Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, Luka Pavelić, Mihael Makek
{"title":"Investigation of the spatial resolution of PET imaging system measuring polarization-correlated Compton events","authors":"Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, Luka Pavelić, Mihael Makek","doi":"arxiv-2409.01238","DOIUrl":null,"url":null,"abstract":"Recent studies of positron emission tomography (PET) devices have shown that\nthe detection of polarization-correlated annihilation quanta can potentially\nreduce the background and creation of false lines of response (LORs) leading to\nimproved image quality. We developed a novel PET demonstrator system, capable\nof measuring correlated gamma photons with single-layer Compton polarimeters to\nexplore the potential of the method. We tested the system using sources with\nclinically relevant activities at the University Hospital Centre Zagreb. Here\nwe present, for the first time, the images of two Ge-68 line sources,\nreconstructed solely from the correlated annihilation events. The spatial\nresolution at two different diameters is determined and compared to the one\nobtained from events with photoelectric interaction.","PeriodicalId":501378,"journal":{"name":"arXiv - PHYS - Medical Physics","volume":"87 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"arXiv - PHYS - Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/arxiv-2409.01238","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Recent studies of positron emission tomography (PET) devices have shown that
the detection of polarization-correlated annihilation quanta can potentially
reduce the background and creation of false lines of response (LORs) leading to
improved image quality. We developed a novel PET demonstrator system, capable
of measuring correlated gamma photons with single-layer Compton polarimeters to
explore the potential of the method. We tested the system using sources with
clinically relevant activities at the University Hospital Centre Zagreb. Here
we present, for the first time, the images of two Ge-68 line sources,
reconstructed solely from the correlated annihilation events. The spatial
resolution at two different diameters is determined and compared to the one
obtained from events with photoelectric interaction.