A Case of Very Early Lead Fracture in an Implantable Cardioverter-Defibrillator: Management and Follow-Up.

IF 1.7 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Pace-Pacing and Clinical Electrophysiology Pub Date : 2025-01-01 Epub Date: 2024-11-19 DOI:10.1111/pace.15103
Marwan Shawki, Sam Mirzaee, Cameron McCormack, Gareth J Wynn, Han S Lim
{"title":"A Case of Very Early Lead Fracture in an Implantable Cardioverter-Defibrillator: Management and Follow-Up.","authors":"Marwan Shawki, Sam Mirzaee, Cameron McCormack, Gareth J Wynn, Han S Lim","doi":"10.1111/pace.15103","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Implantable cardioverter defibrillators (ICDs) lead fractures are rare but serious complications, often leading to inappropriate shocks. The early occurrence of such fractures post-implantation is exceptionally uncommon.</p><p><strong>Methods and results: </strong>We present a case of a 53-year-old male with hypertrophic cardiomyopathy who experienced an inappropriate shock due to a lead fracture just 4 days after ICD implantation. The fractured lead was successfully extracted and replaced using a gooseneck snare passed through an oversized TightRail with no immediate complications.</p><p><strong>Conclusion: </strong>Early lead fractures can occur shortly after ICD implantation, leading to inappropriate shocks. Successful management, including extraction and reimplantation, is possible even in these early cases. Clinicians should remain vigilant, and remote monitoring can play a crucial role in the timely detection of such complications.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"68-71"},"PeriodicalIF":1.7000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15103","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/19 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Implantable cardioverter defibrillators (ICDs) lead fractures are rare but serious complications, often leading to inappropriate shocks. The early occurrence of such fractures post-implantation is exceptionally uncommon.

Methods and results: We present a case of a 53-year-old male with hypertrophic cardiomyopathy who experienced an inappropriate shock due to a lead fracture just 4 days after ICD implantation. The fractured lead was successfully extracted and replaced using a gooseneck snare passed through an oversized TightRail with no immediate complications.

Conclusion: Early lead fractures can occur shortly after ICD implantation, leading to inappropriate shocks. Successful management, including extraction and reimplantation, is possible even in these early cases. Clinicians should remain vigilant, and remote monitoring can play a crucial role in the timely detection of such complications.

一例植入式心律转复除颤器的极早期导线断裂:处理和随访。
导言:植入式心律转复除颤器(ICD)导联骨折是一种罕见但严重的并发症,通常会导致不适当的电击。植入后早期发生此类骨折的情况极为罕见:我们介绍了一例 53 岁男性肥厚型心肌病患者的病例,他在植入 ICD 仅 4 天后就因导联线断裂而发生了不适当电击。使用鹅颈套管通过超大的 TightRail 成功取出并更换了断裂的导联,没有立即出现并发症:结论:ICD 植入后不久即可发生早期导联断裂,从而导致不适当的电击。即使在这些早期病例中也有可能成功处理,包括取出和再植入。临床医生应保持警惕,远程监控在及时发现此类并发症方面可发挥至关重要的作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Pace-Pacing and Clinical Electrophysiology
Pace-Pacing and Clinical Electrophysiology 医学-工程:生物医学
CiteScore
2.70
自引率
5.60%
发文量
209
审稿时长
2-4 weeks
期刊介绍: Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信