Full-Waveform Inversion Imaging of Cortical Bone Using Phased Array Tomography.

IF 4.4 2区 医学 Q2 ENGINEERING, BIOMEDICAL
Lexiu Xu, Yifang Li, Yuan Liu, Qinzhen Shi, Wenyu Xing, Tao Jiang, Gaobo Zhang, Ying Li, Dean Ta
{"title":"Full-Waveform Inversion Imaging of Cortical Bone Using Phased Array Tomography.","authors":"Lexiu Xu, Yifang Li, Yuan Liu, Qinzhen Shi, Wenyu Xing, Tao Jiang, Gaobo Zhang, Ying Li, Dean Ta","doi":"10.1109/TBME.2024.3477708","DOIUrl":null,"url":null,"abstract":"<p><p>Classic ultrasound bone imaging modalities usually demand either a prior knowledge or an advanced estimation on speed of sound (SoS), which not only renders to a burdensome imaging process but also supplies a limited resolution. To overcome these drawbacks, this article proposed a frequency-domain full-waveform inversion (FDFWI) modality using phased array tomography for high-accuracy cortical bone imaging. A transmission scenario of ultrasound wave in 2-D space was presented in the frequency domain to simulate the forward wavefield propagation. Iterations in the inversion process were performed by matching the simulation wavefield to the experimental one from low to high discrete frequency points. Moreover, the association between the maximum initial frequency and the initial SoS model was explored to prevent the occurrence of cycle-skipping phenomenon, which could lead to the outcomes being trapped in local minima. The feasibility and effectiveness of the proposed imaging scheme were testified by simulation, phantom, and ex-vivo studies, with mean relative errors of cortical part being 3.18%, 8.71%, and 9.36%, respectively. It is verified that the proposed FDFWI method is an effective way for parametric imaging of cortical bone without any prior knowledge of sound speed.</p>","PeriodicalId":13245,"journal":{"name":"IEEE Transactions on Biomedical Engineering","volume":"PP ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2024-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1109/TBME.2024.3477708","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Classic ultrasound bone imaging modalities usually demand either a prior knowledge or an advanced estimation on speed of sound (SoS), which not only renders to a burdensome imaging process but also supplies a limited resolution. To overcome these drawbacks, this article proposed a frequency-domain full-waveform inversion (FDFWI) modality using phased array tomography for high-accuracy cortical bone imaging. A transmission scenario of ultrasound wave in 2-D space was presented in the frequency domain to simulate the forward wavefield propagation. Iterations in the inversion process were performed by matching the simulation wavefield to the experimental one from low to high discrete frequency points. Moreover, the association between the maximum initial frequency and the initial SoS model was explored to prevent the occurrence of cycle-skipping phenomenon, which could lead to the outcomes being trapped in local minima. The feasibility and effectiveness of the proposed imaging scheme were testified by simulation, phantom, and ex-vivo studies, with mean relative errors of cortical part being 3.18%, 8.71%, and 9.36%, respectively. It is verified that the proposed FDFWI method is an effective way for parametric imaging of cortical bone without any prior knowledge of sound speed.

利用相控阵层析成像技术对皮质骨进行全波形反演成像
传统的超声骨成像模式通常需要先验知识或先进的声速(SoS)估计,这不仅造成成像过程繁琐,而且分辨率有限。为了克服这些缺点,本文提出了一种利用相控阵层析成像技术进行高精度皮质骨成像的频域全波形反转(FDFWI)模式。在频域中提出了超声波在二维空间中的传播情况,以模拟前向波场传播。反演过程中的迭代是通过将模拟波场与实验波场从低频点到高频点进行离散匹配。此外,还探讨了最大初始频率与初始 SoS 模型之间的关联,以防止出现周期跳跃现象,导致结果陷入局部极小值。模拟、模型和体外研究证明了拟议成像方案的可行性和有效性,皮质部分的平均相对误差分别为 3.18%、8.71% 和 9.36%。验证了所提出的 FDFWI 方法是在不预先了解声速的情况下对骨皮质进行参数成像的有效方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Biomedical Engineering
IEEE Transactions on Biomedical Engineering 工程技术-工程:生物医学
CiteScore
9.40
自引率
4.30%
发文量
880
审稿时长
2.5 months
期刊介绍: IEEE Transactions on Biomedical Engineering contains basic and applied papers dealing with biomedical engineering. Papers range from engineering development in methods and techniques with biomedical applications to experimental and clinical investigations with engineering contributions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信