{"title":"How does intelligent noise reduction software influence the image quality in pelvic digital radiography; a phantom study.","authors":"E D Hussner, S Sundby, C B Outzen, J Jensen, A Tingberg, H Precht","doi":"10.1016/j.jmir.2024.101814","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>This study aims to evaluate the effects of a novel noise reduction software (INR, Canon Europe, Amsterdam, NL) on image quality (IQ) in Digital Radiography (DR) pelvic phantom images.</p><p><strong>Methods: </strong>In total, 53 pelvic phantom images and 360 technical images of a Contrast Detail Radiography phantom (CDRAD) were collected, including 8 different exposure levels (between 0.8 and 40 mAs at 70 kV) and 6 intensities of INR (ranging from settings 0-10). The pelvic images were evaluated by three reporting radiographers using absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed with a CDRAD computer software.</p><p><strong>Results: </strong>The VGA showed that the images with the INR software had higher IQ than the images with no INR. The observers gave a high VGA score to the images with INR of 3.2 mAs and higher. There was a tendency for an INR level of 5 or 7 to give the highest VGA scores. In addition, the CDRAD study showed a significant improvement in IQ with increasing INR levels at the lower exposure levels.</p><p><strong>Conclusion: </strong>An improvement in overall IQ was seen at lower exposure levels when the INR software was used both for the VGA and the CDRAD study. Clinical tests including patient images need to be performed before implementing INR in practice to verify accurate diagnostic performance.</p>","PeriodicalId":94092,"journal":{"name":"Journal of medical imaging and radiation sciences","volume":"56 2","pages":"101814"},"PeriodicalIF":0.0000,"publicationDate":"2024-12-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of medical imaging and radiation sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.jmir.2024.101814","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: This study aims to evaluate the effects of a novel noise reduction software (INR, Canon Europe, Amsterdam, NL) on image quality (IQ) in Digital Radiography (DR) pelvic phantom images.

Methods: In total, 53 pelvic phantom images and 360 technical images of a Contrast Detail Radiography phantom (CDRAD) were collected, including 8 different exposure levels (between 0.8 and 40 mAs at 70 kV) and 6 intensities of INR (ranging from settings 0-10). The pelvic images were evaluated by three reporting radiographers using absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed with a CDRAD computer software.

Results: The VGA showed that the images with the INR software had higher IQ than the images with no INR. The observers gave a high VGA score to the images with INR of 3.2 mAs and higher. There was a tendency for an INR level of 5 or 7 to give the highest VGA scores. In addition, the CDRAD study showed a significant improvement in IQ with increasing INR levels at the lower exposure levels.

Conclusion: An improvement in overall IQ was seen at lower exposure levels when the INR software was used both for the VGA and the CDRAD study. Clinical tests including patient images need to be performed before implementing INR in practice to verify accurate diagnostic performance.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信