Brandon Huynh, Colin Clement, Vuong Nguyen, Stephen O'Hagan, Frank Howes, Peter Macken, David Manning, Ridia Lim, Jed Lusthaus, Mitchell Lawlor
{"title":"12-month Safety and Efficacy Outcomes of a Standalone Trabecular Bypass Device.","authors":"Brandon Huynh, Colin Clement, Vuong Nguyen, Stephen O'Hagan, Frank Howes, Peter Macken, David Manning, Ridia Lim, Jed Lusthaus, Mitchell Lawlor","doi":"10.5005/jp-journals-10078-1447","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To analyze the outcomes of the iStent inject in a real-world clinical setting as a standalone procedure to lower intraocular pressure (IOP) in open-angle glaucoma.</p><p><strong>Materials and methods: </strong>Patients with open-angle glaucoma having undergone iStent inject insertion without concurrent cataract extraction were included in this multicenter observational real-world study in Australia. Patient data was entered into the Fight Glaucoma Blindness! Registry. Assessments through 12 months included glaucoma subtype, IOP, medications, best-corrected visual acuity (BCVA), secondary surgical procedures, and adverse events. Kaplan-Meier survival curves for outcomes were reported according to the World Glaucoma Association (WGA).</p><p><strong>Results: </strong>Sixty-one eyes from 44 patients with a mean age of 76 ± 11.4 underwent standalone iStent inject implantation. The mean ± SD preoperative IOP was 17.5 ± 7.5 mm Hg, and the mean preoperative number of topical medications was 2.5 ± 1.5. At 12 months postoperatively, there was no statistically significant IOP reduction, while the number of glaucoma medications used was reduced to 1.4 ± 1.5 (<i>p</i> < 0.001). Fourteen point one percent of eyes required a secondary pressure-lowering procedure within the 12-month follow-up window.</p><p><strong>Conclusion: </strong>This assessment of standalone iStent inject implantation did not show any significant reduction in IOP, but there was a significant decrease in medication use in the real-world clinical setting. The procedure is safe with minimal adverse outcomes; however, a subset of patients required secondary procedures within 12 months of follow-up.</p><p><strong>How to cite this article: </strong>Huynh B, Clement C, Nguyen V, <i>et al.</i> 12-month Safety and Efficacy Outcomes of a Standalone Trabecular Bypass Device. J Curr Glaucoma Pract 2024;18(3):103-109.</p>","PeriodicalId":15419,"journal":{"name":"Journal of Current Glaucoma Practice","volume":"18 3","pages":"103-109"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11576339/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Current Glaucoma Practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5005/jp-journals-10078-1447","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/29 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To analyze the outcomes of the iStent inject in a real-world clinical setting as a standalone procedure to lower intraocular pressure (IOP) in open-angle glaucoma.
Materials and methods: Patients with open-angle glaucoma having undergone iStent inject insertion without concurrent cataract extraction were included in this multicenter observational real-world study in Australia. Patient data was entered into the Fight Glaucoma Blindness! Registry. Assessments through 12 months included glaucoma subtype, IOP, medications, best-corrected visual acuity (BCVA), secondary surgical procedures, and adverse events. Kaplan-Meier survival curves for outcomes were reported according to the World Glaucoma Association (WGA).
Results: Sixty-one eyes from 44 patients with a mean age of 76 ± 11.4 underwent standalone iStent inject implantation. The mean ± SD preoperative IOP was 17.5 ± 7.5 mm Hg, and the mean preoperative number of topical medications was 2.5 ± 1.5. At 12 months postoperatively, there was no statistically significant IOP reduction, while the number of glaucoma medications used was reduced to 1.4 ± 1.5 (p < 0.001). Fourteen point one percent of eyes required a secondary pressure-lowering procedure within the 12-month follow-up window.
Conclusion: This assessment of standalone iStent inject implantation did not show any significant reduction in IOP, but there was a significant decrease in medication use in the real-world clinical setting. The procedure is safe with minimal adverse outcomes; however, a subset of patients required secondary procedures within 12 months of follow-up.
How to cite this article: Huynh B, Clement C, Nguyen V, et al. 12-month Safety and Efficacy Outcomes of a Standalone Trabecular Bypass Device. J Curr Glaucoma Pract 2024;18(3):103-109.