Daniel Leslie James Morris, Avneet Minhas, Katherine Walstow, Lisa Pitt, Marie Morgan, Tim Cresswell, Marius P Espag, David I Clark, Amol A Tambe
{"title":"Does long-term surveillance of primary linked total elbow arthroplasty identify failing implants requiring revision?","authors":"Daniel Leslie James Morris, Avneet Minhas, Katherine Walstow, Lisa Pitt, Marie Morgan, Tim Cresswell, Marius P Espag, David I Clark, Amol A Tambe","doi":"10.1177/17585732241301356","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Scoping review has identified a lack of evidence guiding long-term follow-up of elbow arthroplasty. We report the effectiveness of primary linked total elbow arthroplasty surveillance in identifying failing implants requiring revision.</p><p><strong>Methods: </strong>A prospective database recording consecutive primary linked total elbow arthroplasty and subsequent surveillance in an elbow unit was analysed. Arthroplasties performed between 01.10.2013 and 31.07.2022 were included, with a minimum 1-year follow-up. Surveillance involves specialist physiotherapist review 1, 2, 3, 5, 8 and 10 years postoperatively. Patient-initiated review could occur between time points. Outcome measures include a number of surveillance reviews offered and attended; and the proportion that identified a failing implant requiring revision.</p><p><strong>Results: </strong>Ninety-seven primary linked total elbow arthroplasties with minimum 1-year follow-up were performed (76 Discovery, 14 Nexel, 7 Coonrad/Morrey). Sixteen patients died prior to 31.07.2023, and three implants required revision <1 year postoperatively. 290 of 328 offered surveillance appointments were attended (88.4%). Five implants required revision ≥1 year post-operatively, with revision requirements identified by surveillance in all cases. Three failures occurred at 5 years postoperatively, and two failures occurred at 8 years postoperatively. Overall, 1.7% attended surveillance appointments identified a failing implant requiring revision.</p><p><strong>Discussion: </strong>This is the first series reporting the effectiveness of primary linked total elbow arthroplasty surveillance in identifying implants requiring revision.</p>","PeriodicalId":36705,"journal":{"name":"Shoulder and Elbow","volume":" ","pages":"17585732241301356"},"PeriodicalIF":1.5000,"publicationDate":"2024-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11629365/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Shoulder and Elbow","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/17585732241301356","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Scoping review has identified a lack of evidence guiding long-term follow-up of elbow arthroplasty. We report the effectiveness of primary linked total elbow arthroplasty surveillance in identifying failing implants requiring revision.
Methods: A prospective database recording consecutive primary linked total elbow arthroplasty and subsequent surveillance in an elbow unit was analysed. Arthroplasties performed between 01.10.2013 and 31.07.2022 were included, with a minimum 1-year follow-up. Surveillance involves specialist physiotherapist review 1, 2, 3, 5, 8 and 10 years postoperatively. Patient-initiated review could occur between time points. Outcome measures include a number of surveillance reviews offered and attended; and the proportion that identified a failing implant requiring revision.
Results: Ninety-seven primary linked total elbow arthroplasties with minimum 1-year follow-up were performed (76 Discovery, 14 Nexel, 7 Coonrad/Morrey). Sixteen patients died prior to 31.07.2023, and three implants required revision <1 year postoperatively. 290 of 328 offered surveillance appointments were attended (88.4%). Five implants required revision ≥1 year post-operatively, with revision requirements identified by surveillance in all cases. Three failures occurred at 5 years postoperatively, and two failures occurred at 8 years postoperatively. Overall, 1.7% attended surveillance appointments identified a failing implant requiring revision.
Discussion: This is the first series reporting the effectiveness of primary linked total elbow arthroplasty surveillance in identifying implants requiring revision.