Nico Fransaert, Allyson Robert, Bart Cleuren, Jean V Manca, Dirk Valkenborg
{"title":"Identifying Process Differences with ToF-SIMS: An MVA Decomposition Strategy.","authors":"Nico Fransaert, Allyson Robert, Bart Cleuren, Jean V Manca, Dirk Valkenborg","doi":"10.1021/jasms.4c00327","DOIUrl":null,"url":null,"abstract":"<p><p>In time-of-flight secondary ion mass spectrometry (ToF-SIMS), multivariate analysis (MVA) methods such as principal component analysis (PCA) are routinely employed to differentiate spectra. However, additional insights can often be gained by comparing processes, where each process is characterized by its own start and end spectra, such as when identical samples undergo slightly different treatments or when slightly different samples receive the same treatment. This study proposes a strategy to compare such processes by decomposing the loading vectors associated with them, which highlights differences in the relative behavior of the peaks. This strategy identifies key information beyond what is captured by the loading vectors or the end spectra alone. While PCA is widely used, partial least-squares discriminant analysis (PLS-DA) serves as a supervised alternative and is the preferred method for deriving process-related loading vectors when classes are narrowly separated. The effectiveness of the decomposition strategy is demonstrated using artificial spectra and applied to a ToF-SIMS materials science case study on the photodegradation of N719 dye, a common dye in photovoltaics, on a mesoporous TiO<sub>2</sub> anode. The study revealed that the photodegradation process varies over time, and the resulting fragments have been identified accordingly. The proposed methodology, applicable to both labeled (supervised) and unlabeled (unsupervised) spectral data, can be seamlessly integrated into most modern mass spectrometry data analysis workflows to automatically generate a list of peaks whose relative behavior varies between two processes, and is particularly effective in identifying subtle differences between highly similar physicochemical processes.</p>","PeriodicalId":672,"journal":{"name":"Journal of the American Society for Mass Spectrometry","volume":" ","pages":"3116-3125"},"PeriodicalIF":3.1000,"publicationDate":"2024-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11622371/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the American Society for Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.1021/jasms.4c00327","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/4 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
In time-of-flight secondary ion mass spectrometry (ToF-SIMS), multivariate analysis (MVA) methods such as principal component analysis (PCA) are routinely employed to differentiate spectra. However, additional insights can often be gained by comparing processes, where each process is characterized by its own start and end spectra, such as when identical samples undergo slightly different treatments or when slightly different samples receive the same treatment. This study proposes a strategy to compare such processes by decomposing the loading vectors associated with them, which highlights differences in the relative behavior of the peaks. This strategy identifies key information beyond what is captured by the loading vectors or the end spectra alone. While PCA is widely used, partial least-squares discriminant analysis (PLS-DA) serves as a supervised alternative and is the preferred method for deriving process-related loading vectors when classes are narrowly separated. The effectiveness of the decomposition strategy is demonstrated using artificial spectra and applied to a ToF-SIMS materials science case study on the photodegradation of N719 dye, a common dye in photovoltaics, on a mesoporous TiO2 anode. The study revealed that the photodegradation process varies over time, and the resulting fragments have been identified accordingly. The proposed methodology, applicable to both labeled (supervised) and unlabeled (unsupervised) spectral data, can be seamlessly integrated into most modern mass spectrometry data analysis workflows to automatically generate a list of peaks whose relative behavior varies between two processes, and is particularly effective in identifying subtle differences between highly similar physicochemical processes.
期刊介绍:
The Journal of the American Society for Mass Spectrometry presents research papers covering all aspects of mass spectrometry, incorporating coverage of fields of scientific inquiry in which mass spectrometry can play a role.
Comprehensive in scope, the journal publishes papers on both fundamentals and applications of mass spectrometry. Fundamental subjects include instrumentation principles, design, and demonstration, structures and chemical properties of gas-phase ions, studies of thermodynamic properties, ion spectroscopy, chemical kinetics, mechanisms of ionization, theories of ion fragmentation, cluster ions, and potential energy surfaces. In addition to full papers, the journal offers Communications, Application Notes, and Accounts and Perspectives