Exploring stereotactic radiosurgery for tremor using the Varian cone planning system

IF 3.3 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Kyra N. McComas , Guozhen Luo , George X. Ding , Stephen Martinez , Michael J. Price , Austin N. Kirschner
{"title":"Exploring stereotactic radiosurgery for tremor using the Varian cone planning system","authors":"Kyra N. McComas ,&nbsp;Guozhen Luo ,&nbsp;George X. Ding ,&nbsp;Stephen Martinez ,&nbsp;Michael J. Price ,&nbsp;Austin N. Kirschner","doi":"10.1016/j.ejmp.2024.104858","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Stereotactic radiosurgery (SRS) is an effective treatment for essential tremor (ET) and Parkinsonian tremor (PT). However, current treatment methods can be time-consuming and expose patients to unnecessary radiation. This study investigates the use of the Varian TrueBeam system with FFF beam modes as a potential solution for these issues.</div></div><div><h3>Methods</h3><div>We recreated and evaluated treatment plans using the Varian TrueBeam and Eclipse Cone Planning system for ten patients previously treated for tremor cessation using a Novalis Tx linear accelerator with Brainlab planning system prescribed a maximum dose of 160 Gy. Analyses included beam-on delivery time, total MU, dose to brain and brainstem tissues, and confirmation of SRS cone accuracy.</div></div><div><h3>Results</h3><div>The TrueBeam cone delivery system is highly accurate. Our results demonstrated a significant reduction in monitor units (MU) and treatment delivery time shortened to 17-26 min beam-on time for all-Varian plans compared to 39 min for non-Varian (iPlan) plans (p &lt; 0.0001). Furthermore, the Varian plans showed improved or similar dose fall-off and radiation dose to brainstem normal tissue.</div></div><div><h3>Conclusion</h3><div>Our findings suggest that the Varian TrueBeam system offers a safe and effective option for SRS tremor treatment, providing similar or superior treatment parameters compared to other LINAC-based SRS treatments. The reduced treatment times and improved patient comfort offered by this system make it a promising solution for future tremor treatments.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"128 ","pages":"Article 104858"},"PeriodicalIF":3.3000,"publicationDate":"2024-11-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179724013267","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Stereotactic radiosurgery (SRS) is an effective treatment for essential tremor (ET) and Parkinsonian tremor (PT). However, current treatment methods can be time-consuming and expose patients to unnecessary radiation. This study investigates the use of the Varian TrueBeam system with FFF beam modes as a potential solution for these issues.

Methods

We recreated and evaluated treatment plans using the Varian TrueBeam and Eclipse Cone Planning system for ten patients previously treated for tremor cessation using a Novalis Tx linear accelerator with Brainlab planning system prescribed a maximum dose of 160 Gy. Analyses included beam-on delivery time, total MU, dose to brain and brainstem tissues, and confirmation of SRS cone accuracy.

Results

The TrueBeam cone delivery system is highly accurate. Our results demonstrated a significant reduction in monitor units (MU) and treatment delivery time shortened to 17-26 min beam-on time for all-Varian plans compared to 39 min for non-Varian (iPlan) plans (p < 0.0001). Furthermore, the Varian plans showed improved or similar dose fall-off and radiation dose to brainstem normal tissue.

Conclusion

Our findings suggest that the Varian TrueBeam system offers a safe and effective option for SRS tremor treatment, providing similar or superior treatment parameters compared to other LINAC-based SRS treatments. The reduced treatment times and improved patient comfort offered by this system make it a promising solution for future tremor treatments.
探索使用瓦里安锥形规划系统进行震颤的立体定向放射外科治疗。
背景:立体定向放射手术(SRS)是治疗本质性震颤(ET)和帕金森震颤(PT)的有效方法。然而,目前的治疗方法可能会耗费大量时间,并使患者遭受不必要的辐射。本研究探讨了使用瓦里安 TrueBeam 系统的 FFF 射束模式作为解决这些问题的潜在方案:我们使用瓦里安TrueBeam和Eclipse锥形规划系统重新创建并评估了10名患者的治疗计划,这些患者之前使用Novalis Tx直线加速器和Brainlab规划系统接受过震颤停止治疗,最大剂量为160 Gy。分析内容包括光束发射时间、总MU、大脑和脑干组织的剂量,以及SRS锥体准确性的确认:结果:TrueBeam锥体投射系统的精确度很高。我们的结果表明,所有瓦里安计划的监测单位(MU)显著减少,治疗投放时间缩短至17-26分钟,而非瓦里安(iPlan)计划的投放时间为39分钟(P 结论:我们的研究结果表明,瓦里安TrueBeam锥体投放系统能显著减少监测单位(MU),治疗投放时间缩短至17-26分钟:我们的研究结果表明,瓦里安 TrueBeam 系统为震颤 SRS 治疗提供了一个安全有效的选择,与其他基于 LINAC 的 SRS 治疗相比,它能提供相似或更优越的治疗参数。该系统缩短了治疗时间,提高了患者的舒适度,是未来震颤治疗的理想解决方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信