{"title":"Low dose threshold for measuring cardiac functional metrics using four-dimensional CT with deep learning.","authors":"Pengwei Wu, Kyle Kim, Lauren Severance, Elliot McVeigh, Jed Douglas Pack","doi":"10.1002/acm2.14593","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Four-dimensional CT is increasingly used for functional cardiac imaging, including prognosis for conditions such as heart failure and post myocardial infarction. However, radiation dose from an acquisition spanning the full cardiac cycle remains a concern. This work investigates the possibility of dose reduction in 4DCT using deep learning (DL)-based segmentation techniques as an objective observer.</p><p><strong>Methods: </strong>A 3D residual U-Net was developed for segmentation of left ventricle (LV) myocardium and blood pool. Two networks were trained: Standard DL (trained with only standard-dose [SD] data) and Noise-Robust DL (additionally trained with low-dose data). The primary goal of the proposed DL methods is to serve as an unbiased and consistent observer for functional analysis performance. Functional cardiac metrics including ejection fraction (EF), global longitudinal strain (GLS), circumferential strain (CS), and wall thickness (WT), were measured for an external test set of 250 Cardiac CT volumes reconstructed at five different dose levels.</p><p><strong>Results: </strong>Functional metrics obtained from DL segmentations of standard dose images matched well with those from expert manual analysis. Utilizing Standard-DL, absolute difference between DL-derived metrics obtained with standard dose data and 100 mA (corresponding to ∼76 ± 13% dose reduction) data was less than 0.8 ± 1.0% for EF, GLS, and CS, and 5.6 ± 6.7% for Average WT. Performance variation of Noise-Robust DL remained acceptable at even 50 mA.</p><p><strong>Conclusion: </strong>We demonstrate that on average radiation dose can be reduced by a factor of 5 while introducing minimal changes to global functional metrics (especially EF, GLS, and CS). The robustness to reduced image quality can be further boosted by using emulated low-dose data in the DL training set.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14593"},"PeriodicalIF":2.0000,"publicationDate":"2024-12-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14593","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Four-dimensional CT is increasingly used for functional cardiac imaging, including prognosis for conditions such as heart failure and post myocardial infarction. However, radiation dose from an acquisition spanning the full cardiac cycle remains a concern. This work investigates the possibility of dose reduction in 4DCT using deep learning (DL)-based segmentation techniques as an objective observer.
Methods: A 3D residual U-Net was developed for segmentation of left ventricle (LV) myocardium and blood pool. Two networks were trained: Standard DL (trained with only standard-dose [SD] data) and Noise-Robust DL (additionally trained with low-dose data). The primary goal of the proposed DL methods is to serve as an unbiased and consistent observer for functional analysis performance. Functional cardiac metrics including ejection fraction (EF), global longitudinal strain (GLS), circumferential strain (CS), and wall thickness (WT), were measured for an external test set of 250 Cardiac CT volumes reconstructed at five different dose levels.
Results: Functional metrics obtained from DL segmentations of standard dose images matched well with those from expert manual analysis. Utilizing Standard-DL, absolute difference between DL-derived metrics obtained with standard dose data and 100 mA (corresponding to ∼76 ± 13% dose reduction) data was less than 0.8 ± 1.0% for EF, GLS, and CS, and 5.6 ± 6.7% for Average WT. Performance variation of Noise-Robust DL remained acceptable at even 50 mA.
Conclusion: We demonstrate that on average radiation dose can be reduced by a factor of 5 while introducing minimal changes to global functional metrics (especially EF, GLS, and CS). The robustness to reduced image quality can be further boosted by using emulated low-dose data in the DL training set.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic