Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation.

IF 2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Emily Draeger, Fada Guan, Min-Young Lee, Dae Yup Han, William Donahue, Zhe Jay Chen
{"title":"Clinical use of Gafchromic EBT4 film for in vivo dosimetry for total body irradiation.","authors":"Emily Draeger, Fada Guan, Min-Young Lee, Dae Yup Han, William Donahue, Zhe Jay Chen","doi":"10.1002/acm2.14574","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In vivo dosimetry is a common requirement to validate dose accuracy/uniformity in total body irradiation (TBI). Several detectors can be used for in vivo dosimetry, including thermoluminescent dosimeters (TLDs), diodes, ion chambers, optically stimulated luminescent dosimeters (OSLDs), and film. TLDs are well established for use in vivo but required expertise and clinical system availability may make them impractical for multifractionated TBI. OSLDs offer quick readout, but recalls have restricted their use. The purpose of this work was to validate the newly available Gafchromic EBT4 film for TBI in vivo dosimetry.</p><p><strong>Methods: </strong>Film calibration curves were created under standard conditions (6MV/15MV, 1.5/3.0 cm depth, 100 cm source-to-surface distance (SSD), 10 × 10 cm<sup>2</sup> field), and films were scanned at several time points (0.5-24 h) to determine the shortest development time that yielded accurate dose measurements. 4 × 4 cm<sup>2</sup> films were placed under 1.5 cm thick bolus on the anterior and posterior sides of a solid water phantom to measure entrance and exit dose under TBI conditions (∼600 cm SSD, 39.5 × 39.5 cm<sup>2</sup> field, 6 MV/15 MV). These measurements were compared to ion chamber and diode readings for validation. Film measurements were also compared to OSLD measurements for three TBI patients.</p><p><strong>Results: </strong>The shortest development time that resulted in accurate dosimetry and allowed for adequate physician review time was 4 h (± 4% dose accuracy). Film entrance and exit dose measurements were within ± 3.8% of ion chamber and diode readings for 6MV and 15MV beams. Patient film measurements were within ∼ ± 5% for the majority of anatomical measurement locations; however, film and OSLD readings for some anatomic locations deviated by > 10%.</p><p><strong>Conclusions: </strong>These results indicate that EBT4 film can be utilized for accurate in vivo dosimetry for TBI patients and shows good agreement with diode and ion chamber measurements. Further investigation into film and OSLD differences was not possible due to OSLD recalls.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14574"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14574","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: In vivo dosimetry is a common requirement to validate dose accuracy/uniformity in total body irradiation (TBI). Several detectors can be used for in vivo dosimetry, including thermoluminescent dosimeters (TLDs), diodes, ion chambers, optically stimulated luminescent dosimeters (OSLDs), and film. TLDs are well established for use in vivo but required expertise and clinical system availability may make them impractical for multifractionated TBI. OSLDs offer quick readout, but recalls have restricted their use. The purpose of this work was to validate the newly available Gafchromic EBT4 film for TBI in vivo dosimetry.

Methods: Film calibration curves were created under standard conditions (6MV/15MV, 1.5/3.0 cm depth, 100 cm source-to-surface distance (SSD), 10 × 10 cm2 field), and films were scanned at several time points (0.5-24 h) to determine the shortest development time that yielded accurate dose measurements. 4 × 4 cm2 films were placed under 1.5 cm thick bolus on the anterior and posterior sides of a solid water phantom to measure entrance and exit dose under TBI conditions (∼600 cm SSD, 39.5 × 39.5 cm2 field, 6 MV/15 MV). These measurements were compared to ion chamber and diode readings for validation. Film measurements were also compared to OSLD measurements for three TBI patients.

Results: The shortest development time that resulted in accurate dosimetry and allowed for adequate physician review time was 4 h (± 4% dose accuracy). Film entrance and exit dose measurements were within ± 3.8% of ion chamber and diode readings for 6MV and 15MV beams. Patient film measurements were within ∼ ± 5% for the majority of anatomical measurement locations; however, film and OSLD readings for some anatomic locations deviated by > 10%.

Conclusions: These results indicate that EBT4 film can be utilized for accurate in vivo dosimetry for TBI patients and shows good agreement with diode and ion chamber measurements. Further investigation into film and OSLD differences was not possible due to OSLD recalls.

求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信