An emf-temperature reference function for Pt–20%Rh versus Pt thermocouples in the temperature range of 0 °C to 1100 °C

IF 2.1 3区 工程技术 Q2 INSTRUMENTS & INSTRUMENTATION
F Edler, P Ederer, K Huang
{"title":"An emf-temperature reference function for Pt–20%Rh versus Pt thermocouples in the temperature range of 0 °C to 1100 °C","authors":"F Edler, P Ederer, K Huang","doi":"10.1088/1681-7575/ad746b","DOIUrl":null,"url":null,"abstract":"An electromotive force–temperature reference function for thermocouples made of platinum–20%rhodium versus platinum (Pt–20%Rh/Pt) was determined for the temperature range from 0 °C to 1100 °C. Three Pt–20%Rh/Pt thermocouples were constructed at the Physikalisch-Technische Bundesanstalt (PTB) and were compared with standard platinum resistance thermometers, an Au/Pt thermocouple and a Pt/Pd thermocouple, which were all calibrated at fixed points of the International Temperature Scale of 1990 (ITS-90). The comparison measurements were performed in stirred liquid baths in the range from 0 °C to 545 °C, in a sodium heat pipe from 560 °C to 980 °C, and in a five-zone furnace at temperatures between 1000 °C and 1100 °C. Furthermore, the Pt–20%Rh/Pt thermocouples were calibrated at fixed points of the ITS-90. The PTB reference function based on two of the three Pt–20%Rh/Pt thermocouples has an expanded measurement uncertainty of 0.25 K in the temperature range from 0 °C to 1100 °C.","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"42 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-09-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1681-7575/ad746b","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

An electromotive force–temperature reference function for thermocouples made of platinum–20%rhodium versus platinum (Pt–20%Rh/Pt) was determined for the temperature range from 0 °C to 1100 °C. Three Pt–20%Rh/Pt thermocouples were constructed at the Physikalisch-Technische Bundesanstalt (PTB) and were compared with standard platinum resistance thermometers, an Au/Pt thermocouple and a Pt/Pd thermocouple, which were all calibrated at fixed points of the International Temperature Scale of 1990 (ITS-90). The comparison measurements were performed in stirred liquid baths in the range from 0 °C to 545 °C, in a sodium heat pipe from 560 °C to 980 °C, and in a five-zone furnace at temperatures between 1000 °C and 1100 °C. Furthermore, the Pt–20%Rh/Pt thermocouples were calibrated at fixed points of the ITS-90. The PTB reference function based on two of the three Pt–20%Rh/Pt thermocouples has an expanded measurement uncertainty of 0.25 K in the temperature range from 0 °C to 1100 °C.
0 °C 至 1100 °C 温度范围内 Pt-20%Rh 与铂热电偶的电势-温度参考函数
在 0 °C 至 1100 °C 的温度范围内,测定了由铂-20% 铑与铂(Pt-20%Rh/Pt)制成的热电偶的电动势-温度参考函数。在德国联邦物理技术局(PTB)制造了三个 Pt-20%Rh/Pt 热电偶,并将其与标准铂电阻温度计、金/铂热电偶和铂/钯热电偶进行了比较,这些热电偶都在 1990 年国际温标(ITS-90)的固定点上进行了校准。对比测量是在 0 °C 至 545 °C 的搅拌液浴、560 °C 至 980 °C 的钠热管以及 1000 °C 至 1100 °C 的五区炉中进行的。此外,还在 ITS-90 的固定点对 Pt-20%Rh/Pt 热电偶进行了校准。在 0 °C 至 1100 °C 的温度范围内,基于三个 Pt-20%Rh/Pt 热电偶中的两个的 PTB 参考函数的测量不确定度为 0.25 K。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Metrologia
Metrologia 工程技术-物理:应用
CiteScore
2.80
自引率
25.00%
发文量
137
审稿时长
12 months
期刊介绍: Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them. The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics. In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信