Zeyad Alawaji, Seyedamir Tavakoli Taba, Lucy Cartwright, William Rae
{"title":"Automated quality control analysis for American College of Radiology (ACR) digital mammography (DM) phantom images","authors":"Zeyad Alawaji, Seyedamir Tavakoli Taba, Lucy Cartwright, William Rae","doi":"10.1002/acm2.14548","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To develop and validate an automated software analysis method for mammography image quality assessment of the American College of Radiology (ACR) digital mammography (DM) phantom images.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twenty-seven DICOM images were acquired using Fuji mammography systems. All images were evaluated by three expert medical physicists using the Royal Australian and New Zealand College of Radiologists (RANZCR) mammography quality control guideline. To enhance the robustness and sensitivity assessment of our algorithm, an additional set of 12 images from a Hologic mammography system was included to test various phantom positional adjustments. The software automatically chose multiple regions of interest (ROIs) for analysis. A template matching method was primarily used for image analysis, followed by an additional method that locates and scores each target object (speck groups, fibers, and masses).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The software performance shows a good to excellent agreement with the average scoring of observers (intraclass correlation coefficient [ICC] of 0.75, 0.79, 0.82 for speck groups, fibers, and masses, respectively). No significant differences were found in the scoring of target objects between human observers and the software. Both methods achieved scores meeting the pass criteria for speck groups and masses. Expert observers allocated lower scores to fiber objects, with diameters less than 0.61 mm, when compared to the software. The software was able to accurately score objects when the phantom position changed by up to 25 mm laterally, up to 5 degrees rotation, and overhanging the chest wall edge by up to 15 mm.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Automated software analysis is a feasible method that may help improve the consistency and reproducibility of mammography image quality assessment with reduced reliance on human interaction and processing time.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"25 12","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2024-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11633796/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.14548","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To develop and validate an automated software analysis method for mammography image quality assessment of the American College of Radiology (ACR) digital mammography (DM) phantom images.
Methods
Twenty-seven DICOM images were acquired using Fuji mammography systems. All images were evaluated by three expert medical physicists using the Royal Australian and New Zealand College of Radiologists (RANZCR) mammography quality control guideline. To enhance the robustness and sensitivity assessment of our algorithm, an additional set of 12 images from a Hologic mammography system was included to test various phantom positional adjustments. The software automatically chose multiple regions of interest (ROIs) for analysis. A template matching method was primarily used for image analysis, followed by an additional method that locates and scores each target object (speck groups, fibers, and masses).
Results
The software performance shows a good to excellent agreement with the average scoring of observers (intraclass correlation coefficient [ICC] of 0.75, 0.79, 0.82 for speck groups, fibers, and masses, respectively). No significant differences were found in the scoring of target objects between human observers and the software. Both methods achieved scores meeting the pass criteria for speck groups and masses. Expert observers allocated lower scores to fiber objects, with diameters less than 0.61 mm, when compared to the software. The software was able to accurately score objects when the phantom position changed by up to 25 mm laterally, up to 5 degrees rotation, and overhanging the chest wall edge by up to 15 mm.
Conclusions
Automated software analysis is a feasible method that may help improve the consistency and reproducibility of mammography image quality assessment with reduced reliance on human interaction and processing time.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic