[Impact of Different Acquisition Modes on Image Quality and Quantitative Accuracy for Bone SPECT Using Ordered Subset Conjugate-gradient Minimizer].

Nihon Hoshasen Gijutsu Gakkai zasshi Pub Date : 2024-11-20 Epub Date: 2024-10-17 DOI:10.6009/jjrt.2024-1455
Yui Orisaka, Takayuki Shibutani, Takahiro Konishi, Anji Yokouchi, Katsutoshi Fujiwara, Hiroto Yoneyama
{"title":"[Impact of Different Acquisition Modes on Image Quality and Quantitative Accuracy for Bone SPECT Using Ordered Subset Conjugate-gradient Minimizer].","authors":"Yui Orisaka, Takayuki Shibutani, Takahiro Konishi, Anji Yokouchi, Katsutoshi Fujiwara, Hiroto Yoneyama","doi":"10.6009/jjrt.2024-1455","DOIUrl":null,"url":null,"abstract":"<p><p>We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and <sup>99m</sup>Tc was filled with the SIM<sup>2</sup> Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuous<sub>view</sub>, continuous<sub>total</sub>) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuous<sub>view</sub>, and continuous<sub>total</sub> methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"1191-1197"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1455","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/17 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and 99mTc was filled with the SIM2 Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuousview, continuoustotal) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuousview, and continuoustotal methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.

[不同采集模式对使用有序子集共轭梯度最小化器进行骨 SPECT 的图像质量和定量准确性的影响]。
我们评估了不同的 xSPECT 骨采集方法对图像质量和定量准确性的影响。在 SIM2 骨模型中注入了设定为 250 HU 和 99mTc 的混合骨当量溶液。xSPECT Bone 采集采用步进拍摄(SS)模式和两种连续模式(continuousview、continuousoustotal),每个视图的采集时间和总采集时间与 SS 模式相同。正常骨骼和肿瘤的放射性浓度比设定为 1:6。用均匀度、对比度、空间分辨率、肿瘤部分的恢复系数(RC)和可探测性对重建图像进行评估。可探测性评分(DS)是对可探测性的总体评价,由均匀度和对比度的结果计算得出。SS、continuousview和continuousoustotal方法的均匀度分别为29.4%、17.4%和11.1%。SS 模式和两种连续模式的空间分辨率、RC 和可探测性相似;然而,与 SS 模式相比,连续模式显示出良好的均匀性和对比度。此外,连续模式的可探测性等于或优于 SS 模式。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信