A comparison of routine [68Ga]Ga-PSMA-11 preparation using Locametz and Illuccix kits

IF 4.4 Q1 CHEMISTRY, INORGANIC & NUCLEAR
Ivan E. Wang, Luke J. Morrissette, Ka Kit Wong, Allen F. Brooks, Marianna Dakanali, Peter J. H. Scott
{"title":"A comparison of routine [68Ga]Ga-PSMA-11 preparation using Locametz and Illuccix kits","authors":"Ivan E. Wang,&nbsp;Luke J. Morrissette,&nbsp;Ka Kit Wong,&nbsp;Allen F. Brooks,&nbsp;Marianna Dakanali,&nbsp;Peter J. H. Scott","doi":"10.1186/s41181-024-00317-4","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Approval of Locametz and Illuccix kits for the manufacture of [<sup>68</sup>Ga]Ga-PSMA-11 (gallium Ga68 gozetotide), a PET imaging agent for prostate cancer, as well as the corresponding therapeutic ([<sup>177</sup>Lu]Lu-PSMA-617 Pluvicto), has led to a rapid increase in demand for [<sup>68</sup>Ga]Ga-PSMA-11 PET imaging. Radiopharmaceutical manufacturers, using <sup>68</sup>Ge/<sup>68</sup>Ga generators, may decide to adopt Locametz and/or Illuccix kits, which requires a comparison to select the most suitable kit for day-to-day use. The objective of this article is to compare both kits and provide guidance for selecting one for routine use, as well as evaluate labeling consistency of both kits during routine production. Additionally, we report our experience during 1.5 years of daily [<sup>68</sup>Ga]Ga-PSMA-11 production at our facility using both kits.</p><h3>Results</h3><p>Locametz (n = 181) and Illuccix (n = 256) kits were prepared using non-silicone coated and silicone-coated needles with <sup>68</sup>Ga activities ranging from 0.53 to 3.16 GBq, with a failure rate of 1 in 128 runs for both kits. With Locametz, a 3.7 GBq generator and 10-min incubation at room temperature gave doses that passed quality control (QC) testing. Use of non-silicone coated needles in the process led to solution discoloration, and QC failure. Additionally, lack of vial inversion led to inconsistent labeling, which improved with subsequent vial agitation. For Illuccix, addition of the acetate buffer to the precursor vial prior to adding the [<sup>68</sup>Ga]GaCl<sub>3</sub> simplifies the workflow. The maximum tolerated activity was 1.85 GBq. Lack of vial inversion led to failures, which were rectified by agitating the vial to properly incorporate the acetate solution with the generator eluate.</p><h3>Conclusions</h3><p>Both kits benefited from using a syringe pump to elute the <sup>68</sup>Ge/<sup>68</sup>Ga generator, vial agitation, and longer length/smaller bore silicone coated needles. Both kits have similar workflows, comparable QC outcomes, and result in equivalent clinical images. Thus, the decision between kits will ultimately be determined by production preferences. Since radiopharmacies have an established “kit-based” workflow, Locametz kits with higher allowed activities and longer shelf-life may offer benefits. Conversely, more traditional PET manufacturing facilities might benefit from using Illuccix kits due to compatibility with cyclotron-produced [<sup>68</sup>Ga]GaCl<sub>3</sub> allowing for kit batching. Ultimately, the commercial availability of 2 approved kits for production of [<sup>68</sup>Ga]Ga-PSMA-11 PET has facilitated ready access to this important new imaging agent.</p></div>","PeriodicalId":534,"journal":{"name":"EJNMMI Radiopharmacy and Chemistry","volume":"9 1","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2024-12-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ejnmmipharmchem.springeropen.com/counter/pdf/10.1186/s41181-024-00317-4","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Radiopharmacy and Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://link.springer.com/article/10.1186/s41181-024-00317-4","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, INORGANIC & NUCLEAR","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Approval of Locametz and Illuccix kits for the manufacture of [68Ga]Ga-PSMA-11 (gallium Ga68 gozetotide), a PET imaging agent for prostate cancer, as well as the corresponding therapeutic ([177Lu]Lu-PSMA-617 Pluvicto), has led to a rapid increase in demand for [68Ga]Ga-PSMA-11 PET imaging. Radiopharmaceutical manufacturers, using 68Ge/68Ga generators, may decide to adopt Locametz and/or Illuccix kits, which requires a comparison to select the most suitable kit for day-to-day use. The objective of this article is to compare both kits and provide guidance for selecting one for routine use, as well as evaluate labeling consistency of both kits during routine production. Additionally, we report our experience during 1.5 years of daily [68Ga]Ga-PSMA-11 production at our facility using both kits.

Results

Locametz (n = 181) and Illuccix (n = 256) kits were prepared using non-silicone coated and silicone-coated needles with 68Ga activities ranging from 0.53 to 3.16 GBq, with a failure rate of 1 in 128 runs for both kits. With Locametz, a 3.7 GBq generator and 10-min incubation at room temperature gave doses that passed quality control (QC) testing. Use of non-silicone coated needles in the process led to solution discoloration, and QC failure. Additionally, lack of vial inversion led to inconsistent labeling, which improved with subsequent vial agitation. For Illuccix, addition of the acetate buffer to the precursor vial prior to adding the [68Ga]GaCl3 simplifies the workflow. The maximum tolerated activity was 1.85 GBq. Lack of vial inversion led to failures, which were rectified by agitating the vial to properly incorporate the acetate solution with the generator eluate.

Conclusions

Both kits benefited from using a syringe pump to elute the 68Ge/68Ga generator, vial agitation, and longer length/smaller bore silicone coated needles. Both kits have similar workflows, comparable QC outcomes, and result in equivalent clinical images. Thus, the decision between kits will ultimately be determined by production preferences. Since radiopharmacies have an established “kit-based” workflow, Locametz kits with higher allowed activities and longer shelf-life may offer benefits. Conversely, more traditional PET manufacturing facilities might benefit from using Illuccix kits due to compatibility with cyclotron-produced [68Ga]GaCl3 allowing for kit batching. Ultimately, the commercial availability of 2 approved kits for production of [68Ga]Ga-PSMA-11 PET has facilitated ready access to this important new imaging agent.

使用Locametz和Illuccix试剂盒制备[68Ga]Ga-PSMA-11的常规方法比较
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
7.20
自引率
8.70%
发文量
30
审稿时长
5 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信