Three-dimensional dosimetry using multiple-energy delivery and a single-layer detector for quality assurance in proton pencil beam scanning

IF 3.2 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Medical physics Pub Date : 2024-12-16 DOI:10.1002/mp.17584
Roberto Righetto, Elena Fogazzi, Francesco Tommasino, Paolo Farace
{"title":"Three-dimensional dosimetry using multiple-energy delivery and a single-layer detector for quality assurance in proton pencil beam scanning","authors":"Roberto Righetto,&nbsp;Elena Fogazzi,&nbsp;Francesco Tommasino,&nbsp;Paolo Farace","doi":"10.1002/mp.17584","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>In Proton Therapy, the presence of implants along the beam path is known to potentially affect the dose distribution. The way such implants are managed in the planning process can vary in the different treatment planning systems (TPSs) and different centers. A specific validation procedure should be accomplished to verify the accuracy of TPS computation in these conditions and accept the applied process before treating patients.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>The aim of this study is to present a quality assurance (QA) tool in pencil beam scanning proton therapy by a method based on multiple-energy delivery and a single-layer two-dimensional detector and to apply it for verifying three-dimensional dose computation and correcting CT calibration in the presence of implants.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Multiple-energy delivery with a single-layer detector (MESL) acquisitions were performed for 80 energy layers (70-150MeV), composed of equally weighted pencil beam spots. MESL measures were acquired using a two-dimensional MatriXX-IBA detector. A transformation of the energy modulation to spatial modulation was obtained by using the power-law relationship of initial energy and range. The setup design involved a reference configuration, allowing to compensate for potential offsets, and the same configuration with an additional phantom to be measured. Both setups were imaged by a CT scanner, and the dose was computed by the TPS. The comparison of TPS-computed and MESL-measured data of the phantom was performed by producing a 2D map of range-error. For testing the procedure, plastic slabs and rods made of tissue equivalent materials (TEMs), with known water equivalent path length (WEPL), were used. Range error mapping was then applied to verify dose computation with a titanium cylinder and a titanium implant. Numerical procedures were obtained by modifying at the TPS the segmented volume, or the value in the CT calibration curve for the titanium objects. The optimal values were then determined by identifying the one that minimizes residual range error.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The results of the consistency test on the plastic slabs and the TEM rods showed differences between measured and expected WEPL below 1%, confirming the reliability of the method and the energy-spatial transformation. In the titanium cylinder, the optimal volume and the point in the calibration curve (relative to the titanium saturated value), to be used for TPS simulation is about the real size of the cylinder and the tabulated stopping power value. However, the optimal value to be assigned to the CT calibration curve might depend on the type and shape of the object, as they were different for the cylinder and the implant with screws.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The availability of a QA tool, like the one presented, paves the way for systematic studies of all the parameters that impact computation accuracy, and the methods to improve the accuracy of TPS computation.</p>\n </section>\n </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 3","pages":"1769-1778"},"PeriodicalIF":3.2000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/mp.17584","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

In Proton Therapy, the presence of implants along the beam path is known to potentially affect the dose distribution. The way such implants are managed in the planning process can vary in the different treatment planning systems (TPSs) and different centers. A specific validation procedure should be accomplished to verify the accuracy of TPS computation in these conditions and accept the applied process before treating patients.

Purpose

The aim of this study is to present a quality assurance (QA) tool in pencil beam scanning proton therapy by a method based on multiple-energy delivery and a single-layer two-dimensional detector and to apply it for verifying three-dimensional dose computation and correcting CT calibration in the presence of implants.

Methods

Multiple-energy delivery with a single-layer detector (MESL) acquisitions were performed for 80 energy layers (70-150MeV), composed of equally weighted pencil beam spots. MESL measures were acquired using a two-dimensional MatriXX-IBA detector. A transformation of the energy modulation to spatial modulation was obtained by using the power-law relationship of initial energy and range. The setup design involved a reference configuration, allowing to compensate for potential offsets, and the same configuration with an additional phantom to be measured. Both setups were imaged by a CT scanner, and the dose was computed by the TPS. The comparison of TPS-computed and MESL-measured data of the phantom was performed by producing a 2D map of range-error. For testing the procedure, plastic slabs and rods made of tissue equivalent materials (TEMs), with known water equivalent path length (WEPL), were used. Range error mapping was then applied to verify dose computation with a titanium cylinder and a titanium implant. Numerical procedures were obtained by modifying at the TPS the segmented volume, or the value in the CT calibration curve for the titanium objects. The optimal values were then determined by identifying the one that minimizes residual range error.

Results

The results of the consistency test on the plastic slabs and the TEM rods showed differences between measured and expected WEPL below 1%, confirming the reliability of the method and the energy-spatial transformation. In the titanium cylinder, the optimal volume and the point in the calibration curve (relative to the titanium saturated value), to be used for TPS simulation is about the real size of the cylinder and the tabulated stopping power value. However, the optimal value to be assigned to the CT calibration curve might depend on the type and shape of the object, as they were different for the cylinder and the implant with screws.

Conclusions

The availability of a QA tool, like the one presented, paves the way for systematic studies of all the parameters that impact computation accuracy, and the methods to improve the accuracy of TPS computation.

利用多能量传输和单层探测器进行三维剂量测定,以保证质子铅笔束扫描的质量。
背景:众所周知,在质子治疗中,沿着射束路径存在植入物可能会影响剂量分布。不同的治疗计划系统(TPS)和不同的治疗中心在计划过程中管理植入物的方式可能会有所不同。目的:本研究旨在通过基于多能量传输和单层二维探测器的方法,介绍铅笔束扫描质子治疗中的质量保证(QA)工具,并将其用于验证三维剂量计算和修正植入物存在时的 CT 校准:方法:对80个能量层(70-150MeV)进行了多能量传输和单层探测器(MESL)采集,这些能量层由权重相等的铅笔束点组成。MESL 测量使用二维 MatriXX-IBA 探测器采集。利用初始能量和范围的幂律关系将能量调制转换为空间调制。设置设计包括一个参考配置(可补偿潜在偏移)和一个附加待测模型的相同配置。两种设置均由 CT 扫描仪成像,并由 TPS 计算剂量。通过生成二维范围误差图,对 TPS 计算的数据和 MESL 测量的模型数据进行比较。为了测试该程序,使用了由组织等效材料(TEM)制成的塑料板和塑料棒,它们具有已知的水等效路径长度(WEPL)。然后将范围误差绘图应用于验证钛圆柱体和钛植入物的剂量计算。通过在 TPS 上修改钛物体的分割体积或 CT 校准曲线中的值,获得了数值程序。然后通过确定最小残余范围误差值来确定最佳值:塑料板和 TEM 棒的一致性测试结果表明,测量值和预期 WEPL 之间的差异低于 1%,这证实了该方法和能量空间转换的可靠性。在钛圆柱体中,用于 TPS 模拟的最佳体积和校准曲线上的点(相对于钛饱和值)与圆柱体的实际尺寸和表格中的停止功率值差不多。但是,CT 校准曲线的最佳值可能取决于物体的类型和形状,因为圆柱体和带螺钉的种植体的类型和形状不同:质量保证工具的出现,为系统研究影响计算精度的所有参数以及提高 TPS 计算精度的方法铺平了道路。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Medical physics
Medical physics 医学-核医学
CiteScore
6.80
自引率
15.80%
发文量
660
审稿时长
1.7 months
期刊介绍: Medical Physics publishes original, high impact physics, imaging science, and engineering research that advances patient diagnosis and therapy through contributions in 1) Basic science developments with high potential for clinical translation 2) Clinical applications of cutting edge engineering and physics innovations 3) Broadly applicable and innovative clinical physics developments Medical Physics is a journal of global scope and reach. By publishing in Medical Physics your research will reach an international, multidisciplinary audience including practicing medical physicists as well as physics- and engineering based translational scientists. We work closely with authors of promising articles to improve their quality.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信