Commissioning and implementation of a pencil-beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator.

IF 2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Sameer Taneja, Hesheng Wang, David L Barbee, Paulina Galavis, Mario Serrano Sosa, David Byun, Michael Zelefsky, Ting Chen
{"title":"Commissioning and implementation of a pencil-beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator.","authors":"Sameer Taneja, Hesheng Wang, David L Barbee, Paulina Galavis, Mario Serrano Sosa, David Byun, Michael Zelefsky, Ting Chen","doi":"10.1002/acm2.14590","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To commission a beam model in ClearCalc (Radformation Inc.) for use as a secondary dose calculation algorithm and to implement its use into an adaptive workflow for an MR-linear accelerator.</p><p><strong>Methods: </strong>A beam model was developed using commissioning data for an Elekta Unity MR-linear accelerator and entered into ClearCalc. The beam model consisted of absolute dose calculation settings, output factors, percent depth-dose (PDD) curves, mutli-leaf collimator (MLC) transmission and dose leaf gap error, and cryostat corrections. Beam profiles were hard-coded by the manufacturer into the beam model and were compared with Monaco-derived profiles. The beam model was tested by comparing point doses in a homogenous phantom obtained through measurements using an ionization chamber in water, Monaco, and ClearCalc for various field sizes, source-surface distances (SSDs), and point locations. Additional testing including point dose verification for test plans using a heterogeneous phantom and patient plans. Post clinical implementation, performance of ClearCalc was evaluated for the first 41 patients treated, which included 215 adaptive plans.</p><p><strong>Results: </strong>PDDs generated using ClearCalc fell within 1.2% of measurements. Field profile comparison between ClearCalc and Monaco showed an average pass rate of 98% using a 3%/3 mm gamma criteria. Measured cryostat corrections used in the beam model showed a maximum deviation from unity of 1.4%. Point dose and field monitor units (MUs) comparisons in a homogenous phantom (N = 22), heterogeneous phantoms (N = 22), and patient plans (N = 57) all passed with a threshold of 5%/5MU. Clinically, ClearCalc was implemented as a physics check post adaptive planning completed prior to beam delivery. Point dose and field MUs showed good agreement at a 5%/5MU threshold for prostate stereotactic body radiation therapy (SBRT), pelvic lymph nodes, rectum, and prostate and lymph node plans.</p><p><strong>Discussion: </strong>This work demonstrated commissioning and clinical implementation of ClearCalc into an adaptive planning workflow. No primary or adaptive plan failures were reported with proper beam model testing.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14590"},"PeriodicalIF":2.0000,"publicationDate":"2024-12-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14590","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To commission a beam model in ClearCalc (Radformation Inc.) for use as a secondary dose calculation algorithm and to implement its use into an adaptive workflow for an MR-linear accelerator.

Methods: A beam model was developed using commissioning data for an Elekta Unity MR-linear accelerator and entered into ClearCalc. The beam model consisted of absolute dose calculation settings, output factors, percent depth-dose (PDD) curves, mutli-leaf collimator (MLC) transmission and dose leaf gap error, and cryostat corrections. Beam profiles were hard-coded by the manufacturer into the beam model and were compared with Monaco-derived profiles. The beam model was tested by comparing point doses in a homogenous phantom obtained through measurements using an ionization chamber in water, Monaco, and ClearCalc for various field sizes, source-surface distances (SSDs), and point locations. Additional testing including point dose verification for test plans using a heterogeneous phantom and patient plans. Post clinical implementation, performance of ClearCalc was evaluated for the first 41 patients treated, which included 215 adaptive plans.

Results: PDDs generated using ClearCalc fell within 1.2% of measurements. Field profile comparison between ClearCalc and Monaco showed an average pass rate of 98% using a 3%/3 mm gamma criteria. Measured cryostat corrections used in the beam model showed a maximum deviation from unity of 1.4%. Point dose and field monitor units (MUs) comparisons in a homogenous phantom (N = 22), heterogeneous phantoms (N = 22), and patient plans (N = 57) all passed with a threshold of 5%/5MU. Clinically, ClearCalc was implemented as a physics check post adaptive planning completed prior to beam delivery. Point dose and field MUs showed good agreement at a 5%/5MU threshold for prostate stereotactic body radiation therapy (SBRT), pelvic lymph nodes, rectum, and prostate and lymph node plans.

Discussion: This work demonstrated commissioning and clinical implementation of ClearCalc into an adaptive planning workflow. No primary or adaptive plan failures were reported with proper beam model testing.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信