Brain MR-only workflow in clinical practice: A comparison among generators for quality assurance and patient positioning.

IF 2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Mathilde Levardon, Damien Autret, Thomas Le Dorze, Camille Guillerminet, Stéphane Dufreneix
{"title":"Brain MR-only workflow in clinical practice: A comparison among generators for quality assurance and patient positioning.","authors":"Mathilde Levardon, Damien Autret, Thomas Le Dorze, Camille Guillerminet, Stéphane Dufreneix","doi":"10.1002/acm2.14583","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong>Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.</p><p><strong>Materials and methods: </strong>On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.</p><p><strong>Results: </strong>Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.</p><p><strong>Conclusions: </strong>PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14583"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14583","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose: Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.

Materials and methods: On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.

Results: Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.

Conclusions: PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.

临床实践中的纯脑磁共振工作流程:质量保证和患者定位生成器之间的比较。
背景和目的:基于人工智能(AI)的 sCT 仍需要常规质量控制程序来验证发生器的性能。本研究旨在评估三种基于人工智能或体密度(BD)分配的发生器,以便在临床常规中对另一种基于人工智能的发生器进行患者特异性质量保证(PSQA)。此外,还进行了一项基于 2D/2D kV-image 的患者定位研究,比较了四种 sCT 发生器的性能:在我院现有的四台商用发生器中,选择一台作为临床使用,其他三台用于 PSQA。计算了几个剂量指标,如平均误差、剂量-体积直方图指标和 1%/1毫米伽马分析。与 CT 的比较被视为参考。基于 sCT 的病人定位过程中发现的平移和旋转与基于 CT 的病人定位过程中发现的平移和旋转进行了比较:结果:根据 CT 计算出的一些指标显示患者超出了所选的容差范围(点指标为 1%;伽马通过率为 90%)。在所有研究指标中,没有一个生成器能识别出这些异常值。使用其他 sCT 生成器进行 PSQA 时,会出现一些假阳性和假阴性。在所有研究指标中,没有一个发生器能够明确识别金属植入物导致的真正 sCT 故障。BD 分配 sCT 的定位偏差最小,而唯一不基于 T1 Dixon 磁共振序列的人工智能发生器的定位偏差最大:对一台 sCT 发生器和另一台 sCT 发生器进行 PSQA 时应非常小心。患者定位是评估 sCT 发生器时需要考虑的一个重要方面。这项研究的结果将有助于医学物理学家根据二维/二维 kV 图像患者定位建立仅用于脑部的磁共振工作流程。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信