99.90; 164.56 ± 87.25 vs 111.12 ± 42.43; 118.83 ± 68.76 vs 76.01 ± 35.48, P < 0.001). There was no significant difference in ADC values of liver and lesions between the two techniques: ADCliver-L and ADCliver-R: (FT 1479.3 ± 270.0 vs FB 1529.3 ± 275.5 and FT 1219.6 ± 127.4 vs FB 1248.7 ± 168.2, P > 0.05); ADC lesion:FT(969.0 ± 261.3) vs FB (1017.5 ± 240.4, P > 0.05).</p><p><strong>Conclusion: </strong>For multi-b-value liver diffusion-weighted imaging, FT technique has higher image quality and better lesion visibility than conventional FB technique and there is no significant difference in ADC values of liver and lesions between the two techniques.</p>","PeriodicalId":18165,"journal":{"name":"Magnetic resonance imaging","volume":" ","pages":"110312"},"PeriodicalIF":2.1000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic resonance imaging","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.mri.2024.110312","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: The aim of this study was to evaluate a novel respiratory frequency-modulated continuous-wave radar-trigger (FT) technique for multiple -b-value diffusion-weighted imaging (DWI) of liver and compare it with conventional free breathing (FB) DWI technique.
Material and methods: 39 patients with focal liver lesions underwent both frequency-modulated continuous-wave radar-trigger (FT) and conventional free-breathing (FB) multi-b-value diffusion-weighted imaging (DWI,b = 0,50,400,800 s/mm2). Two abdominal radiologists independently assessed the quality of liver DWI images obtained using both techniques, measured and compared liver signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) at different b-values, as well as apparent diffusion coefficient (ADC) values calculated from all b-values.
Results: In terms of image quality, the FT technique is superior to the conventional FB technique, with overall image quality scores (Reader 1, 3.56 ± 0.50 and Reader 2, 3.59 ± 0.55)vs (Reader 1, 2.90 ± 0.75 and Reader 2, 2.97 ± 0.71), respectively. The liver SNR (at b-values of 50,400,and 800 s/mm2) obtained by FT was (138.5 ± 43.48, 96.67 ± 31.95, 71.54 ± 22.03), respectively, which was significantly higher than that obtained by conventional FB (110.90 ± 39.28, 80.86 ± 29.13, 60.43 ± 18.61, P < 0.05). The lesion CNR with FT was significantly higher than that with conventional FB (258.99 ± 151.38 vs 174.60 ± 99.90; 164.56 ± 87.25 vs 111.12 ± 42.43; 118.83 ± 68.76 vs 76.01 ± 35.48, P < 0.001). There was no significant difference in ADC values of liver and lesions between the two techniques: ADCliver-L and ADCliver-R: (FT 1479.3 ± 270.0 vs FB 1529.3 ± 275.5 and FT 1219.6 ± 127.4 vs FB 1248.7 ± 168.2, P > 0.05); ADC lesion:FT(969.0 ± 261.3) vs FB (1017.5 ± 240.4, P > 0.05).
Conclusion: For multi-b-value liver diffusion-weighted imaging, FT technique has higher image quality and better lesion visibility than conventional FB technique and there is no significant difference in ADC values of liver and lesions between the two techniques.
期刊介绍:
Magnetic Resonance Imaging (MRI) is the first international multidisciplinary journal encompassing physical, life, and clinical science investigations as they relate to the development and use of magnetic resonance imaging. MRI is dedicated to both basic research, technological innovation and applications, providing a single forum for communication among radiologists, physicists, chemists, biochemists, biologists, engineers, internists, pathologists, physiologists, computer scientists, and mathematicians.